test 9 third generation Flashcards
Heartware HVAD
Small continuous flow rotary pump with a centrifugal and non-contact bearing design.
Impellar is held in place by passive magnetic and hydrodynamic bearing systems, avoiding mechanical contact and wear.
Physical contact between the housing and the impeller is prevented by thin blood film generated by the hydrodynamic bearings.
Placed within the pericardial cavity at the apex of the LV
No need for abdominal pocket
first human implanted Heartware HVAD
2006
Heartware HVAD approved BTT in
Approved November 20, 2012
Flow through the heartware HVAD
- Three blood paths
- Primary blood flow
- Secondary blood flow
- Tertiary blood flow - enhances blood flow and reduces blood trauma while reducing the time blood travels through the device
DuraHeart LVAS
By Terumo Heart, Inc.
Continuous-flow rotary pump
Upper housing with levitation system and impeller
Bottom housing containing the external-drive motor
Flow rates of 2-8lpm with motor speeds of 1200-2400rpm.
First device to reach clinical trial that combined centrifugal pump and magnetic levitation technology
Placed in abdominal pocket
Clincally available in Europe.
Momentum 3 Trial Heartmate 3
US IDE Clinical trial
Prospective, Randomized, Multi center, unblinded
Heartmate III vs. Heartmate II
DT/ BTT
Heartmate III
Fully magnetically levitated impellar No friction Minimize complications Artificial pulse technology Help with clot formation Textured blood surface Prevent clot formation More like native tissue Pocket controller Implanted above diaphragm Up to 10 Liters/min