Maintenance + testing of washer-disinfector Flashcards
General
service engineer and in-house tests
some have automatic data logging devices incorporated that produce print outs go their operational parameters - validated record to prove each cycle ran efficiently and had produced clean and disinfected items
Automatic control test
ensure the cycle parameters set are actually achieved - regard to temp, time and drying
Chemical dosing
ensure the detergent/disinfectant is released into the machine correctly during cycle and that low levels of either are indicated as necessary
Thermometric disinfection test
using a heavily soiled load, the temps achieved are tested to ensure those reached are suitable to ensure disinfection has occurred