What is the frequency of ACR Accreditation?
Every 3 years
The material used for the exit port of the mammography tube is necessary because
Regular glass would harden the emerging beam
The intensity of the xray beam from the cathode side of the tube is generally higher because
The heel effect causes variation in the intensity of the xray beam
The design of the lip of the compression paddle (both height and angle along the chest wall) affects all of the following except
A) it prevents the posterior and axillary fat from overlapping the body of the breast
B) it allows uniform compression of the posterior breast tissue
C) it helps increasing the structural strength of the compression paddle
D) insurers higher compression of the anterior breast tissue (ANSWER)
The primary goal of compression is to
Allow uniform penetration of structures within the breast
What features of the high-transmission cellular (HTC) grid allows these grids to maintain equal or less radiation dose to the patient compared with linear grids?
They have copper as the strips and air as the interspace material
The grid ratio can vary in modern mammogram units. A common grid ratio used in
5:1
The chest wall edge of the compression paddle should be aligned just beyond the chest wall edge of the detector to
Avoid projecting the chest wall edge of the paddle on the mammogram
Which of the following affects the focal spot size?
Angle of the anode
Which of the following is the commonly used focal spot size for routine work mammography?
0.3 mm
Which of the following mammography quality control tests is performed as needed?
Repeat analysis
The general criteria to pass the new American College of radiology— digital mammography (ACR DM) phantom require a minimum of ___ masses
2
What is the compression thickness indicator test?
A check to ensure that the indicated compression thickness is within normal range
What is the name given to the test to determine whether the unit has consistent detector signal to noise level for all breast thickness?
Visual checklist
A technologist test that is performed semi annually is
Compression force test
Repeats are
Images that involve unnecessary exposure to the patient
To get meaningful analysis, the repeat analysis needs a patient volume of at least ___ pts
250
Mammography facilities can receive certification from
FDA or States as Certifiers (SACs)
Aluminum filtration is likely to be matched with ___ targets
Tungsten
Which of the following visual checklist tests must be performed before clinical use?
Paddles/face shield must not be cracked
What tool is used to evaluate for artifacts?
Phantom
A single binary digit of data, such as 0 or 1 is called a
Bit
In general, higher magnification will require the use of a
Smaller focal spot size
A grid is not necessary during magnification because
The large OID produces the same effect as a grid