Mixed Review Deck 7 Flashcards
Oblique L spine: correct obliquity is demonstrated by the pedicles (eye of Scottie dog) demonstrated where?
Halfway between the midpoint of the vertebral bodies and the lateral border of the vertebral bodies.
On an oblique L spine, the pedicles seen _____ (in relation to the vertebral bodies) indicates over-rotation? Under-rotation?
Pedicles seen closer to the midline of the vertebral body indicates over rotation. If they are closer to the lateral border of the vertebral body, this indicates under-rotation.
In evaluating lateral knee radiographs, the distal margin of the medial condyle is ___ in shape, while the lateral condyle margin is ___.
Medial is convex, lateral is flat.
Lateral knee: if the medial condyle is seen the joint space (its distal margin seen distal to the distal margin of the lateral condyle), what is the correction?
Cephalic CR angulation was insufficient.
Give the voltage wave (VW) correction factors (or generator correction factors) for calculating heat units
single phase: 1
for 3-phase or high frequency: 1.4
Point lesions (damage to a structural molecule, such as DNA) occur when there is disruption in …
single chemical bonds
What is the most common cause of tube failure?
electrical arcing
What is electrical arcing?
When, during the final stage of an exposure, the filament electrons jump to the glass of the envelope, possibly causing a crack in the glass.
During processing of the PSP image, what layer of the IP is responsible for directing the (isotropically emitted) blue light toward the processor’s light collecting optics and photodetector?
the IP’s reflective layer
The continued emission of light by a phosphor after the activating source has ceased is termed …
phosphorescence
Medical and dental radiation accounts for what percentage of the general public’s exposure to human-made radiation?
90%
Congruence of the x-ray beam with the light field is tested using …
radiopaque objects (placed at each corner of the light field before an exposure is made)
Why should 70 kV NOT be exceeded during an IVU?
Because the iodine-based contrast given during this exam gives optimal opacification at 60-70 kVp
What is the BEST way to reduce magnification distortion (and why?)
Decrease OID. This is better than the other option (increasing SID), because an increase in SID would require an increase in exposure factors as well.
What quality control test of a CR reader is done using a small lead block?
erasure thoroughness test (for complete erasure) done by exposing the lead block, then erasing, then re-exposing the plate without the block
Quality assurance deals with ____ while quality control deals with ___
QA: people
QC: equipment
What does TQM refer to?
Total Quality Management (another term for quality control)
How often is filtration tested, and what does this measure?
Tested annually; it measures beam quality by doing a half value layer measurement.
What is used to do a quality control test of filtration/beam quality?
digital dosimeter
How often is collimation accuracy (light field/beam congruence) tested, and what is the tolerance?
tested semi-annually; tolerance within 2% of SID.
Remember it was always wrong at Riverbend so it definitely should be semi annual!!!
How often is the effective focal spot tested, and what is the tolerance?
annually; within 50% of equipment specs
What are the 3 tests of effective focal spot size, and which is most accurate?
slit camera, star test pattern, and pin-hole camera. slit camera is most accurate
kVp and exposure linearity are both tested ___ (how often?) and have a tolerance level of ___
annually; 10%
How often is the timer tested for accuracy? It must be accurate within ___ % of exposures over ___ ms
annually; within 5% of exposures over 10 ms
How often are lead aprons tested for cracks, and how are they tested?
annually they are tested using fluoroscopy equipment
The spinning top test and synchronous spinning top are used to test what 2 things?
timer accuracy AND rectifier function
What type of equipment does the synchronous spinning top test vs. the spinning top teset?
synchronous spinning top tests 3-phase equipment and will display arcs rather than dots; the spinning top test tests single phase equipment and displays a series of dots.
Synchronous spinning top test: A 3-phase full-wave rectified current would expose a 360 degree arc each ___; a 1/12 second exposure should expose a ___ degree arc.
second
1/12 sec exposure: 30 degree arc
For spinning top tests and synchronous spinning top tests: if the test is off by a few dots or arcs, the issue is probably with the ____. If the variance is exactly half the expected number of dots or arcs, the problem is likely to be _____.
a few dots or arcs: timer accuracy
exactly half: rectifier failure
What tools are used to test exposure linearity and reproducibility?
step wedge and densitometer
Fluoroscopy equipment exposure rate is tested with a ____ and exposure should not exceed _____
digital dosimeter
100 mGy(air)/minute
Fluoroscopic exposure rate must not exceed …
100 mGy(air)/minute
Digital display monitors must be calibrated to the ….
DICOM Grayscale Standard Display Function (GSDF)
DICOM PS3.14 is a standard that controls the …
lookup table of the monitor display
Widening of intercostal spaces is characteristic of…
Emphysema
Pneumothorax is typically accompanied by…
Atelectasis
An increase of 1.0 mm added aluminum filtration would have what effect on the tube’s mGy(a) output?
Decreased mGy(a), because the filtration would absorb low energy photons, decreasing beam quantity.
To superimpose the femoral condyles on a mediolateral knee, the CR angle for each of these patients should be …
-average pt
-short pt with wide hips
-tall pt with narrow pelvis
Avg: 5 to 7 cephalad
Short/wide: 7 to 10 ceph
Tall/narrow: 5 degrees ceph