Sensing Flashcards

Provides an understanding of the concepts used by devices to appropriately detect intrinsic activity. Currently weighted 2.5% in the CCDS exam.

1
Q

What 3 components determine polarisation effect?

A
  1. Pulse width
  2. Stimulation output
  3. Electrode radius
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Amplitude and slew rate can abruptly decline 1 week post implant - when do values typically return?

A

6-8 weeks post implant.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

True / False

Unlike pacing, sensing circuits should have low/no resistance.

A

True.

Sensing circuits that follow pacing circuit design may result in diminished sensing.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Define the difference between source impedance and input impedance.

A

Source impedance = Voltage drop from origin of IEGM to proximal portion of lead.

Input impedance = Impedance of sensing amplifier itself.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

The IEGM seen by the pacemaker is the ratio between which two impedances?

A

Source impedance vs Input impedance.

Large difference = less signal attenuation.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

What 5 factors affect sensing accuracy?

A
  1. Lead integrity
  2. Electrode characteristics and placement
  3. Lead polarity
  4. EMI
  5. Properties of the myocardium
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Oversensing / Undersensing

A fractured wire with intermittent contact will lead to which outcome?

A

Oversensing.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Oversensing / Undersensing

A fractured wire with no contact will lead to which?

A

Undersensing.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

What are historically acceptable sensing values for P and R waves in pacemakers?

A

P wave >2mV

R-wave >5mV

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Where are myopotentials most likely to arise from when ICD is programmed integrated bipolar?

A

Diaphragm or Pectorals.

These muscles are in the devices ‘Field of View’. More likely to occur in integrated bipolar leads vs true bipolar.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

List 4 ways to reduce ICD over sensing (x4 programmable / x1 interventional).

A
  1. Reduce max sensitivity
  2. Decrease sensitivity decay
  3. Decrease LRL or Increase AV delay (Minimise RV pacing)
  4. Prolong detection interval - Transient over sensing will terminate
  5. Implant separate RVOT sensing lead (Further from diaphragm)
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

List 3 ways to stop T-wave over sensing.

A
  1. Reduce max sensitivity
  2. Increase refractory periods
  3. Reduce auto sensitivity aggressiveness
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

The IEGM shows two non-physiologic sensed intervals. List two potential sources.

A
  1. Loose set screw in header
  2. Lead defect
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Yes / No

Is it necessary to have both near field and far field sensing channels in ICD systems?

A

Yes.

Far field helps to show morphology changes and also can detect atrial signals - thus showing VT vs SVT.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

If ICD sensing can’t be sorted with programming, what is the next course of action?

A

Re-intervention should strongly be considered E.g. Small R-wave // Big T-wave.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Generally speaking, how does the ICD noise detection algorithm work in ICDs?

A

Identifies marked differences between near field and far field IEGMs.

Tip - Ring = Near Field

Coil - Can = Far Field

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

Noise secondary to a conductor failure will typically be seen on which IEGM?

A

Near field (Tip - Ring).

A conductor fracture is less likely to interfere with far field EGM.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

Noise detection algorithm is ineffective against EMI - Why?

A

Noise will appear on both near and far field channels - thus therapy will likely be delivered.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

Is ICD sensitivity more or less sensitive following a paced beat?

A

More sensitive.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

With ICD sensing are post pacing blanking intervals longer or shorter than post sensing intervals?

A

Post pacing blanking periods are longer.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

List 3 programmable ways to decrease TWOS

A
  1. Decrease max sensitivity
  2. Prolong refractory periods
  3. Alter auto sensitivity (Amplitude and Delay Decay)
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

Near Field / Far Field

Which is more likely to detect morphology changes during tachycardia?

A

Far Field.

A greater field of view gives a larger observed degree of vector change.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

Define the difference between consecutive and probabilistic counters.

A

Consecutive = one following other.

Probabilistic = certain quantity in a certain time period.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

What do VF counters generally accept % wise.

25
How long should an ICD take to charge.
5-12 seconds.
26
The following statement best describes which form of sensing circuit? 'Both anode (+) and cathode (-) are located in the heart'.
Bipolar sensing.
27
The following statement best describes which form of sensing circuit? 'Anode (+) is in the generator pocket and Cathode (-) is in the heart'.
Unipolar sensing.
28
Atrial and Ventricular Slew rates should exceed what respective values at implant?
P-wave = 0.5V/s. R-wave = 0.75V/s.
29
The ability of a device to adequately sense is determined by which two values?
1. Impulse Amplitude 2. Slew rate
30
The following formula describes what? ΔV/ΔT
Slew rate. ## Footnote *Delta change in voltage / delta change in time.*
31
Sensing amplifiers are typically most sensitive between what Hz bandwidth?
30-40Hz.
32
What is the approximate sensing field of view for both Unipolar and Bipolar sensing circuits? (in cm)
Unipolar = 30-50cm - FOV is lead tip to Generator can. Bipolar = 3-5cm - FOV is lead tip to lead ring.
33
# Unipolar / Bipolar Which pacemaker sensing circuit is most susceptible to inappropriate pacing inhibition due to oversensing?
Unipolar. ## Footnote *Due to larger sensed field of view. Impulses of non cardiac origin are 'seen' by the device and confused for genuine cardiac signals - hence inhibition of output.*
34
# Overpace / Underpace If a device is oversensing it is most likely to do which of the following?
Underpace (Inhibit inappropriately). ## Footnote *The device is fooled into thinking there are intrinsic cardiac impulses, thus will inhibit pacing output.*
35
What Hz bandwidth do device bandpass filters typically allow?
20-40Hz.
36
The following statement best describes what generator component? 'Permits frequencies in a defined range yet attenuats frequencies outside that range'
Bandpass Filter.
37
# True / False Review this statment with respect to a single lead system 'Oversensing will lead to under-pacing, while under sensing will lead to over-pacing'
True. ## Footnote *Extraneous impulses of non cardiac origin will fool the pacemaker to inhibit necessary output. An inability to sense intrinsic events will fool the pacemaker to think the myocardium is inactive, thus delivering inappropriate output.*
38
# True / False Review the following statement respect to a dual lead system programmed DDD. 'Atrial oversensing will lead to ventricular under-pacing'.
False. ## Footnote *DDD is a tracking mode. Thus it is likely the pacemaker will track the oversensed impulses on the atrial channel and over-pace the ventricle in response.*
39
What is the Atrial impulse density as an expression of Hz bandwidth?
Atrial signals are typically 80-100Hz.
40
What is the Ventricular impulse density as an expression of Hz bandwidth?
Ventricular signals are typically 10-30Hz.
41
# Increase / Decrease What effect will an increase in stimulation pulse width have on the corresponding polarisation effect?
Increase. ## Footnote *Increasing pulse width or 'time the current is switched on' will heighten the polarisation effect.*
42
# Increase / Decrease What effect will an increase in electrode radius have on the corresponding polarisation effect?
Increase. ## Footnote *Fractal coatings on electrode tips help to increase surface area without increasing physical radius, thus keeping polarisation effect low.*
43
# Increase / Decrease What effect will a decrease in stimulation voltage have on the corresponding polarisation effect?
Decrease. ## Footnote *Stimulation voltage is directly related to polarisation effect. Lower voltages correspond to lower polarisation.*
44
Amplitude and slew rate can abruptly decline 1 week post implant - discounting dislodgement, why does this occur?
Localised Oedema / Inflammation ## Footnote *Anti-inflammatories are typically affixed to the lead tip to largely negate this in modern systems.*
45
What are acceptable sensing values for P and R waves in ICDs?
P wave \>2mV R-wave \>10mV
46
Why is ICD sensitivity is typically more sensitive following a paced beat?
To prevent the phenomenon of pacing into VF.
47
# True / False 'Voltage difference between two electrodes creates the sensed signal sent to the device'.
True.
48
# True / False 'A generator with sensitivity fixed to 3mV will sense only cardiac signals 3mV and greater'.
True. ## Footnote *Anything smaller than 3mV will not be sensed.*
49
The following statement best describes what? 'The change in signal amplitude vs. time'.
Slew Rate.
50
List 3 examples of extra-cardiac signals that present with very high slew rates.
1. Skeletal Muscle Myopotentials 2. Electrical Interference 3. Electromagnetic Interference ## Footnote *The signals are normally attenuated via the bandpass filter.*
51
List an example of a cardiac signal that presents with a very low slew rate.
The T-wave. ## Footnote *Normally attenuated via the bandpass filter.*
52
# True / False 'A bandpass filter discriminates signals by their amplitude'.
False. ## Footnote *A bandpass filter discriminates via frequency (Hz) not amplitude (mV).*
53
# True / False 'A level detector discriminates signals by their amplitude'.
True.
54
# True / False 'A V-sensed event with no preceding A-event (as detected by the pacemaker) will be labelled a PVC'
True. ## Footnote *This also holds true if the atrial event falls within the PVAB.*
55
Define the difference between true and functional undersensing.
True = Amplitude of signal is too small to be detected at current programmed sensitivity levels. Functional = Signal falls within a programmed blanking or refractory period and is thus not detected.
56
List one possible explanation for undersensed atrial events post PPM defined PVC, despite them occurring after the normal programmed PVARP interval.
Most systems extend the PVARP interval following a PPM defined PVC. ## Footnote *This interval is significantly longer than the normal PVARP interval, upwards of 400msec.*
57
Which of the following is the correct description of each of the four recordings?
E.
58
Which is important for sensing evoked potentials necessarry for capture confirmation algorythms? 1. Low impedance lead 2. Higher french lead 3. High impedance lead 4. Low polarization lead 5. High polarization lead
4 - Low polarization lead.
59
A patients device is undersensing at 5mV. What would be the next appropriate step. 1. Decrease sensitivity to 7mV 2. Increase sensitivity to 4mV 3. Increase the output 4. Increase sensitivity to 10mV
2 - Increase sensitivity (by decreasing the number) to an acceptable threshold.