Troubleshooting Flashcards

1
Q

How many digits make up the error code

A

6

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Error severity level 1

A

Indictates an easy to resolve problem, keep going

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Error level 2

A

Non-locking error1. retry2. ignore3. cancel and re-start

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Error level 3

A

Locking error, must stop procedure

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

What does the second number mean in the error code

A

Used by tech support to know which phase the error occured

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

last 4 number error code

A

Needed to be recorded and shared with EDAP after the procedure

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Error 1102

A

Probe tempterature is too high

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

resolve probe temperture too high

A
  1. pump is running2. fokal pac head/circulation pump close around tube3. no tubing is clamped4. ablasonic fluid leaking
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Error 1101

A

Probe temp too low

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Probe temp too low troubleshooting steps

A

insert and allow the body to warn to target

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Error 1202

A

Patient moved

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

resolve patient movement error

A

Check reflector to make sure it isnt coveredmodify contours

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Error 1302

A

Distance from the wall to the probe is greater than 4 mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Error 1303

A

Distance from wall to probe is between 1-4 cm, system will try autocatch up

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Wall position error where the first measurement is less than the second numbert

A

probe too close to rectal wall

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Wall/probe position error, 1st number greater than 3rd

A

Probe too far from wall

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

Causes of error 1302/1303 6

A
  1. patient moved2. us image too light/dark3. artifacts in balloon4. rectum not clean5. balloon burst6. ballon not adhering to rectum
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

Resolve error 1302 or 1303 5

A
  1. remove any artifects from balloon2. low difference, press retry 3 times3. determine if probe is too far/short4. patient movement, re-do localization5. Change balloon
19
Q

Error 1450

A

Key is pressed on keyboard or probe

20
Q

Error 1601

A

Communication problem with US

21
Q

Resolve error 1601

A
  1. check green light2. if off, turn on3. if on call service
22
Q

Error 1604

A

Probe cannot be unblocked

23
Q

Resolve error 1604

A

check/re-connect image connector and ensure US probe is locked

24
Q

Error 1607

A

US probe disconnected

25
Q

Error 1901

A

Incorrect Focal Pak code

26
Q

Error 1902

A

Focal Pak already used

27
Q

What is the duration of the Focal Pak after probe insertion is validated

A

5 Hours

28
Q

Error 1903

A

Focal Pak expired

29
Q

Error 5003

A

Emergency stop depressed

30
Q

If patient moves during scanning, course of action

A

redefine UL/LL

31
Q

Steps to adjust orange line to the rectum if not synced

A
  1. increase/decrease Ablasonic Fluid2. Move probe transverse3. Correct patient positioning
32
Q

Causes “fleeing”:

A

Shoulders are not pushed inwareds and parrellel to the long axis of the table

33
Q

What is fleeing

A

the organge line moves away from the rectum

34
Q

Min distance from the probe to the wall

A

23 mm

35
Q

Hyperchoic is id’ed by what color

A

white

36
Q

Intrinsic phenomenom when micro bubbles are formed and seen on US

A

Cavitation

37
Q

Actions in case of cavitation

A

Pause and wait for the bubbles to clear

38
Q

Process after recieving warning

A
  1. Read2. document3. Make corrections4. Abort5. Inform Focal One
39
Q

4 ways to tell if the patient moved or reflector isn’t properly positioned

A
  1. No reflection2. Blue light off3. Error4. Procedure will not progress
40
Q

what are the 3 types of cavitation

A

popcorn, foggy, barrier

41
Q

White clouyds in the treatment area

A

Popcorn

42
Q

Can’t ID limits between prostate and other strcutures

A

Foggy

43
Q

What casues fogging

A

accumulation of energy

44
Q

Barrier cavidation

A

happens after fogging and the rectal wall thickness increases