Follow-up: Programming Flashcards
Provides exposure of various programming scenarios encountered during clinic follow up. Currently weighted 6% in the CCDS exam.
Programming triggered modes (AAT, VVT) is useful to determine what?
To determine when sensing threshold is achieved.
i.e. At what point during the P-wave does the device register the event.
What mode can be programmed to temporarily correct myopotential oversensing and/or noise due to lead failure until a permanent solution can be achieved?
Triggered (AAT/VVT) - Prevents under-pacing due to oversensing. Non-competitive against intrinsic conduction*. Good for non-dependent pts.
Asynchronous (AOO/VOO/DOO) - Paces at programmed rate but will compete against intrinsic conduction. Good for dependent pts.
*Triggered beat will fall within refractory of intrinsic beat - Functional non-capture.
How would you reduce safety pacing // functional non sensing in patient with frequent ectopy?
Shorten the ventricular blanking period.
List two indicators that show a devices battery has hit ERI.
- Percent or fixed decrease in pacing rate
- Change to a simpler pacing mode E.g. DDD to VVI
Define the following
- BOL
- ERI
- EOL
- BOL = Beginning of life
- ERI = Electrive replacement indicator
- EOL = End of life
What are the 4 types of data one receives when interrogating a device?
- Administrative data - E.g. Patient details
- Programmed data - E.g. Mode, Rate
- Real-time data - E.g. Testing values
- Stored data - E.g. Events
Is the EGM of an event stored in the RAM or the ROM of the device?
EGM is stored on RAM = read and write capabilities.
ROM = installed by manufacturer and can’t be modified.
In clinic you are presented with this atrial lead noise. List two ways to program around it.

- Program the lead uniploar and test via provocation manouvers.
- P-wave amplitude is considerably larger in amplitude than the noise. Thus decreasing atrial sensetivity (bigger number) to filter out noise while retaining ability to ‘see’ the p-wave may be appropriate.
Patient mentions palpitations - what programming changes would you make with relation to Mswitch?

Reduce Mswitch entry criterion & Increase Mswitch exit criterion.
Mswitch resolved inappropriately due to atrial undersensing. By making exit criteria more difficult (more beats necessary at normal rate) undersensing is accounted for to some degree.
Reduce Mswitch entry criteria - this means fewer beats are necessary to enter Mswitch, which means less likely to experience symptoms associated with faster V-pacing rate.
You see dropped p-waves on this IEGM, what would you adjust to ensure appropriate sensing?

Increase atrial sensetivity (lower number).
The amplitude of the dropped p-waves is significantly less than the preceeding sensed p-waves. Thus making the atrial channel more sensitive is likely to correct the problem.
Would you program a shorter AVD and promote pacing in this patient?

Yes.
Although intrinsic rhythm is generally something to be preserved, an AV delay >450msec is unlikely to promote AV synchrony. Thus it is advisable to program normal physiological AV delays and increase V-pacing.
What is the likely programming change seen here?

Both leads have changed from unipolar to bipolar sensing configurations.
Note how ventricular IEGM goes from broad to narrow (represents sensing field of view going from lead tip-can (large) to lead tip-ring (small)).
Note how far-field ventricular signal is removed from the atrial channel.
What is the appropriate mode for each number?

- AAIR
- DDDR
- DDDR
- AAIR
What is the appropriate mode for each number?

- VDD / DDD
- DDDR
- VVIR
What would be an appropriately programmed output voltage?

1.5V
Double the threshold voltage value = approriate programmed output.
Atrial rates above ______bpm will trigger a switch to DDI pacing?

>160bpm.
ATR mode switch rate as illustrated.
Yes / No
Would IRS+, MVP, VIP or RYTHMIQ be appropriate for this patient?

No.
These are all algorhythms designed to reduce RV pacing. This patient has CHB as evidenced by 100% RVp. These algorhythms would likely invoke symptoms.
Yes / No
Would IRS+, MVP, VIP or RYTHMIQ be appropriate for this patient?

Yes.
These algorhythms aim to reduce RVp% by promoting intrinsic rhythm. This patient is clearly suitable for such treatment as 99% of V-events are sensed - thus evidencing intrinsic conduciton.
From the below sensing test what would be the most appropriate mode to program?

DDD.
Rhythm shows good sinus function with 2:1 conduction. DDD will allow for appropriate sinus tracking and AV synchrony.
You observe the following, how could one program around this without changing the pacing mode?

Program atrial lead from bipolar to unipolar.
Removing the outter ring conductor (most likely to fracture) will eradicate the noise. However be aware unipolar programming leads to increased risk of oversensing myopotentials. Reprograming the mode to VVI would risk sacrificing AV synchrony should they become dependent and could bring about paceaker syndrome.
Yes / No
Would IRS+, MVP, VIP or RYTHMIQ be appropriate for this patient?

No.
These are all algorhythms designed to reduce RV pacing. This patient has CHB as evidenced by ~100% RVp. These algorhythms would likely invoke symptoms.
To remove the observed ‘AT’ sensed events, would one increase or decrease sensitivity?

Decrease sensitivity by Increasing the programmed mV number.
What would be an appropriately programmed output for this patient?

Double threshold voltage (0.9 x 2) = 1.8V
What two programming changes could be made to remove the observed far field atrial sensing?

- Decrease atrial sensitivity (intrinsic activity appears double the amplitude of the far field signal).
- Extend the atrial blanking window (Far Field Protection).
























