MAAC: Analytical Spectrophometry 2 Flashcards
What can UV/Vis absorption be used for in relation to drug analysis?
Quantification of drugs in formulations (without excipient interferences):
–Determine pKa of some drugs
–Determine partition coefficients and solubilities of drugs
–Determine release of drugs from formulations with time (dissolution testing)
–Monitor reaction kinetics of drug degradation
–UV spectrum used as one of the pharmacopoeial identity checks
What happens when polychromatic light passes through a sample?
Why is this?
The intensity decreases due to 3 processes:
- Reflection at phase boundaries, i.e. liquid/air, glass/liquid etc. This is caused by differences in the refractive index of the different materials through which the light is passing.
- Scattering of light caused by non-homogeneity of the sample.
- Absorbance by atoms or molecules in solution.
How do you prevent these?
- Use matched pair cuvettes (Cuvettes normally come in pairs) for the reference and sample. This way the cuvette is the exact same material for both so refractive indexes for these are both taken into account. The use of the same solvents in both refence and sample (bar drug) takes into account the effect of the liquid on the reflection of light too.
- Sample may not be completely dissolved therefore particles remain, air bubbles in cuvette, finger prints on cuvette. These will scatter light away from the detector and it will be assumed the light has been absorbed and hence give incorrect reading. To remove airbubbles the cuvette can be tapped on the table and the bubble with float to the top. To remove particles the solution must be filtered using fine filter paper. Cuvettes must be made sure they are clean inside and outside and when handling them should be held at the top corner or edges.
What wavelength is selected for the analysis?
•Usually use the wavelength corresponding to the absorption peak (λmax)
–Scan sample over region of interest and note the wavelength at highest absorption
–Can also use British Pharmacopoeia to find λmax of a particular drug or clarks if not medicinal sample
Wavelength max gives greatest sensitivity and peaks tend to be fairly flat
The following gives a reminder of molecular orbitals.
Things to remember, the larger the change in energy, the shorter the wavelength and larger the frequency
What is Beer’s Law and what is Lambert’s Law?
What is the equation that combines them?
- Lambert - Exponential decrease in light intensity as Path length increases
- Beer’s - Exponential decrease in light intensity as concentration increases
A= log I0/I = kcl
A= Absorbance
k= constant
c= concentation
l = Path length
What is the relationship between A, L and C
A is directionally proportional to L
A is directionally proportional to C
the Absorbance constant can be expressed in 2 different ways depending on the units of concentration.
When are is molar absorpitivity constant and specific absorbance used?
- If the units of concentration are molarity (i.e. number of moles per litre), constant is ε (the Greek letter epsilon) and is known as the molar absorptivity.
–ε has units of L mol -1 cm -1 although the units are seldom expressed.
–ε is equal to the absorbance of a 1 M solution in a cell of path length 1 cm
- if the concentration of the sample is expressed in percentage weight by volume (% w/v), then the constant used is A 1%, 1 cm usually written as A11
–% w/v = 1 g in 100 mL
–Called the specific absorbance, with units of dL g-1 cm-1
dL = decilitre (1/10 L) = 100 mL
When can the A11 valvue in beers law be useful?
•The A11 value is useful in pharmaceutical analyses where the molecular weight of the sample may be unknown
–e.g. when analysing a protein or DNA, or
–where a mixture of several components is being analysed in the same sample (e.g. an ointment or suspension).
What equation relates A11 and E (episilon)?
E = A11 x Mr / 100
What are examples where there may be deviations from Beer’s Law of A is linear with C?
- Instumental:
- The law is obeyed when monochromatic light (e.g. lazer) is used. A lot of the time polychromatic light is used and a monchromatic used to pick a particular wavelength. This therefore means there is a slight deviation in the results.
- Stray light in the instrument at high absorbances can lead to underreporting value of A (and curve in graph). Stray light is ight not suppose to reach the detector that may bounce of the multiple opti in the instrument. Above a value of 3 this becomes a very big factor there for low Absorbances sould be worked at
- Chemical Deviations:
- Also arise since ε depends on refractive index, n, of the medium.
- ε is only constant for the absorbing molecule or ion in a particular solvent at a particular wavelength
- pH causes the greatest deviation
- Some chromophores will shift wavelength with a change in pH
- Also arise since ε depends on refractive index, n, of the medium.
- Concentration affects
- Beer’s law describes the absorption behaviour of dilute solutions only
- Below 10 mM, concentration effects not usually observed
- Above 10 mM the close distance between absorbing molecules can alter ability to absorb a given λ
- How do you stop pH effects?
- How do you stop concentations effects?
- pH effects can be overcome in 2 ways:
- All solutions are buffered
- Take measurements at the isosbestic point (the point where the 2 overlapping curves intersect) as opposed to lamda max of both as they share a common point.
- If graph starts to curve due to conc, dilute sample further
What are the 2 methods used when using spectroscopic measurements in drug analysis?
They are different in the US and EU
Comparative (US):
- The sample concentration is worked out if you have the concentration of the standard
- Absorbance Test/ A Std = [Test]/ [std]
- Adv - Can be used even if drug undergoes chemical reaction during the assay
- Dis - Need authentic sample of drug in question for comparision (expensive)
Absolute (EU)
- Absorbance measured and BL equation used to calculate [Drug]
- Range of standard solutions created and absorbance of each measured. Calibration curve created and unknown worked out using ththis
Why should at least one other analysis method be used with Lamda max to determine the drug?
- Lamda max is characterisitic of a particular chromophore - not a particular drug. Therefore if 2 drugs have a similar structure and contain the same chromophores these can be the same.
- Lamda max can move position as the drug is ionised
What is a chromophore?
What is an auxochrome?
- The part of the molecule responsible for absorbing light - conjugated part.
- This is a functional group attached to the chromophore which does not absorb light energy itself, but which influences the wavelengths of light absorbed by the chromophore.