Interventional Flashcards
how many inches equal 1mm
0.039
what are microwire sizes 2
0.018, 0.014
how many french = 1mm
3 french
diameter in mm =
French/3
french size is external or interal of a catheter?
external
sheath size given by which diameter?
inner diameter
how to predict size of skin hole?
add 2 French to sheath to calculate outer diameter
1 french - how many mm?
0.3mm
wire diameters given in?
inches
wire lengths given in?
centimeters
18G needle will accept what inch guidewire
0.038
19G needle will accept what inch guidewire
0.035
21G needle accepts what inch guidewire
0.018
Standard wire length?
180cm
Long wire length?
260cm
bentson used for what?
thrombus lysing
floppy
lunderquist used for what?
aortic stent grafting
Stiff!
end hole catheter used in what type of injection?
hand only
catheter needed for acute angle? 2
berenstein or headhunter
catheter needed for angle of 60-120?
RDC or cobra
catheter needed for obtuse angle?
recurved catheter.
sidewinder or sos omni
rate 1-2ml/sec for 4-10ml volume
bronchial artery
intercostal artery
rate 4-8ml/sec for 8-15 ml volume
carotid SC renal femoral IMA (higher vol 15-30)
rate 5-7ml.sec for 30-40 volume
celiac
sma
rate 20-30 ml/sec for 30-40 ml volume
aorta
aortic arch
ivc
pulmonary artery
each french gives what ml for flow rate? (with 3F as baseline)
8ml/s
4F rate
16 ml/s
5F rate
24ml/s
What is a contraindication for translumbar aortic puncture?
supraceliac aortic aneurysm
what is typical access level for translumbar aortic puncture?
high: T12 endplate
low: L3
Order of preference for venous access:
RIJ>LIJ>REJ>LEJ
thrombin dose
0.5-0.1ml (500-1000 units)
Self expandable stents are good for what?
superifical locations
balloon expandable stents are good for what?
precise deployment
aortic balloon size
10-15 mm
common iliac balloon size
8mm
external iliac balloon size
7mm
CFA, Prox CFA balloon size
6mm
distal SFA balloon size
5mm
Indications for EVAR
AAA over 5cm
AAA growing >0.5cm in 6 months
landing zone requirements for EVAR
10mm long
<3.2cm wide
<60 degree of toruousity
iliac angulation > what angle may difficult device deployment?
over 90
iliac artery diameter less than what makes for difficult device deployment
less than 7mm
What is a crawford type 4 aneurysm?
12th intercostal space to iliac bifurcation with involvement of renal superior mesenteric and celiac arteries
type 1 endoleak
leak at top or bottom of graft
type 2 endoleak
filling of sac by feeder arery
type 3 endoleak
defect/fracture in graft
type 4 endoleak
porosity of graft
type 5 endoleak
endotension
varicocele tx
coils
uterine fibroid embo tx
pva or microspheres 500-1000 micrometers
generic trauma tx
gel foam