portal dosimetry Flashcards
can you for portal dosimetry at any SSD?
Yes, but system will auto distance-scale the resultant dose image to SSD 100 and will not apply IS to scale the dose
explain gamma 3D test
Our gamma is 3 %/3 mm, normalize to dmax, only looks at points within 10 % of dmax
– Gamma- looks at distance to agreement and difference in dose value
-calculates (actual distance to agreement/ 3 mm)^2
-calculates ((dose error/Dmax)/3%)^2
gamma value is square root of sum of the two above
Pros and cons of different detectors wrt absolute dose
vented ion chamber- can be calibrated and measure absolute dose
diodes- need cross calibration
liquid filled IC- need cross calibration
pros and cons of different detectors wrt detector size and resolution
vented IC: large
liquid filled IC: smaller
diodes: smallest
pros and cons of different detectors wrt angular dependence
almost all detectors experience angular dependence
- use inclinometer and internal corrections
- make phantoms that account for the issue
- irradiate only perpendicular to the detector
pros and cons of different detectors wrt energy dependence
- diodes have energy dependence- need separate calibration files
- ionization chambers have minimal E dependence
different detectors that do 4D
- Octavius 4D- flat panel inside a cylinder and panel rotates
- ark check- diodes lie along the cylinder
- delta 4- x shaped panels- don’t rotate
- arck check and delta 4 may have worse resolution but you can rotate arc check by a few mm and retake measurements to get better resolution
detector type in arc check, octavius, and delta 4
arc check and delta 4 are diodes
octavius is liquid filled PP IC (for SRS) or vented IC
resolution of arc check, octavius, delta 4, matrixx
mapcheck and matrixx ~ 7 mm
octavius vented IC is 10 mm
octavius liquid filled PP IC is 2.5/5 mm
delta4 is 5/10 mm
do plane parrelel liquid filled ion chambers (Octavius) have dose rate dependence?
yes
type of ion chambers in profiler
parrallel -plate
IMRT/VMAT QA process
- plan is calcualted on phantom geometry
- calculated dose for a known plane location is exported