Mode Switch Flashcards

1
Q

How does blanked flutter search work?

A

Monitors A-A intervals that may indicate blanked flutter waves. If there are 8 consecutive intervals that are < 2 x TAB and the atrial rate is > 1/2 the mode switch intervention rate, then PVARP is extended. If a P wave is unmasked, the device immediately mode switches. If the pattern persists, this process is repeated every 90 seconds.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

What criteria must be met in order for a pacemaker to mode switch?

A

4 out of 7 intervals must be fast than the programmed MSw Intervention rate.