Stimulation Flashcards

Provides an understanding of the concepts used by devices to appropriately deliver pacing therapy. Currently weighted 3.5% in the CCDS exam.

1
Q

True / False

‘Rheobase is described as the lowest stimulation voltage amplitude at an infinitely long pulse duration’.

A

True.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

True / False

‘Rheobase = Threshold Voltage at Pulse Width of 2.0ms’.

A

True.

Strength Duration Curve is essentially flat at a pulse width of 2.0ms.

Rheobase = Measured Threshold @ 2.0ms.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

True / False

‘Chronaxie is described as the lowest stimulation voltage at an infinitely long pulse duration’.

A

False (Thats the Rheobase definition).

Chronaxie is pulse width at double rheobase voltage Threshold pulse duration at twice the rheobase voltage amplitude (i.e. pulse duration at double the voltage threshold).

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

The following statement best describes what?

‘Approximates the point of minimum threshold stimulation energy (micro Joules/μJ)’

A

The Chronaxie.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Are modern pulse generators constant voltage or constant current generators?

A

Constant voltage.

Internal resistance is very low compared with the load resistance it is giving power to.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Define strength duration curve differences between constant voltage and constant current generators.

A

Constant current = current decline is more gradual.

Constant voltage = current decline is more abrupt.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

The following statement best describes what?

‘Stimulation thresholds measured by decrementing the stimulus voltage until loss of capture are usually 0.1-0.2V lower than when the stimulus voltage is gradually increased from sub-threshold until capture is achieved’.

A

The Wedensky effect.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Which equation calculates Stimulation Energy Consumption of a device circuit?

A

E = V2 / R x t

E = Energy / Joules / uJ
V = Voltage / Volts / V
R = Resistance / Ohms
T = Time / Pulse Width / Msec

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Doubling the pulse duration does what to the respondent stimulation energy?

A

Respondent two-fold increase in stimulation energy.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Doubling the stimulation voltage does what to the respondent stimulation energy?

A

Respondent four-fold increase in stimulation energy.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

True / False

‘Longer pulse durations than necessary dont result in a wastage of stimulation energy without providing a substantial increase in output safety margin’.

A

False.

Longer pulse durations than necessary will waste energy.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

In CRT systems, is the RV electrode considered the anode or cathode?

A

RV Electrode is the Anode (Positive).

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

In CRT systems, is the LV the anode or cathode?

A

LV Electrode is the Cathode (Negative).

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

With respect to CRT, list 3 reasons why Anodal capture is undesirable.

A
  1. Earlier activation of the RV
  2. Not ideal in patients who require a desirable offset between the LV and RV
  3. May prove deleterious to biventricular
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

True / False

‘Auto threshold testing can use evoked response to determine capture/non-capture’.

A

True.

Commonly evoked response is used solely or in conjunction with other variables.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

The following statement best describes what?

‘Electrical event that results from myocardial capture after an output pulse’.

A

Evoked Response.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

Evoked response is commonly mistaken for what type of Artifact?

A

Polarization artifact.

High stimulation energy = Greater Polarization artifact. If auto capture keeps failing, try lowering starting voltage to reduce polarisation artefact.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

How does autocapture measure differences between evoked response and polarisation artefact?

A

Delivers Paired impulses.

First pulse captures myocardium and gives evokes response.

Second pulse occurs in the absolute refractory period, thus causing only polarisation artefact.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

What is retry mode? (Boston scientific)

A

Automatic Capture is active, the device is operating in beat-to-beat mode and is unable to confirm capture due to reasons such as loss of capture or due to signal/noise issues. Retry pacing amplitudes will be at double previous measured threshold.

20
Q

Automatic capture algorithms use beat to beat capture apart from which manufacturer?

A

Medtronic.

Utilises algorithm called capture management.

21
Q

What are the two benefits of automated capture management?

A
  1. Improve safety related to pacing by avoiding situations where loss of capture could occur due to a rise in the stimulation threshold
  2. Improve battery longevity by reducing the programmed safety margin closer to the stimulation threshold
22
Q

Voltage drop across lead body (distal to proximal) results in what?

A

Heat.

A form of wasted energy that unnecessarily depletes battery.

23
Q

Increases / Decreases

A small electrode tip size does what to the impedance?

A

Increases impedance.

24
Q

Why can’t we have even smaller electrode tip sizes than that found presently?

A

A smaller fixation helix will increase likelihood of micro-dislodgement.

25
If high impedances are good for battery longevity - why dont we install higher impedance leads?
Implantation of high pacing impedance leads for ventricular stimulation does not result in a benefit with respect to pacemaker longevity as compared with standard impedance leads.
26
Define polarisation with respect to the cathode.
Movement of positive IONs to cathode, cathode repels negative ions.
27
What has a greater effect on polarisation, Voltage or Time?
Time. ## Footnote *Effect magnitude directly relates to the duration of stimulus.*
28
List two ways to minimize polarization (1x Programmable, 1x Structural).
1. Shorter pulse durations 2. Larger surface area
29
Smaller electrode surface area = better impedance yet Larger surface area = reduced polarisation. What's the solution?
Electrodes with small geometric radii but a complex, porous surface that increases microscopic surface area may be used (aka Fractal Coating). In addition, various coatings (e.g. platinum or iridium oxide) may help minimize polarisation.
30
# Lead tip / Device can When programmed unipolar, which structure is the cathode and which is the anode?
Lead tip = cathode (negative) Can = Anode (positive)
31
# Lead tip / Lead ring When programmed bipolar, which structure is the cathode and which is the anode?
Ring electrode = Anode (Positive) Tip = Cathode (Negative)
32
In Quadripolar ICD leads, is the tip the anode or cathode?
Tip = Cathode = Negative
33
In Quadripolar ICD leads, is the ring the anode or cathode?
Ring = Anode = Positive
34
# True / False 'Chronaxie is the threshold pulse duration at twice the rheobase voltage amplitude'.
True. ## Footnote *Chronaxie = Pulse width at double the voltage threshold.*
35
# True / False 'Constant current systems use less energy than constant voltage systems'.
False. ## Footnote *Energy requirements in a constant current system are higher than in a constant voltage system. All implantable systems are constant voltage for this reason.*
36
# True / False 'Typically all constant current pacemakers are implantable pacemakers'.
False. ## Footnote *All implanted pacemakers are constant voltage. Most external / temporary pacemakers are constant current as energy is plentiful and easily accessible from the mains power socket.*
37
# True / False 'Larger pulse widths result in a greater polarisation effect'.
True. ## Footnote *Polarisation is resistance to flow. A longer pulse width will increase the respondent polarisation effect.*
38
# Yes / No Is it possible to program Bipolar with the following configuration? 'Lead Tip = Cathode, Device Can = Anode'.
No. ## Footnote *The device cannot be included in a bipolar circuit. Device can only form part of the circuit in unipolar configuration.*
39
Capture hysteresis is also known as '\_\_\_\_\_ effect'
Wedensky effect.
40
List 3 potential causes of stimulation exit block
1. Antiarrhythmic drugs 2. Electrolyte imbalance 3. Myocardial Infarction at pacing site
41
# True / False Energy requirements for pacing and defibrillation differ by a factor of: 1. 10 2. 100 3. 100,000 4. 1,000,000
4 - 1,000,000.
42
# True / False Exercise increases myocardial stimulation threshold.
False ## Footnote *It decreases stimulation threshold.*
43
At an pulse duration of 2.0ms the threshold voltage is 1.5V. At a programmed amplitude of 3.0V the pulse width threshold is 0.4ms. What is most likely? 1. Rheobase 3.0V 2. Rheobase 0.4ms 3. Chronaxie 0.4ms 4. Chronaxie 2.0ms 5. Chronaxie 0.8ms
3 - Chronaxie is pulse width capture at x2 rheobase voltage. ## Footnote *Rheobase is 1.5V in this example.*
44
When connecting a unipolar lead to the generator - what is the next step? 1. Ensure both screws are fully torqued onto lead pin 2. Verifty pacing before placing device into pocket 3. Place device into pocket, then verify pacing
3 - Place device into pocket, then verify pacing. ## Footnote *Unipolar leads only have one set screw and will not pace until generator is in contact with the skin.*
45
For bipolar pacing systems, which is true? 1. Neither the anode or cathode are in the heart 2. Both the anode and cathode are in the heart 3. The anode is in the heart, cathode is in the device pocket 4. The cathode is in the heart, anode is in the device pocket
2 - Both anode and cathode are in the heart.
46
Calculate the energy delivered below. '5V, 565ohm, 0.4ms'
E = (V2/R)\*T E = (52/565)\*0.4 E = 0.018 E = 18 microjoules