portal dosimetry Flashcards

1
Q

can you for portal dosimetry at any SSD?

A

Yes, but system will auto distance-scale the resultant dose image to SSD 100 and will not apply IS to scale the dose

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

explain gamma 3D test

A

Our gamma is 3 %/3 mm, normalize to dmax, only looks at points within 10 % of dmax
– Gamma- looks at distance to agreement and difference in dose value
-calculates (actual distance to agreement/ 3 mm)^2
-calculates ((dose error/Dmax)/3%)^2
gamma value is square root of sum of the two above

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Pros and cons of different detectors wrt absolute dose

A

vented ion chamber- can be calibrated and measure absolute dose
diodes- need cross calibration
liquid filled IC- need cross calibration

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

pros and cons of different detectors wrt detector size and resolution

A

vented IC: large
liquid filled IC: smaller
diodes: smallest

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

pros and cons of different detectors wrt angular dependence

A

almost all detectors experience angular dependence

  • use inclinometer and internal corrections
  • make phantoms that account for the issue
  • irradiate only perpendicular to the detector
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

pros and cons of different detectors wrt energy dependence

A
  • diodes have energy dependence- need separate calibration files
  • ionization chambers have minimal E dependence
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

different detectors that do 4D

A
  • Octavius 4D- flat panel inside a cylinder and panel rotates
  • ark check- diodes lie along the cylinder
  • delta 4- x shaped panels- don’t rotate
  • arck check and delta 4 may have worse resolution but you can rotate arc check by a few mm and retake measurements to get better resolution
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

detector type in arc check, octavius, and delta 4

A

arc check and delta 4 are diodes

octavius is liquid filled PP IC (for SRS) or vented IC

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

resolution of arc check, octavius, delta 4, matrixx

A

mapcheck and matrixx ~ 7 mm
octavius vented IC is 10 mm
octavius liquid filled PP IC is 2.5/5 mm
delta4 is 5/10 mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

do plane parrelel liquid filled ion chambers (Octavius) have dose rate dependence?

A

yes

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

type of ion chambers in profiler

A

parrallel -plate

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

IMRT/VMAT QA process

A
  • plan is calcualted on phantom geometry

- calculated dose for a known plane location is exported

How well did you know this?
1
Not at all
2
3
4
5
Perfectly