why do we need 30x30x30 tank?
because water should extend 10 cm beyond the field you are measuring on all 4 sides
why do we have to correct for stopping power with depth for electrons but not for photons?
Electron stopping power ratios change a lot as a function of depth because they deposit energy along their path. Photons on the other hand have infrequent but powerful interactions- the new electrons set in motion have roughly the same energy
why does TG51 use 10 cm depth?
beyond the depth of highest energy contaminant electrons, and TCPE exists for the highest energy beam
issues with different detectors for electron PDD measurements
- in buiulkd-up region of electrons, cylindrical chambers are inaccurate compared to PP/diamond- issue is more pronounced for 6 and 9 MeV because you have a more prominent build up region – important when commissioning!
- Cylindrical- correct for Prep and stopping power ratios
- PP-correct for stopping power ratios
- diamond - don’t correct at all cause it should be like water
explain how TG21 measures the dose to the gas in the chamber (Dgas)
- kerma at point P in free space is Kc=X * (W/e)
- the collision kerma at an infinitesmal piece of wall placed at P is Kc,w = Kc * (uen/p) wall to air
- The little wall P is centered within the spherical. wall attenuation reduces Kc,w to AKc(uen/p)wall to air
- the absorbed dose in the wall material at P is Dw=Kc,w
- dose to gas in the sphere is Dg = Dw * (L/p) from gas to water