A protective curtain, or sliding panel, with a minimum of 0.25-mm lead equivalent, should normally be positioned between the fluoroscopist and the patient to intercept which of the following types of radiation?
Scattered radiation above the tabletop
For C-arm devices with similar fields of view, the dose rate to personnel located within a meter of the patient is ____________ the dose rate to personnel from ______________.
comparable to; routine fluoroscopy.
For primary radiation, what is the term that represents a portion of beam-on time that the x-ray beam is directed at a primary barrier during the week?
Use factor
If the peak energy of the x-ray beam is 100 kVp, a protective lead apron must be the equivalent to at least which of the following measures?
0.25-mm thickness of lead
The annual effective dose limit for individual members of the general population not occupationally exposed is ___________ for continuous or frequent exposures from artificial sources other than medical irradiation and natural background and ____________ for infrequent annual exposure.
1 mSv; 5 mSv
The maximum weekly permitted equivalent dose for a controlled area must not exceed
1 mSv.
When a radiologic technologist declares her pregnancy to her employer, which of the following is appropriate?
She continues to perform her duties without interruption of employment, provided that she has received radiation safety counseling and follows established radiation safety practices.
When doors to radiographic and fluoroscopic rooms are closed during radiation exposures, a substantial degree of protection is provided for persons in areas adjacent to the room door. This is because in most facilities room doors have attenuation for diagnostic energy x-ray equivalent to that provided by _______ mm of lead.
0.8
While standing behind the control-booth barrier, a radiographer makes a radiographic exposure. The x-rays scattered from the patient’s body should ____________________.
scatter a minimum of 2 times before reaching any area behind the control-booth barrier.
Who should determine the exact shielding requirements for a particular imaging facility?
Medical physicist