2nd Year: Commissioning and General Machine/TPS Knowledge Flashcards

1
Q

If you take a PDD curve, and the curve is parallel but offset to what gold beam data expects, is this an issue with the Linac? Why or why not?

A

Most likely not

If energy/beam quality differs, the curve will be skewed in a non-parallel manner (attenuation differences)

If it just appears offset, you most likely just have a bad setup on measurement

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

If energy of a flattened beam is higher than expected, what happens to the horns?

A

Horns increase

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

What beam characteristic is generally the best distinguisher of beam energy?

A

Flatness (for flattened beams)

Ex. for a 6M beam, a 1% error in energy presents as a 4% error in profile

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

List some equipment that is required for acquisition of commissioning data

A

3D Water Tank with large enough dimensions for full scatter

Ion chamber for TG-51

Small field detector

Scanning chambers/detectors

Levels for the tank and gantry head

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

For water tank scanning, how is a CAX correction applied? How is it measured?

A

Measured using geometry that extracts inplane offset, crossplane offset and angles based off in-plane and crossplane profiles at 2 or 3 different depths

Translational corrections can be applied to all data collected therafter, but rotational corrections cannot be.

For large fields, not applying rotational corrections does not make a big difference. For small fields, it does so you want rotational CAX measurement to be as close to 0 as possible

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

For tanks that are unable to measure inplane, crossplane and diagonal profiles for the largest field sizes, what method is used to do this?

A

Shifted tank method

Center field at a quadrant of the tank, with approx 5 cm from any tank border from crosshair and measure only towards center of tank

With this method you are measuring half the profile, which is fine because in TPS you would just mirror the profile anyway

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Why for commissioning do we do 10 x 10 PDD profiles for FFF beams with and without the lead foil?

A

TG-51 requires it with the lead foil

TPS requires it WITHOUT the lead foil. Which makes complete sense as you never treat a patient with lead foil

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

At what depths does Eclipse TPS want inplane and crossplane profiles for photons?

A

max, 5 cm, 10 cm, 20 cm, 30 cm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

How do you measure Sp?

A

You don’t

You measure Scp in a water tank

You measure Sc in air

You then divide Scp/Sc for all respective field sizes to get Sp

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

How is DLG measured?

A

Varian created template plan where there’s small slit of opening with MLCs that slides across the field with varying gap sizes

Detector measures integrated dose for each slit size plans

Create a plot of integrated dose vs gap size

Extrapolate down to a gap of 0 mm and that gives you your DLG

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

What is the difference between inplane and crossplane?

A

Inplane is along the axis of the linac (LNG)

Crossplane is perpendicular to the axis of the LINAC (LAT)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Why are you allowed during commissioning to assume perfect symmetry of your profiles?

A

Because you normally check ahead of time during acceptance testing

Afterwards, during all your QA, you ensure symmetry is close to perfect

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

What is the most ideal small field dosimeter for scanning? What about output factors?

A

Scanning: diode detector such as Edge or MicroDiamond

OF: W1 or W2

Note: Although W2 has scanning capabilities, it’s not too great. Software kinda stinks. Also scanning requires high doses, and this over time causes plastic yellowing throughout course of commissioning

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

What criteria is used to compare measured profiles to VRB data?

A

2% / 2mm gamma analysis for inplane, crossplane and PDD

3% / 3 mm gamma analysis for diagonals

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Where does the VRB data come from?

A

Real measured data which is a composite of multiple sets of data measured across multiple varian truebeams at Duke, with different physicists

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

What does VRB stand for?

A

Varian Reference Beam

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

How is VRB used during commissioning?

A

Two approaches based off the client

  1. VRB is used in the TPS to make models, and all commissioning measurements are used to validate the VRB
  2. The measured data is used in TPS to create the models and are double checked with VRB

In both cases, you need to ensure VRB data is close to measured data anyway. It’s then just a matter of which data do you want to use for your model? Which do you trust more?

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

How is data processed for PDD curves?

A
  • Shift
  • Smooth
  • Normalize
  • For electrons, software should have a depth dependent PDI to PDD correction for chamber
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

How is data processed for inplane and cross plan measurements?

A

Smooth
CAX correction
Make symmetric

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

The microdiamond and edge detectors are both diode detectors. What is the main compositional difference between them though?

A

Microdiamond is artifical diamong with a very small dye and is near tissue equivalent

Edge is silicon

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

Down to what field size is PDD data measured?

A

1 x 1 cm2

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

What field sizes are MLC and Jaw collimated small field profiles taken for in plane, crossplane and PDD curves?

A

1x1 –> 3x3 cm2 for PDD

2x2 and 3x3 for all profiles and PDDs

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

Down to what field size does eclipse accept in and crossplane data?

A

3x3 cm2

Anything below will not be accepted by Eclipse

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

Why for small field PDD can you feel confident measuring directly and not needing a per depth correction factor?

A

Although best practice is to do a per depth correction, it is deemed non-essential by commissioning standards, as the actual energy spectrum, and thus perturbation correction factor barely changes vs depth. Most commissioning don’t apply any factors for scans, only point measurements

This is due to field spreading increasing low energy scatter, while beam hardening increases energy, thus resulting in a somewhat balance. The factors will change, but not by a significant amount

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
For a output factor corrected reading at an intermediate field, how do you find output factor for fields smaller than the intermediate field using daisy chaining?
Measure response at the intermediate field This gives you a dose/response Then field response at smaller fields Multiply responses by correction factor to go from small field to intermediate field and multiply by dose/response of intermediate field
26
How are small field scatter factors normalized to a 10x10 field?
Two steps, First the corrected scatter factor is normalized to intermediate field, then normalized to 10x10 That is, daisy chaining
27
True or False Varian machines tend to behave very closely to their pre-calculated VRB?
True!
28
True or False Elekta machines tend to act very closely to their representative data?
False For vendors besides Varian, there is no representative data. All models must be based off user collected data.
29
Assuming you use VRB data to create TPS model, how do you perform validation during commissioning?
You can do it as you take the real data (this is most efficient) Generate plans in liquid water with the exact same setup as the measurements you're taking, then you can do either a point by point or gamma analysis comparison between the two In addition you will need to generate series of test plans for delivery of course
30
Assuming you didn't want to use VRB data for your TPS models, how does validation work?
You cannot validate data as you collect it. Meaning you must collect everything first, enter into TPS, generate models, then create plans to compare to the measured profiles
31
What critieria is needed for two linacs to be "matched"
Energy is exactly equal MLCs are the same DLG and transmission factors are the same **Note: **There is a tuning factor that allows you to have the same DLG. But that doesn't exist for transmission facotrs
32
How is transmission factor measured?
100 SAD, 5 cm depth 10x10 cm2 measurement first Take a measurement with MLC bank A covering field Then another with MLC bank B covering Transmission factor is average of A and B covering divided by open field
33
For small fields, why should you NOT measure PDD? Only TMR
It's very difficult to measure PDD due to chamber movement from CAX as the chamber changes depth. Your alignment is critical and has to be centered perfectly the entire range of depths For TMR, since the chamber only stays in one spot, it only needs to be aligned once If you are gonna measure PDD, you better make sure your chamber alignment is excellent at all depths
34
What data must be measured during commissioning for CDC/ECDC?
10 x 10 cm2 open field absolute dosimetry Output factor for all cones (95 SSD, 5 cm depth) TMR for at least three cones OAR for at least three cones measured at ATLEAST one SSD and one depth
35
What data did we collect for commissioning of CDC and ECDC?
Profiles at 80, 90, 100 SSD for every cone at d = 5 cm TMR at STD = 100 for all cones Output factor for each cone at 95 SSD, 5 cm depth for all cones, 5x5 cm2 jaws
36
Which detectors did we use for cone commissioning?
Edge and W1 They were compared to one another, and W1 was compared to MC
37
In addition to data collection for TPS, what else is done for Cone Commissioning?
* E2E test * Comparison of data vs golden beam provided by varian for 5x5 cm2 jaw setting * Winston-Lutz isocentricity test * Center check output factor * Cone Jaw setting check * Cone concentricity
38
When commissioning your 3D tank for scans, what are 4 things you must measure/determine extent of?
Alignment Orthogonality Distance Accuracy Hysteresis effects
39
For small fields, where you want to measure TMR but can't, how do you convert from PDD to TMR?
British journal of radiology 25 is a supplement that explains the conversion
40
True or False Sc and Sp data are never actually used by the TPS
True You do not input this data into the TPS. It is purely for your own databook in case someone wants to do a hand calc They contribute nothing to the models The output factor tables already include all scatter info and the TPS can just derive a model from that
41
What are some tests used to commission RapidArc?
Plans with slits at various gantry angles to test affect of gravity on MLCs PF at cardinal angles PF during rapid arc delivery (gantry rotation affect vs MLC positioning) Intentional PF error during rapid arc and ability of system to detect Dose rate and gantry speed control (deliver same dose with 7 different dose rates and gantry speeds) DMLC dosimetry with MLC slit at 4 different angles
42
Briefly, how is portal dosimetry commissioned?
PVA calibration all scans During PVA calibration MV dosimetry mode, deliver the beam, normalize response across all sub imagers in panel array, then upload the beam data for your TPS model and specify dose at CAX This will tell the panel exactly what dose was delivered to each point and it gives it a reference to calibrate all it's sub arrays to Repeat for all energies Run series of test plans (3D, IMRT, chair) to validate
43
What does the chair sample plan test? (3 major components of IMRT)
1. Inverse planning module 2. Leaf motion calculator 3. Dynamic MLC control
44
How does your TPS create its model given input dataset?
It's essentially an optimization process The model tweaks its parameters and optimizes to to try best match input data
45
What is the photon source size for the linac?
It is not one singular value, source size is defined by the algorithm, NOT physical size, and depends on if you collimate MLC or Jaws For AAA: x source size is 1 mm and y is 0 for MLC. Both are 0 for jaw For acuros: x source size is 1.5 mm and y is 0. Both are 0 for jaw Both models just need to use those values for their calcs, taking into account MLC or jaw transmissions in either direction
46
Per TG-106, where should physical wedge factors be measured?
100 SSD at dmax
47
Per TG 106 where should scatter factors be measured?
100 SSD either at dmax or 10 cm depth
48
Per TG-106, where are cone output factors defined?
100 SSD, dmax
49
What do the three regions of the chair test, test?
1. Investigate an area of only leaf transmission 2. Provide a homogenous area for absolute dose verification 3. Area of zero fluence between legs which require leaves to move over this area at max speed to minimize dose in the region. When the banks move over the region, they need minimal leaf gap. Errors that occur when modeling leaf gap will be evident by a mismatch in the zero fluence region between calculated and measured
50
True or False Per TG-106, it is actually recommended NOT to use VRB data for your TPS models?
TRUE Although a lot of clinics and commissioning teams do it anyway.
51
Per TG-106, what is the bare minimum data to be measured for photon beam commissioning?
* PDD curves * In-plane and cross-plane profiles * Wedge factors * MLC data (leakage, penumbra, tongue and groove, DLG) * Scatter factors * Tray factors
52
Per TG-106, what is the bare minimum data to be collected during electron commissioning?
* PDD * In-plane and cross-plane profiles * Cone factors * Insert factors * VSD
53
Typically, how many beam models would you have for a given photon energy?
5 or more One for open field, the rest are for physical and/or dynamic wedges These are all individual models
54
What is the minimum dimension for a water tank to be considered "full-scatter"?
5 cm margin around the largest field size dimension projected at all depths in the tank Down toa depth of 40 cm, a conservative size tank would be about 75x75 cm2
55
When may you see detector arrays used for commissioning?
Wedge profile measurements They are really good at measuring these
56
When scanning, what must the chamber direction of motion be level with, the tank or the water?
ALWAYS THE WATER Remember, **WATER IS ALWAYS LEVEL**. A tank is not As long as the chamber runs parallel with water surface, you're leveled
57
What is wrong with the setup that acquired the dotted data?
The gantry is tilted
58
What is an idea detector for scanning of electron beams?
Diodes! Stopping power ratio is unaffected by depth and energy of beam, meaning easy PDI to PDD conversion BUT, they're not great in the bremsstrahlung tail
59
Why are diodes good for scanning electrons, bad at measuring in the bremmstrahlung region of the curve?
Because bremmstrahlung region is all photons. Diodes made to scan electrons are not good at scanning photons
60
For scanning, is it best to scan slow or fast? Why?
Slow Less noise, and less water movement effects
61
Should you measure PDD shallow to deep or deep to shallow?
Deep to shallow Less water ripple effect
62
Regarding water in the 3D tank, what are some considerations you may want to take into account?
Use distilled water (avoid algae getting in mechanisms) Check water level atleast once daily for evaporation
63
Per MPPG 5a, what is the only recommended data collection for your CT sim in the TPS?
HU mapping data with a wide range of material densities
64
Down to atleast what field size should small field output factors be measured?
2x2 cm2 or below
65
Why does TPS not accept small field data below 2x2 cm2?
It understands there is a lot of uncertainty on the data collection It's better at that point to just extrapolate the data to get the really small field stuff
66
Before validating measured data, what are three useful practices you can do to assess the quality of your data?
1. Compare to reference data 2. Compare to expected physics (Ex. PDD should mimic what's roughly expected) 3. After TPS processes, make sure there was no issues with smoothing, mirroring or importing
67
What two types of validation datasets should you do? What are the uses of both?
1. Validation data calculated with plans exactly mimicing the collection conditions of data put into model (this makes sure the optimization didn't stray too far from the input data) 2. Tests with different setups than those used for collection (identify model's capability to calculate a dose for a scenario that was not used for the model)
68
What is tolerance per MPPG 5a for point dose comparisons of sample plans vs collected data used in model?
2%
69
Which two reports give a full list of recommended tests to do for commissioning validation?
MPPG5a and TG 106
70
Per MPPG 5a, what level of agreement should dose per MU at reference condition have with TPS model post commissioning?
0.5%
71
In what regions of curves/profiles do you expect the most inherent uncertainty/difficulty of the TPS model to predict dose in?
Buildup region Profiles with oblique incidence Penumbra Tails
72
Before conducting validation tests of heterogenity correction algorithms, what two pre-requisites are required?
Beam models have already been created an validated HU curves have been created from the CT Sim
73
What report gives a full list of recommended sample plans for IMRT/VMAT validation?
TG 119
74
**Advice: ** In general, know that you don't need to memorize all the recommended tests and tolerances in every report. Just know where to find the info. What are the three major reports you should know?
TG 106 and MPPG5a for commissioning and validation TG 119 for IMRT/VMAT validation All three of them have very detailed explanations of each test and recommended tolerances
75
In general for point dose validation, which regions have the most generous tolerances?
Regions that are difficult to model, - Penumbra - Tails - Buildup region
76
What tests does TG 119 recommend for IMRT commissioning?
A catalogue of sample plans ranging from minimal complexity to most complex. The report gives detailed descriptions on how to optimize these plans The sample plans are... - AP/PA - Bands - Multi-target - Mock prostate - Mock H&N - Complex C-Shape - Complex C-Shape 2
77
For the TG 119 recommended sample tests, what is the recommended phantom to be used?
A large water equivalent slab where a chamber smaller than a farmer chamber, and film, can be inserted for analysis Film is analyzed with gamma analysis Chamber is point dose %diff
78
Per MPPG 5a, what are the recommended TPS tests meant to check (two purposes)? How are they done?
1. Check that TPS has not been unintentionally modified (use checksums or uncalc and recalc standard beams. We do both monthly) 2. Dose calculation is consistent following TPS upgrades (uncalc and recalc standard beams)
79
Which TG report gives recommendations for commissioning of MLC, and what are the recommendations?
TG-50 Transmission test PDD, output factors and TMRs for MLC-generated fields spot check against jaw collimated fields In-plane and cross-plane profiles of MLC collimated fields spot checked against jaw collimated fields
80
What is the setup when measuring wedge and scatter factors?
It's completely vendor dependent. Whatever they want for their model Ex. Varian wants wedge profiles at dmax and wedge factors at d10 SAD setup
81
What setup does the TPS want for measurement and modeling of EDW factors/profiles?
Trick question, you never feed the model EDW. It's calculated by the TPS after the model is already made All you do for EDW factors and profiles is validate against a sample plan of EDWs that you generate on the TPS. There's no data to be taken for the model itself
82
What setup does Varian want scatter factors measured at for the model?
Sc and Sp do not go into the model, they are purely for the data book Scp is measured at 10 cm depth, 100 SAD. It's effectively just an output factor
83
What is the primary and secondary objectives of a commissioning job. What are the two parts that you are commissioning.
Two parts: 1. Which machine are you commissioning? 2. Which TPS are you commissioning? Two goals of commissioning 1. Collect data for the TPS (this is required) 2. Collect extra data for data books in case hand calcs are needed (some jobs do this, but this is absolutely NOT REQUIRED)
84
How is VSD calculated for electrons?
Take scans at 3 different distances from source Use software to find FWHM at all 3 depths Allows you to plot out the beam divergence Backward project divergence until you get to a point
85
Why for electron commissioning do we take uncollimated air scans at the defaulted field sizes per energy, even though we will never treat without the cones?
eMC needs it to help model scatter off jaws
86
What data does eMC need for modeling?
Un-collimated air scans Uncollimated 40x40 water scans Collimated inplane, crossplane and PDDs for all cones and all energies
87
In what tab in aria are DLG and transmission factors input into?
RT Admin, MLC modeling tab But with Eclipse 16, u can also add it in beam config
88
What is a "test box"?
Play area essentially where everything is configured for a version of TPS that isn't FDA approved or clinically used yet Allows for commissioning of models in a safe area that won't effect any current models
89
True or False, For varian commissioning jobs, they don't need to calculate models on the spot, they just need to import the models from a library
True They have all the models pre-calc'd ahead of time. This helps because some wedge and electron models take a VERY long time to calc.
90
When calculating a model, what is the TPS actually doing?
It takes your input data and tweaks your beam parameters in such a way that it gives you an optimized model to match your data
91
What are all algorithms that need to be modeled during an eclipse commissioning? (6)
PO - photon optimizer PDIP - portal dosimetry AAA Acuros eMC ECDC if you have cones
92
In general, what is your measured data validated against in a commissioning?
VRB data AAA calc'd plans Acuros calc'd plans eMC calc'd plans PDIP calc'd plans with multiple test plans These are all either point does (like absolute point doses, OARs, output factors, scatter factors, etc), and/or scan validations (3%/3mm gamma, such as in-plane cross plane, PDDs, wedge profiles, etc)
93
True or False Each scanning and analysis software (IBA, SNC, Mephysto) have a different file format that you need to convert between in order to share the data with customers that don't have the same syste as you.
True
94
What is the difference between gold beam data and VRB? Which does varian use during commissioning?
Gold beam data was notorious for having some issues and processing inconsistencies VRB is an updated version of gold beam data that was acquired from 3 truebeams in Duke Varian during commissioning used VRB for everything other than small fields. This is because duke measured small field scans with a cc13. THe varian commissioning team has improved data for small fields to replace the VRB data
95
What impact does tweaking the target spot size in beam config have on your model?
It adjusts the geometric penumbra of your beam model In addition, it also have a very small impact on the hotspot
96
What are some of the values that model tweaks during optimization? Where are these characteristics held in the model, what is that all encompassing data called?
Energy spectrum Mean energy Head scatter and leakage Energy intensity vs radial Electron contamination Phase space is characterized by these
97
Most of the time the default field size settings for cones with electron energy remain what varian sets them as. When would this not be the case and what must be tweaked in TPS if this occurs?
Sometimes the engineers during installation cannot get beam steering into spec, so they may have to change default field sizes to do this If this happens, you need to make sure they update the default value list in RT Admin
98
**An important note on one of the purposes of commissioning/validation. Read back**
You may see during validation that model calc'd profiles or point doses may not perfectly agree with measured. At times they may even be out of tolerance Part of commissioning and validation is to know where your uncertainties are Ex. TPS is not great at modeling dose beyond the boundaries of your wedge, So you often see gamma failure outside of hard wedge fields. It's good to know this limitation but to also know outside wedge is often low dose and of minimal clinical impact
99
How is absolute dose defined in the TPS?
In beam config,each model has a subsection for absolute dosimetry parameters You need to define a single geometry and associated dose/MU there It can, in theory, be any setup as long as you are consistent in your calibration and QA Ex. Varian actually recommends 95 SSD, 5 cm as the point of dose calibration, but your dose/MU there is not 1 cGy/MU, that is dependent on whether you're SSD or SAD calibrated
100
Conceptually, what is DLG?
Parameter in RT Admin used to model the rounded leaf effect of the leaf collimator (MLC). This parameter attempts to model the physical difference between the radiation and light field and accounts for inherent leakage between adjoining leaves. It has units of cm and is the FWHM of the field created by leaves touching each other
101
Why does varian recommend setting absolute output in parameters at 95,5 or 90,10 instead of dmax like we calibrate in our clinic?
Because TPS gives the best absolute output match at whatever your reference condition is **in the beam config**. Since targets are almost never at dmax, and are usually at depth instead, you want the best matching at depth
102
True or False During the model calculation, the TPS can change parameters set by you.
False These are all hard set and include reference conditions for absolute dose calibration, effective target spot size, SAD distance, etc **NOTE: ** Optimization does NOT tweak the effective target spot size. You set that based on varian recommendation that best models penumbra
103
What are some things that your model calculates during optimization based on input data?
Absolute dose scaling factor Mean radial energy Intensity profile Particle fluence spectrum Head leakage/scatter Electron contamination Calculated profiles Calculated depth dose Calculated gamma error histogram Calculated wedge parameters (also dose scaling)
104
Why does eclipse use DLG but no other TPS does?
Eclipse does not model MLC leaf ends, so instead they created DLG which is a factor that acts as essentially a fudge factor that accounts for leakage of the leaf ends
105
For validation point doses, what sorts of setups, in general, tend to yield the biggest expected %diff?
Those where you change multiple parameters from reference Example, non-10x10 field + off axis + deep depth + wedged field, etc The more changes you have to your reference condition, the worse you expect uncertainty to be
106
For Sc measurements, is it better to use acrylic or brass buildup caps? Why?
Brass You get your buildup with a smaller volume of material, meaning the cap+detector can be fully in the field down to smaller fields and up to higher energies relative to acrylic
107
What do the marked 3 regions of the below test allow the user to inspect?
Region 1: Has an aea of completely blocked dose. This lets user inspect accuracy of modeled transmission factor Region 2: Is completely open, allowing the user to inspect accuracy of homogenous open field absolute dose verification Region 3: Has a center zero region in which the leaves must move very quickly from left to right while maintaining a zero gap in order to minimize dose in the region. Inspects travel speed and leaf gap modeling
108
Why does TG-106 recommend use of daisy chaining if diodes are used for SF output factors?
To minimize energy dependence effects
109
per MPPG 5B validation with point sources, what are the recommended tolerances for difference between calc and measured for... High dose region with 1 parameter change from reference - High dose region with multiple parameters changed from reference - Penumbra distance to agreement - Low dose tail region up to 5 cm away -
High dose region with 1 parameter change from reference - 2% High dose region with multiple parameters changed from reference - 5% Penumbra distance to agreement - 2 mm Low dose tail region up to 5 cm away - 3% of field max dose
110
per MPPG 5a, what is the tolerance for dose per MU at reference condition for TPS vs measured?
0.5%
111
Per MPPG 5a, what is the heterogeneity point dose tolerance for a setup of heterogeneity above and below measurement point?
3%
112
True or False Dosimetry for SFD is often extrapolated by the TPS?
True Most TPS do not take field size data at 2x2 or below. Instead they extrapolate because the process of measuring that data presents too much uncertainty So during commissioning, when we take SFD data, it's purely for validation. Ex. MLC collimated small field PDDs and small field profiles and PDDs and output factors
113
In addition to dose comparison for heterogeneity validation, what else does MPPG 5B recommend for validation of heterogeneity calculation?
Confirmation of lookup tables, bulk densities , HU to electron density and mass density curves, and material assignment tables for the TPS
114
Per IIROC, what percentage of facilities pass the anthropromorphic phantom E2E with very loose gamma criteria? What about slightly more strict but still pretty loose criteria?
Very loose - 90% of facilities pass Less loose but still loose - 70% f facilities pass
115
Per MPPG 5B, what four types of tests are recommended for validation of electron model?
116
What fraction of commissionings actually do a full MPPG 5B validation?
Around 1/4th
117
the DLG is a value defined in cm. How is the value used by the TPS?
When calculating fluence from the beam, the TPS will automatically bring back the leaf ends by half of the dosimetric leaf gap. This is because the TPS does not actually model transmission through leaf, as a result they have to bring the leaf back virtually to account for the unaccounted for leakage fluence From beam config guide: "shifting the leaf tip positions in the actual fluence calculation. Leaf tips are shifted by pulling each of them back by half the value of the dosimetric leaf gap parameter so that the gap between a fully closed leaf pair equals the dosimetric leaf gap parameter."
118
In Varian TPS model, what is the "second source?" What does it help model?
The second source is a Gaussian plane source located at the bottom plane of the flattening filter. It models the photons that result from interactions in the accelerator head outside the target (primarily in the flattening filter primary collimators and secondary jaws).
119
What three "sources" do both AAA and Acuros model?
Primary source Secondary source Electron contamination source All three sources and their characteristics are contained in the phase space created by the model
120
True or False To create the model for PDC, you don't have to take additional data?
True The Portal Dosimetry pre-configuration package consists of: ■ Pre-configured beam data for the “Portal Dose Image Prediction” calculation model. ■ A two dimensional beam profile correction file to be imported during “Dosimetry Calibration” on the 4DITC workstation. ■ A set of verification plans ensuring the proper installation of the package.
121
What 5 sources make up the electron source in eMC model?
■ Main source (electrons and photons) as a point source near the scattering foil ■ Jaws source (electrons and photons) ■ Scraper sources for electrons scattered at the upper applicator scrapers ■ Edge source for electrons scattered at the last applicator scraper or insert ■ Transmission photons