radiation protection aka dose reduction part 2 Flashcards
handheld, portable dental xray unit
NOMAD, NOMAD Pro and NOMAD Pro II
position and distance
stand 6 feet away (stand straight)
90-135 degrees to primary beam
NOMAD use (does/doesn’t) violate current radiology statutes
does violate
exemption for licensed held units
- backscatter shield at all times
- operators wear personal monitor device
- training in the safe use of these systems and records
this on NOMAD pro acts as barrier against backscatter radiation eliminating the need for the operator to leave the room
secondary shielding
Maximize the operator’s area of protection by doing the
following:
- Place the backscatter
shield at the end of the
PID (closest to the
patient). - Align the PID close to the
patient. - Keep all body parts within
the area of the shield’s
protection.
protection diminishes when the shield
1
2
3
- is not at the outer edge of the cone, use positioning kits that do not require the shield to be slid back
- is distanced from the subject, hold the cone close to the patient
- is not parallel to the operator, avoid this by asking pt to slightly tilt their head
more operator exposure if the sensor (does/does not) parallel the standing of the operator
does NOT
less operator exposure and better shield protection if sensor (does/ does not) parallel the standing position of the operator
does