Cleaning and Sterilization Flashcards

1
Q

Probe removal steps 7

A
  1. Close Ablacsonic pouch and lower gallow2. open clamp3. open pump head4. put shield on and remove probe5. select main menu6. Move system7. Remove and disenfect probr
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

When cleaning the probe, how much and many flushed need to occure

A
  1. 60 cc2. Each lumin twice
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Probe rinsing steps

A
  1. Soak in warm bath 1 minute1. flush channels with tap water3. Repeat flush 3 more times for a total of 4
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Drying probe steps 5

A
  1. Remove from guide2. use compressed air to clear channels3. wipe with soft, lint free cloth4. Inspect 5. Attache probe protector
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Cleaning cords, rinsing and brush probe steps 4.

A
  1. Start at the distal end of the cord s and move proximally using a sanity whipe or damp cloth2. Detach probe protector3. Rise with luke warm tap water for 75 seconds3b. transucer indirectly rised for 30 seconds4. flush channels
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

What is the ideal distance between the orange dotted line and the rectal wall

A

23 mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Cleaning, immersion and brushing probe steps 5

A
  1. Fully emerce in cleaning solution2. let soak for 5 minutes3. Halfway through the use soft bristled brush to manually clean the probe and proctor for 1 minute4. remove visable debris, look at small crevases5. flush each channel twice
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

High level cleaning steps 5

A
  1. prep cleaning solution2. clean cords/rinse brush probe3. Immersing and brushing4. Rinse5. Dry
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Manual disinfection steps

A

1, wipe cords up to protectors2. remove protector from probe3. sumerge probe and proctector in high level disinfection bath and flush with 60cc4. Let all elements soak

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Manual rinsing steps

A
  1. prep sterile bath2. transfer from bath to sterile water3. soak probe and protector 1 minute4. flush 240cc
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

How often does the probe need to be calibrated

A

every 6 months

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

How is the transducer dried

A

Air dry

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Type of dissenfection on probe

A

High Level

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Recommended probe cleaning solution

A

ph neutral enzematic

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

In what order should the workstation componets be cleaned 10

A
  1. screen arm2. screen back3. tablet4. keyboard5. cooling tank6. probe support7. superior cover8. metallic rail9. 4 lateral covers10 breaks, wheels, pads
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

What is recommended to clean temperature sensistive componets

A

Paracetic acid

17
Q

Common causes for fusion errors 4

A
  1. surgery between DX and fusion2. Hormon therapy3. artifacts generated from the BX4. Prostate limts are different
18
Q

Computer controlled medical device for the transrectal HIFU ablation of prostate tissue

A

Focal One Description