Mersey US Flashcards
- Pressure varies along an US wave
- Pressure is highest at a compression & lowest at a rarefaction
- Wavelength = distance between corresponding compressions
- Frequency = number of wavelengths emitted each second
- Amplitude = increase in pressure at a compression
Acoustic Impedance (Z)
- A measure of how a tissue resists the passage of the US wave
- Z is affected by the density & compressibility of tissue
- For any tissue: Z = density x wave speed
- Z is measured in units of rayl
- NB – Z is independent of the US frequency
- Bone and PZT crystals have high Z
- Air & lung have low Z
- Differences in Z between tissues are important for
creating reflections (echoes)
- Wavelength range 0.1-1.0mm in clinical images
- Audible range of the human ear is 20Hz - 20kHz
- US imaging uses a frequency range 2-15 MHz
- In US, frequency is inversely proportional to wavelength
➢ At 1.5MHz: wavelength = 1.0mm in soft tissue
➢ At 15MHz: wavelength = 0.1mm in soft tissue
Speed of Propagation (Wave Speed)
- Waves speed depends on the stiffness or
compressibility of the tissue or medium - Wave speed, c = frequency x wavelength
- US machines assume the waves travel at 1540m/s in the patient
- Speed of propagation varies for different tissues
➢ Bone 4000 ms-1
➢ Fat 1450 ms-1
➢ Lung 600 ms-1
Wave speed is independent of the US frequency
NEED TO KNOW FOR EXAM
Interactions of US with Tissue
There are 4 basic interactions of US radiation with tissue:
* Reflection – partial reflection can occur at boundaries of 2 dissimilar tissues.
* Scattering – structures smaller than the wavelength of the beam scatter US radiation in all directions
- Refraction – change in direction of US wave when striking a boundary of 2 tissues obliquely
- Absorption – US energy is converted into heat as it passes through tissue
Specular Reflection
- Specular reflection occurs at large, smooth boundaries between
tissues where there is a change in acoustic impedance - Incident wave is partially reflected
- The reflected ‘echo’ is weaker than the incident wave
- The incident wave is transmitted with reduced amplitude
- The transmitted wave can create further echoes
- When the incident beam strikes the boundary obliquely specular
reflection still occurs - The angle of reflection = angle of incidence
- The intensity of the reflected wave is still reduced
Scattering
- Structures < wavelength cause scattering
- Very weak echoes produced in all directions
- Wavelength depends on frequency, so scattering will also depend on frequency
- At 1.5MHz, scattering occurs from objects < 1mm
- At 15MHz, scattering occurs from objects < 0.1mm
Specular reflection – echo strength
- The intensity of the reflected wave (with respect to the incident wave) depends on the difference in acoustic impedance between the 2 tissues (Z1 & Z2)
- Relative echo strength is independent of US frequency
Diffuse Reflection
- Diffuse reflection occurs if the reflecting surface interface isrough or has undulations that are smaller than the wavelength of the radiation
- Multiple reflected waves are generated & emitted in different
directions
Refraction
- Refraction is the change in direction of a wave passing between 2 tissues where there is a change of wave speed
- It only occurs when the incident beam strikes the tissue boundary obliquely (there is no refraction for normal incidence)
- The transmitted wave moves in a different direction from the incident wave
- The magnitude of the change in direction depends on the ratio of the wave speeds in the 2 tissues
- The reflected wave is unaffected
US Beam Attenuation
- As ultrasound pulses (and reflected echoes) travel
through tissue, their intensity is reduced or
attenuated.
- Attenuation of the US beam is due to:
➢ Reflection and scattering removes/redirects energy
from the US beam
➢ Beam divergence – energy is spread over a wider area thus reducing intensity
➢ Frictional losses – results in absorption of US energy by tissue and is converted into heat.
US Beam Attenuation
* The intensity of any US beam decreases exponentially
with depth in tissue.
* Attenuation increases with US frequency
US Attenuation Coefficients
NEED TO KNOW
- US attenuation coefficients are quoted in units of dB/cm/MHz
- Soft tissue attenuation coefficient ~0.8dB/cm/MHz:
– 0.8 dB/cm at 1 MHz (17% reduction in intensity per cm)
– 4 dB/cm at 5MHz (60% reduction in intensity per cm)
- Water/blood attenuation coefficient
~0.15dB/cm/MHz:
– 0.15 dB/cm at 1 MHz (3% reduction in intensity per cm)
– 0.45 dB/cm at 5MHz (10% reduction in intensity per cm) - Bone attenuation coefficient ~15dB/cm/MHz:
– 15 dB/cm at 1 MHz (97% reduction in intensity per cm) - Lung attenuation coefficient >30dB/cm/MHz
Ultrasound Transducer
- An ultrasound transducer converts electrical energy into ultrasound waves (mechanical energy) & vice versa
- Transducer acts as both the source & detector of US waves using the Piezoelectric effect
- US production - alternating voltage applied to the end faces of the transducer crystal causes it to vibrate
- US detection – waves arriving at the crystal face cause it to vibrate & generate an alternating voltage
- PZT (lead zirconate titanate) often used as the transducer crystal
Ultrasound Transducer
- The crystal is backed with a damping material to prevent the crystal from vibrating
- This is important for the production of short pulses
- An “acoustic window” is fitted to the front surface of the transducer to protect the crystal from damage and transmit as much of the US radiation as possible into the patient
ACOUSTIC FIELDS
- US imaging is performed in the near field as far as the focal zone.
- The shape of the beam is affected by the frequency of the US radiation and
the diameter of the transducer crystal. - The near field is narrower at higher frequencies
Unfocused beam - parallel - then diverges
Focused converges and then diverges
Basis for US Imaging
- All US imaging derives spatial information about the location of tissue
boundaries using the “pulse echo” principle - A short pulse of US (typically ~1µ sec in duration) is transmitted into the body
& the time taken for it to return to the transducer is measured - If the velocity of the pulse in tissue is known (or assumed) then the depth of
the reflecting structure relative to the transducer can be determined
US Pulses
- A pulse is a short burst of US radiation
- Each pulse has finite duration and size (length)
- Length is typically 2-3 wavelengths
- Duration is typically ~1s
- The number of pulses emitted each second by the transducer is called the
pulse repetition frequency (PRF) - In clinical US imaging the PRF is typically 3-6kHz
US Scanning
* There are various ways data can be recorded
➢ A-mode
➢ B-mode
➢ M-mode
➢ CW Doppler
➢ Pulsed Doppler
➢ Harmonic imaging
A-Mode Scan
- In A-mode (“A” = “amplitude”) pulses are sent down a single scan line and
produces a graphical display, not an image - Depth is represented by the horizontal axis of the display
- The amplitude of each echo is represented on the vertical axis
- A-scans can accurately depict the depth of structures in the patient
- Used in opthalmology to measure the axial length of the eye & in obstetrics to
measure the size of the fetal head - A-mode uses high US frequency (typically 10MHz) to improve accuracy
B-Mode Scan
Brightness Mode
We fire a scan line into the patient, we’re going to get partial reflection at these two positions. And in the image, we see two dots which match the depth of the reflecting surfaces. The brightness of the dots reflects the magnitude of the echo at those places, and then we do that for subsequent lines across the image.
M Mode Scan
stands for motion. With this, it’s like a B mold scan, except you only scanned on a single line, so there’s a single scan line.