extra stuff to know for test from all weeks Flashcards
how to get into kohler
- close radiant field diaphragm
- open condesner aperture cimpletely
- raise condesner to highest position
- looking into microscope lower condenser and stop when diaphragm image is in sharp focus- crisp edges
- center image using screws
- open radiant field diaphragm until view field is completely filled with light-stop when light is just outside field.
- remove ocular and close condenser until only 70% is of light coming in
- adjust brightness to personal preference
how to set up microscope
turn microscope on put everything to middle setting ensure nosepiece in park clean oculars clean lens in increasing order adjust eye distance raise condenser to highest position
what are volumetric pipettes used for
accurate measurement of aqueous fluids such as controls or standards
what are mohr pipettes used for
used to measure fractions of a total volume-same solution into multiple test tubes
not as accurate as volumetric pipettes
used to measure reagents or water
not accurate enough for samples or standards
what are semiautomatic pipettes and what are they used for
they are MLA pipettes, eppendorf pipettes and hamilton pipettes and they are used to measure small volumes of liquids in microlitres
use of biological safety cabinet class 2A2
uv off drain closed
-flourescent light on
-run blower for 5 minutes and record pressure gauge
-ensure sash is open no more than 10 inches
-disinfect all interior surfaces of BSC
-place everything you need in in cabinet
-wear PPE use absorbent pad for spills
-keep area clean
keep intake grills free of obstacles
work 10-15 cm inside of the cabinet
-limit arm movements in order to not disrupt air flow
-wait one minute after you have moved arms into the cabinet veer you start to work to let airflow stabilize
-dont use flames
what to do when done using the bunsen burner
- ensure containers are closed and covered outsides and disinfected before properly disposing of all contaminated materials
- disinfect with cavicide or other appropriate disinfectant that will not damage metal
- allow cabinet to purge by leaving the blower on for at least 5 minutes after cleaning.
- remove gloves and dispose of them in the BSC
How to use chemical fume hood
- fluourescent light off
- check are vents free of obstruction
- make sure sash at proper height
- make sure hood is working properly