Ortho Flashcards
what does ortho calibration factor yield?
dose to small mass of tissue in air
why can we use uen/p and not stopping power ratios for dosimetry in ortho?
small range of electrons (order of microns, < 0.5 mm)
-not the case in MV beams
what is uen/p a function of?
- uen/p in air is a function of HVL only, HVL defined for narrow beam conditions. Exception is for bone, where it also depends on depth, field size, and SSD
- in water (for in phantom calibrations), it depends on both depth and field size also
For simplicity, only HVL is used to specify the
beam quality. One should therefore keep in mind that the uncertainty on this quantity is no better than %1.5
where is the effect of photon scattering from the stem icluded?
in the calibration factor NK for the beam quality
and photon field used in the calibration
patient has a lead cutout-what do you need to account for?
backscatter will change
Is Krad significant in ortho?
No, it is neglificle so can say dose = K
What is used to define beam quality?
kV and HVL
does BSF depend on energy?
Yes
daily QA checkmate- what happens if you don;t put space or solid water between the source and checkmate?
dose saturates the detectors
use of short cones vs long cones
- small cones= higher dose rate; large cones= easier to reach hard-to-reach places, but slower dose rate and heavier
- long cones are used for larger FS- easier to get a more uniform field because the IS effect is less significant than it is at 30 cm cone (dose at edge of field vs dose at center). Also, for irregular IS on a patient (say on the face), IS effect is again less significant with long cone vs short cone.
Dose fall off for brachy (400 kV) is faster than ortho (300 kV)
In brachy we are on the steep decline of the IS curve- thus fall off is fast. In ortho we are on the flatter part- thus fall off Is slow (ortho PDD starts on the flatter part of the curve). IS is very important for these small distances and heavily impacts the PDD (PDD depends on attenuation, IS, and scatter).
why do long cones have closed ends?
electron contamination becomes problematic throughout the cone so the bottom is closed
explain virtual source
use virtual source (or isocenter, same with electrons) for IS, as due to forward scatter of electerons (more electrons scatter in than out), the actual source is closer than the physical one. To find the virtual source, have to fit the data to an equation to find the virtual sourced- different for each cone or electron applicator
What does Pq,chamb correct for in-water method?
PQ,cham accounts for the effect of the change in chamber
response due to photon energy and angular variation between chamber calibration
in air and measurement in water, the effect of chamber stem between calibration in
air and measurement in water, and the effect of displacement of water by the chamber
in the measurement in water. In water, dose is measured at 2 cm depth
at what depth is measurement made for in-water method?
2 cm
what detector is preferred for x-ray energies below 70 kV?
calibrated
parallel-plate chambers with a thin entrance window are preferred to minimize
the effect of the chamber wall attenuation. Thin plastic build-up foils should
be added to the entrance window, if necessary, to provide full electron buildup and
to eliminate electron contamination (thicker foil for higher energies, but below 100 kV (i.e. in PP dosimeter range)
-need sufficient build-up to make sure electrons created in wall of chamber are absorbed within that same wall.