Mersey XR Flashcards

1
Q

X RAY PRODUCTION

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

BREAKING RADIATION

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

MAX ENERGY = TUBE POTENTIAL

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Characteristics XRAYS for tungsten

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q
A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

X-Ray Tube

A LOT OF HEAT is produced
Thermal conduction dissipates heat
- down anode shaft
- surrounding mineral oil for thermal INSULATION
-Surrounding lead shield absorbs x rays

A

Inherent filtration is 0.5-1mm Al

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

There will be SOME leakage radiation

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

RADIATION PROTECTION

A

Berilium - lower atomic number - used as a window for lower energy mamography

Normal XR - Aluminimum
CT - Copper and Tin
Mammography - Molybdenum or Rhodium Filters

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

RAYLEIGH SCATTERING

NO ENERGY TRANSFER
ONLY OCCURS AT LOW ERGY IN SOFT TISSUE
<10% OF XRAY INTERACTIONS

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Photoelectric effect - provides contrast

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

1kEV soft tissue characteristic soft tissue x ray

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Compton Scattering

A

Small scattering angle - x ray still has most of its energy
Larger scattering angle - x ray loses or transfers MORE energy

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Graph showing P() of angles at different energy
At LOW energy = scattering through larger angles and smaller angles

You get LESS backscatter at HIGHER energies

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

P() of Compton scattering LOWERS with higher energy
INCREASES with MASS DENSITY (ρ) and number of electrons per gram

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

X-Ray interactions in tissues

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

In soft tissue >25KeV - Compton Scatter predominates

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Attenuation

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

Linear Attenuation coefficient
- increases with density
- decreases with energy

LAC/DENSITY - MASS ATTENUATION COEFFICICIENT

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

Comparing air to soft tissue and bone

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

K edge filters - used to remove high energy x rays to IMPROE CONTRAST

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

K edge filters preferential attenation

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

K EDGE OF IODINE = 33.2kEV
So you want x-ray energies ABOVE this

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

So here’s a typical setup that you would use for plane or X ray imaging. You’ve got the X ray tube, obviously your patient supported on imaging table. Underneath the imaging table is your x ray image receptor, which we’ll look at. There’s different types an x ray image receptor. In between the table and the image receptor, there are two, two other objects that are often used. So there’s ionization chambers which are which are used to automatically control the exposure of the X ray. So the the AEC, as it’s often referred to, there’s that these ionization chambers are interspersed between the detector and the patient, and in some studies. There will also be an anti scatter grid placed between the patients and the detector

A

99% of the energy that they have gets converting converted into heat, and only about 1% gets converted into X rays. - thats why it rotates

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

the X rays are emitted from the target, from an area known as the focal spot. So the focal spot is where the electrons from the cathode interact with the target So the focal spo

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

SMALLER ANGLE = SMALLER EFFECTIVE FOCAL SPOT SIZE
Larger real spot = more heat dissipation

The target angle affects the effective focal spot size, leading to a smaller focal spot size with a larger heel effect.

A
25
Q

More self attenuation in direct of ANODE

Tungsten has got a high atomic number and a high density, which makes it a good absorber of X rays, a good producer of X ray.

BUT MORE HEEL EFFECT at SMALLER TARGET angles
Heel effect MORE APPARENT at LOWER energies due to INCREASED PHOTOELECTRIC EFFECT.

X rays emitted towards cathode have less self attenuation, resulting in non-uniform intensity along axis.
Heel effect is more noticeable at small target angles, exaggerating non-uniformity at low tube voltages.

A
26
Q

Spatial Resolution

Measurable ways to say it’s the ability to resolve objects as separate entities.

How far apart do you need to have two objects for them to be seen as separate in the image. So that’s that’s a distance that you can measure.

The size of the penumbra determines the space resolution in plain radiography, with smaller penumbras resulting in higher resolution.

A

This is what the test patterns typically look like, a series of attenuating and non attenuating rectangles of different thicknesses and separations. So with each of these, we have what’s known as a line pair. We have an non attenuating and attenuating rectangle of the same width and the same separation, and then you go through the next pair same with separation and so o

So in spatial resolution, in X ray imaging is typically referred to in this in this way. Line pairs per millimeter.

So typically you will get planar X ray imaging with resolutions of five line pairs per millimeter, that sort of magnitude.

27
Q

BROAD = 1.2mm
FINE = 0.6mm

Focal spot size is actually selected by choosing a different coil. in the cathode.

Focusing also helps to direct electrons into a narrower beam as they have a tendancy to divge

The size of the penumbra determines the space resolution in plain radiography, with smaller penumbras resulting in higher resolution.
The focal spot size and source-to-image receptor distance are important geometric factors that affect the size of the penumbra.

A
28
Q

X RAY IMAGE RECEPTORS

Instead, the three sets of detectors that we’re interested in for planar X ray imaging are

A
29
Q

COMPUTED RADIOGRAPHY

Imaging plates with phosphor in a casette

Expose the imaging plates with X rays.
Electrons within the phosphor within this barrier of fluoro halide, electrons are excited to higher energy states and the trapped there. These electrons jump up in energy to a different energy states, and stay the, they don’t, they don’t drop back.

A

The READ OUT
Red light stimulation
Blue light released and detected by photomultiplier, amplifier and ADC

Plate erased by bright white causing all remaining -e’s to drop back

30
Q

ADC conversion

Continuous analogue signal from photomultiplier tube. Each signal stored as a digital value in each pixel.

Sampling pitch = pixel size
Sampling frequency = pixels per unit length = 1 / pixel size

A
31
Q
  1. CR Plates

Matrix size = no. pixels in both axes

Cesium iodide and gadolinium sulfide are common scintillators used in X-ray detectors.

A

Smaller imaging plate with same matrix = smaller pixel = better spatial resolution

32
Q

Spatial resolution:

CR imaging resolution depends on sampling frequency, pixel size, and laser beam width.

so space resolution in CR imaging depends
- the sampling frequency, the pixel size during the read out process.
- by the size of the laser beam that’s used in the reader. if you have a very broad laser beam,
- the scattering of that laser light within the phosphor. So the thicker the phosphor is on the plate, more scattering of laser light will occur, and then the divergence of the light that’s emitted from the plate.

A
32
Q

So digital radiography using flat panel detectors is another type of imaging detector that we use in planar X ray work.

Flat panel detector is a matrix of individual detectors. Cr play is just one continuous detector that’s digitized.

Digital radiography consists of a matrix of detectors, and you detect a signal that each individual detector to produce your image.

A

ith CR plates. You normally just stick they go under the imaging table in a holder, these different sized imaging plates. You take them away to read them out with flat panel detectors. They’re usually part of the X ray system. So it’s an example of one sort of fixed in place, or it can be fixed in place underneath. When you irradiate them, you don’t need to read out. You don’t put them in a reader. Afterwards, they produce a digital signal, almost immediately connected to a computer and produce a digital signal. There’s two types of the flat panel detector, indirect, direct. So look at each of those

33
Q

INDIRECT FLAT PANEL DETECTORS

XRAYS to LIGHT to ELECTRICAL

Indirect flat panel detectors are better for general radiography and fluoroscopy due to lower patient doses and higher resolution images.

Different phosphors, different scintillators that you can use. Cesium iodide is, is a common one, gadolinium, oxysulfide is another one.

A continuous piece of scintillator, and it sits on top of matrix of photo diodes. So a photodiode is just a light detector
light that’s emitted from the scintillator will be detected by these filter diodes.

A matrix of light detectors detecting light from the scintillator.

The laser light scattered within phosphor spreads out, causing blurring of signal.

A
34
Q
A
35
Q

DIRECT FPDS

Direct conversion of XRs into electric current

The direct flat panel detector converts X rays into electric charge directly, without emitting light, and collects the electric charge through a high electric field on a transistor array.

Direct flat panel detectors have a lower divergence of signal compared to indirect flat panel detectors, resulting in higher resolution available.

AMORPHOUS SELENIUM - PHOTONCONDUCTOR

PIXEL SIZE = TRANSISTER SIZE

A
36
Q

Comparing the two

Indirect = better with higher energies = lower patient dose

Direct = better with lower energies = better in mammography

A
37
Q

Image intensifier

X-Ray Source
Detector = Image intensifier = detects WEAK XRs

Flux and minification gains determine brightness gain, with accelerated electrons producing more light at output phosphor.

A
38
Q

So is side a schematic view side through of an image intensifier. X rays are entry through a very thin entrance window here, just on the inside of that there’s a phosphor or a scintillator, if you like, typically, caesium iodide that converts X rays into light. The light is then incident upon a photocathode, which converts the light into electrons. And the electrons that are liberated from here are accelerated by series of electrodes, they’re accelerated towards an output window. So these electrons, they’re generated accelerated by these electrodes.

A

And the electrons that are liberated from here are accelerated by series of electrodes. They’re accelerated towards an output window. So these electrons that are generated accelerated by these electrodes, you’ll notice that the output window, the output phosphor, is much smaller than the input phosphor, so we’ve got a sort of a concentration of energy going from a wide area to a small area, and that produces that helps to produce image brightness helps to amplify the brightness of the signal from here going down to here. So when these electrons reach this output phosphor, they convert their kinetic energy gets converted into light, so we get a visible image produced on a much smaller screen. that reflects what’s happening on the input screen.

39
Q

Gives BRIGHTNESS GAIN - brightness between two screens

FLUX GAIN = acceleration by electrodes

MINIFICATION GAIN = Output image is much smaller = square of input screen diameter / square of output screen diameter

BG = FG X MG

A
39
Q

INTENSIFIER ARTIFACTS

Pin cushion distorsion = Non-linear pathway of accelerated electrons through intensifier

Vignetting = image brighter at the centre

Veiling Glare = depicted here is a lead disc, which should be totally attenuating, so we shouldn’t get any signal. This is the profile of signal intensity through through the central line. We shouldn’t get any signal here, because this is totally attenuated led disc. But you’ll see that there’s a signal.

A
40
Q

Magnification in Fluoroscopy

Magnification REDUCES MINIFICATION GAIN = reducing BRIGHTNESS GAIN = COMPENSATING BY INCREASING XRAYS

A
41
Q

Digital subtraction angiography

MASK subtracted from pulse contrast or LIVE images

LOGARITHMIC SIBTRUCTION (due to exponential xray attenuation)

SUBTRACTION PROCESS = INCREASED NOISE

Uses >70kEVP as it uses iodine

A
42
Q

Digital subtraction angiography

A
43
Q

XRAY CONTRAST = VARIATION IN BEAM ATTENUATION

What we mean by contrast? First of all, contrast is the difference in sit and strength between two parts of an image. Obviously, if we’re going to detect features in an image, we need to have contrast, differences in signal between what we’re interested in and what we sort of our target features and background features.

Higher energies, there’s less attenuation,

More attenuation if high density or atomic number

A
44
Q

Radiographic contrast

Amount of scattered radiation that reaches the image receptor will interfere with the amount of contrast that we can see, so we may need to remove scattered radiation. A

The actual display information on the computer screen, the way it’s displayed can affect our perceived contra

A

Subject contrast

45
Q

Subject contrast

So we’ve got 6x rays in this schematic interacting with long 6x rays going through bone, six through soft tissue. You’ll notice that in each case, there’s a different number of X rays being transmitted through lung is a very weak attenuator. So most of the radiation pass through the soft tissue, there’s partial attenuation, and through bone, there’s much greater attenuation

A

Effect of Kevp

46
Q

Effect of scatter

A

About two thirds will be absorbed in the patient as radiation dose, and only about 1% actually gets the image receptor.

46
Q

Controlling Scatter

So if we can reduce the volume of tissue that’s irradiated, we will reduce the number of scattered photons being produced. So we do that by making the beam smaller. So we use beam coordination to make the beam area as small as possible. So that makes the area of the beam small, and we if we compress the tissue, we make the thickness of tissue that the radiation is passing through smaller as well. So these two techniques are simple ways of trying to reduce the amount of scatter that’s produced. So tissue compression is used principally in mammography. It’s used routine mammography, and that’s why to reduce the amount of scatter and enhance contrast.

A
47
Q

Scatter Vs kVP

Higher energy XR = FEWER SCATTERED EVENTS = BUT MORE FORWARD SCATTERS = SO MORE GET TO THE DTECTOR

Increasing KVP leads to fewer scatter events, but those that do occur are more likely to reach image receptor.

A
48
Q

Scatter Grid = series of attenuating strips

A

Lower grid ratio = less scatter
Grid ratio = H/d

49
Q

Radiographic noise

Unwanted variation in signal intensity - INCREASE NUMBER POF XRAYS

A
50
Q

Mammography

  1. Compression of breast to reduce thickness
  2. Xray tube angle cathode side along chest wall = where beam is stronger through thicker tissue
  3. Lower tube voltage 25-35KVP
  4. Mo Rh
A
51
Q

looking at radiation output from X ray tubes, the focal spot is effectively a point source of X rays, which means we can use the inverse square law to determine the intensity of radiation with distance from the fogged spot. So the beam radiation is emitted in all directions. So the radiation field diverges with distance from a point source. The Inverse Square Law says that the intensity of the radiation falls off with the square of the distance that you are from the source. So this makes distance a very good way of protecting yourself against radiation. Coronavirus in lecture, looking under the radiation output from an x ray machine

A
52
Q

Radiation output from XRAY tubes = Measured in AIR KERMA

Kerma = TOTAL ENERGY liberated by PE and CS events events per unit mass of air - mGy

Air Kerma Rate mGy/s = measured by IONISATION CHAMBER

A

Intensity Vs KvP - proportional to voltage SQUARED

52
Q

Minimum dose to receptor to get useful image

Low KVp - higher ESD, Low penetration
Higher kVP - lower ESD, higher penetration - BUT LESS CONTRAST

A
52
Q

Intensity Vs tube current - proportional to current

A

Entrance surface dose

53
Q

DOSE AREA PRODUCT

Flat ionising chamber measures dose x beam area to measure how much going to patient

DOSE NOT VARY WITH DISTANCE

DAP is independent of distance from X-ray tube, making it useful for measuring patient dose

A

PATIENT DOSE PARAMETERS

DAP or kap
Reference Air Kerma
Entrance Surface Dose
Organ Dose
Receptor Dose

54
Q
A