patient dosimetry Flashcards
1
Q
issue with high Z detectors
A
over-response for low energy scattered photons
-issue in larger fields, heavily blocked fields
2
Q
considerations for IMRT dosimetry
A
- 3D CRT- can measure a subset of beams is sufficient
- IMRT: fluence is varied during treatment so requires integrative dosimetric techniques
- IMRT- sometimes independent dose calculation methods are used as QA
- have to consider oblique-incidence response of clyndrical ion chambers for non-coplanar beams
- Volume averaging is an issue for detectors in small fields- dose will be underestimated due to averaging over the entire size of the detector
- if measuring profiles of dose distributions with rapidly spatially varying doses, the shape of the dose distribution will be blurred by volume averaging – use higher spatial resolution detector if accurate measurement of profile shape is needed
3
Q
what is arc check?
A
4D array which “measures and correlates gantry angle, leaf-end position, absolute dose and time to identify any potential source of error throughout the patient volume