Pacemakers Flashcards
Transcutaneous Pacemaker
External gelled pads are places on chest and attached to pulse generator
Can be used to defibrillate or as a pacer
Electrical charge delivered through skin
Transvenous Pacemaker
A lead wire is placed through the venous access to either the atrium or ventricle and attached to an external pulse generator
A pulse generator must have a …
rate, sensitivity, and milliamps
Permanent Pacemaker
Pulse generator implanted under skin with lead wire passed into atria and/or ventricle via venous access
What are the methods of pacing
Can be used to stimulate atria, ventricles or both
Atrial pacing is used to mimic normal conduction and produces atrial kick
Ventricular pacing stimulates ventricular contraction
Dual chamber pacing stimulates both atria and ventricle
What is the set rate
Number of impulses/min to produce adequate CO
What is the demand mode
Will generate a beat only if patient’s own intrinsic rate drops below pacer rate
What is the Asynchronous mode
Paces the heart at a set rate, doesn’t respond to patient’s own intrinsic rate
What is Sensitivity
Ability of the pacemaker to sense pt’s own intrinsic beat so there is no competition between pacemaker and pt’s heart rate.
Post-op care/teaching
Assess for bleeding, hematoma or infection
Pt should be quiet in bed, not moving the left arm much
May have to wear a sling to immobilize the left arm
No lifting more than 10lbs
Monitor for dysrhythmias or pacer failure for first 24hrs
Avoid magnets of any kind (MRI, cell phones close to PM)
Will alarm in airport - carry a card
What are the three pacemaker malfunctions
Failure to pace
Failure to capture
Failure to sense
Failure to capture and sense
Explain failure to pace and what are the causes
PM does not initiate an electrical impulse when it is supposed to. Causes: Generator/Battery problem Fractured wire Loose of connection
Explain failure to capture and what are the causes
Pacer spike but no depolarization (no P-wave or QRS) Cause: Milliamp too low Wire broken Displaced wire
What is a nursing intervention and the patient has a failure to capture d/t a displaced wire?
Turn patient to the left side to facilitate contact of pacing wire with the endocardium and septum
Explain failure to sense and the causes?
Pacer doesn’t sense the patient intrinsic rhythm and initiates an impulse
Causes:
May be caused by lead displacement