ICM - Equipment Flashcards

1
Q

Pulse oximetry (RCoA old book)

A
Red 660/ infrared 940  Isobestic points 590/805Beer-Lambert laws (concentration and thickness respectively) SpO2<70% is inaccurate/extrapolated Causes of inaccuracy Time lag 30-60sBurns possible CO-oximeter - part of modern blood gas analyser; measures carboxyHb and methHb as well as oxyHb; therefore useful for elucidating causes of hypoxia, and demonstrating falsely low/high pulse oximetry readings
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Defibrillator (Mendonca)

A
A step-up transformer converts 240V to 5000V. A rectifier converts AC to DC. Electrical charge is stored in a capacitor before release. An inductor slows the discharge from the capacitor, lengthening the shock duration, and also absorbs some of the energy. Thoracic impedance (about 50 Ohms) is reduced by gel pads, large paddle size, and by delivering shock in expiration. Only 4% of the delivered energy passes through the heart. Critical mass theory = a critical mass of cells needs to be depolarised. Capacitance is determined by surface area of the capacitor plates and thickness of the insulating layer between. Monophasic = single current pulse in one direction. Biphasic = two consecutive pulses in opposite directions. Defib threshold lower in the latter - more efficient, smaller capacitor and battery needed. Energy = 1/2 QV. Q = charge. ICD: energy 0.1-30J. Can perform sync/unsynced shocks and anti-tachycardia pacing. Always also has PPM function for anti-bradycardia backup. External defib over ICD can cause EMI and myocardial burns. Difference between defibrillation and cardioversion = syncing. 90% success but high relapse rate. CI to cardioversion: digoxin toxicity and AF >48h without anticoagulation (need anticoag 3/52 before and 4/52 after). Remember: press sync every time, and need to press and hold shock button. Elective: keep SV, deep sedation or GA, pre-O2 but no airway required; propofol/sevo but not opioid.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Capnography (Mendonca, past Q)

A

Confirmation of ETT: clinical signs, capnography, fibreoptic bronchoscopy, oesophageal detector. Gold standard = direct visualisation and presence of EtCO2 for 6 consecutive breaths. Pattern recognition y axis can be % or kPa Normal trace - see loose notes

Phase 1 - dead space, no CO22 - alveolar gas mixes with dead space 3 - plateau, EtCO2 is right at end 0 - inspiration Alpha and beta angles Can get phase 4 in obesity and pregnancy (upswing after phase 3 - reduced thoracic compliance) Reverse phase 3 - emphysema Camel hump - lateral position Notched phase 2 - leak Dual capnogram - single lung transplant, kyphoscoliosis, endobronchial (sometimes)
Normal EtCO2 = 38mmHg/5% Flat trace: disconnection, extubation, oesophageal intubation, analyser not in circuit, analyser malfunction/blockage, resp/cardiac arrest, apnoea testing, complete obstruction High CO2 - Excess production: fever, sepsis, MH, bicarb, tourniquet release, CO2 embolism, overfeeding, high CO/BP- Reduced clearance: hypoventilation, endobronchial intubation, partial airway obstruction, rebreathing, leak, inadequate FGF, exhausted soda limeLow CO2- Reduced production: hypothermia, hypotension, cardiac arrest, PE, apnoea, total airway obstruction, disconnection, leak, condensation in sample line - Excess clearance: hyperventilationSudden rise: correction of obstruction, ROSCEtCO2 high/low: actual changes in CO2 production, altered alveolar perfusion or ventilation, technical factors. Volume capnography- Gives V/Q info - Can measure dead space - Gives CO2 elimination info- Gives mixed expired CO2 conc IR detects molecules of differing atoms (i.e. not O2). Mainstream - cuvette with quartz window; beam of IR light passes through stream. Sensor must be heated to 39C to prevent water condensing on the sensor and causing inaccuracy. Sidestream - sample aspiration 50-150ml/min. More convenient but time delay (transit and rise time) - esp in MRI. Factors affecting IR capnography - N2O - collision broadening (false high) - Water vapour (false high) - Response time - faster = more accurate. Total of transit time and rise time. - Atmospheric pressure - increases density of IR-absorbing molecules Last ICS review (2016) recommended use of capnography in cardiac arrest and CPAP (as disconnection/occlusion alarm). Existing recommendations (2011) are for intubation/tracheostomy, continuous use in all intubated pts, and on transfers. Capnography/capnometry/colourimetric devices, Severinghaus electrode, infrared, mass spec/Raman spec/photoacoustic spec.Collision broadening Mass spec Separates substances according to MW. Sample aspirated into vacuum chamber where an electronic beam ionises and fragments the sample. Ions are then accelerated by an electric field, and deflected by a magnetic field according to their MW. Lighter ions are deflected more. Detector plates measure molecules. Rapid response time and accurate but bulky and expensive. Raman Uses ‘Raman scattering’ for CO2. Sample is aspirated into a chamber where molecules are scattered by an argon laser beam. A measurable change of wavelength of light occurs. Photo-acoustic Uses pulsatile infra-red radiation. An acoustic signal is generated and measured by microphone.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

BP measurement (Mendonca, past Q)

A

Non-invasive: - Discontinuous (sphyg (Hg or aneuroid), DINAMAP or von Recklinghausen oscillotonometer) - Continuous (Penaz/Finapress, Doppler)Von R has 2 cuffs, DINAMAP 1 cuff. At SBP, oscillations start. At MAP, oscillations are maximal. Diastolic is calculated. Bladder needs to cover 80% of arm circumference. Width should be 20% greater than arm diameter. Midline of bladder to be over brachial artery. Too small a cuff over-reads, too large under-reads. Obese pts have conical arms - difficult. NIBP inaccurate at extremes of BP (over-read at low, under-read at high). Arrhythmias cause inaccuracy, as can movement/shivering. Can cause nerve damage and petechial haemorrhage. Finapress = short for FINger Arterial PRESSure. Infra-red photo-plethysmograph detects finger blood volume. Finger cuff pressure constantly adjusts to keep volume constant (null deflection). The applied pressure correlates with arterial pressure and is displayed as waveform. Invasive: arterial line Arterial cannula, tubing, transducer, pressurised flush system (300mmHg), cable and monitor. Cannula: Teflon (lower thrombosis risk), short, wide and stiff so less effect on natural frequency and damping. Fluid-filled tubing transmits arterial pressure to the transducer. Filled with saline +/- heparin (anything more viscous would cause over-damping). Flow = 3-4ml/h to reduce clot formation. Tubing is non-compliant and should be bubble-free. Length max = 122cm and number of 3-way taps minimised (they have narrower lumen and reduce natural frequency so cause damping). Info from arterial trace: HR (+ regularity of rhythm), BP, indication of preload (respiratory swing), contractility (upslope), afterload (downslope and position of dichrotic notch) and stroke volume (AUC up to dichrotic notch). CO can then be calculated. Zero calibration to RA eliminates the effect of atmospheric pressure on the measurement. Damping: tendency to resist oscillation. Caused by bubbles, clots, kinking, 3-way taps, narrow/long/compliant tubing. Damping coefficient of 1 = critical damping. DC of 0 = oscillation indefinitely. Optimal damping = 0.67 - best compromise between speed of response and accuracy (oscillation/overshoot). Over-damping under-reads SBP and over-reads DBP (values get closer). Under-damping over-reads SBP and under-reads DBP (values diverge). MAP stays same. Testing for optimal damping: square wave test/fast flush. 300mmHg is applied, causing a square waveform then oscillation. In optimal damping, 2-3 oscillations occur before values settle. Over-damped = no oscillations. Under-damped >3-4 oscillations. Comps of art lines: haemorrhage, haematoma, thrombosis, ischaemia, dissection, nerve damage, infection, aneurysm, accidental drug administration, AV fistula. Overdamped - coefficient >1Underdamped <1Critical 1 Optimal 0.64

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Fibreoptic bronchoscope (Mendonca, past Q)

A

Body, insertion cord, light source, working channel, eyepiece, dioptre ring Body houses a lever that manipulates tip Working channel can accommodate suction, epidural catheter or O2 Cord 55-60cm; light transmission bundle, working channel and control wires to tip Light carried to end of scope in fibres Light from object reflected onto lensLight focused onto image transmission bundle which carries image to eyepiece Optical fibres are 6-10microns in diameter, arranged in bundles Each fibre has a glass centre (core) where light travels, an outer optical material surrounding it (cladding) and a plastic coating (buffer coating) which protects the fibre from damage and moisture. Light hitting the cladding is either absorbed (very little), transmitted/refracted or reflected. At a critical angle of incidence, all light is completely reflected (total internal reflection). This is repeated until the light emerges at the other end. Fibreoptic relay system: transmitter, optical fibre, optical regenerator and optical receiver (photodiode).

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Ultrasound Principles, modes, uses

A

2002 NICE
- US guidance of CVCs US > 20kHz Piezo-electric crystal in probe generates US waves
- Waves reflected at tissue interfaces, converted into electrical signal
- Tissues can be hyper/hypo/anechogenic (latter = blood, air)
- High frequency = better resolution but poorer depth penetration, and vice versa
- Modes: A (single transducer), B (linear array), M (motion, for valves etc), 2D (commonest), Doppler.
- Doppler: transcranial, echo, cardiac output, fetal BART - blue away, red towards probe
- US uses: lines, blocks, drains/taps, echo, FAST, CO, MCA flow, VTE diagnosis PTX on US - loss of lung sliding, loss of comet tails (B lines (short)) but A lines (long) still present, lung point (junction between sliding and non-sliding parts) (NB lung bullae can appear similar) Spine sign on US - indicative of basal consolidation as normally can’t see the spine through aerated lung
- Pleural effusion estimated volume: max depth in mm x 20ml FAST can detect 200ml with 90% sensitivity.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Discuss ICP monitoring

A

Raised ICP: HA worse in am, N/V, altered mental status, papilloedema/visual dist. M-K curve steeper with blood and CSF as non-compressible so decompensate faster.
Cushing: HTN, bradycardia, Cheyne-Stokes respiration

Managing raised ICP: approaches to reduce blood, brain and CSF

Measurement
Indirect - clinical features, imaging evidence, papilloedema

Direct - EVD and ICP bolts (extraparenchymal - extradural, subdural, subarachnoid; intraparenchymal).

Bolts work by wire strain gauge or fibreoptic pressure transducer.
EVD ‘kept at 15cm’ means CSF column is allowed to reach 15cm above the zero point before it starts to drain. Can be recalibrated at the bedside and can treat raised ICP.

Extra/intraparenchymal devices are subject to drift. All devices carry risk of CNS infection. If invasive ICP monitor CI, can use transcranial Doppler (Lindegaard ratio - MCA to carotid flow ratio. >3 = vasospasm).

Waveform: percussion (systolic), tidal (brain compliance), dichrotic (aortic recoil)

Lundberg A - plateau waves, always pathological
Lundberg B - absence indicates compromised autoregulation
Lundberg C - can be normal

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Discuss RRT - indications, Types, Modes, Complications, PD

A

Indications: pH, K+, fluid, urea, sepsis, drug toxicity, severe hyperthermia Types: HD, PD; intermittent/continuous, VV/AV
Modes:
- Haemofiltration - water moves across by hydrostatic pressure/ultrafiltration and other molecules follow by convection (solute/solvent drag); removes larger molecules and water; no dialysate fluid
- Haemodialysis - diffusion; countercurrent dialysate; removes small solutes down conc gradt, does not remove water
- Haemodiafiltration - combined mode
- SLED - slow low efficiency HD. hybrid, faster than CRRT but slower than HD - SCUF - slow continuous ultrafiltration - slow water removal.
- Haemoperfusion - no semi-permeable membrane, blood passed over activated charcoal or ion exchange resin. Good for removing some drugs but also removes glucose and platelets.
- PLEX - removes substances up to 500kDa. Replaced with HAS/FFP.
- Plasmapheresis - centrifugation rather than membrane.

Choosing mode: aim of therapy (size of particles to be removed), CV stability (CRRT>HD), resources/staffing. Pore size 5nm. MW up to 50kDa. SA 0.5-2 m2.

Complications:
- line-related, activation of clotting cascade/inflammatory response,
- anticoagulation-related
- RRT-related (hypotension, disequilibrium, hypothermia, haemolysis, thrombocytopenia), exsanguination if line left open. Altered drug PK.

  • Dose: 25-35 ml/kg/h.

Sepsis: can remove TNF, IL-1/6/8, complement, plt activating factor on CVVH with high flow rate. Controversial. No ev improved outcome.

PD not used in ICU because: inefficient, not able to remove fluid, sepsis risk, need surgically placed catheter, causes respiratory embarrassment. However used in PICU because vascular access is challenging and peritoneal surface area is relatively larger than in adults.

Why pts dislike PD: have to be fluid restricted, body image (large abdo/catheter), causes constipation which then causes poor drainage, peritonitis/catheter infections, hernias, discomfort.

MARS/liver replacement - blood passed over a mesh of animal hepatocytes.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Cardiac output monitoring - CVP issues, types monitoring, ideal properties

A

CVP inaccuracy: pt (PEEP, TR), equipment (transducer height, damping/resonance, bag pressure)
Non-invasive e.g. NICOM, transthoracic electrical bioimpendance
Minimally invasive e.g. pulse contour methods, FloTrac
Invasive e.g. oesophageal Doppler, TOE, PAC
Ideal properties: accurate, reproducible, non-invasive, quick, simple, cost-effective, operator-independent, continuous measurement, safe, minimal drift fast response time.
Fick principle and partial CO2 rebreathing - easy and non-invasive but need to be on mandatory ventilation and is inaccurate in any condition with V/Q mismatch or barrier to CO2 diffusion, so limited usefulness in critical illness.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

ECGs (past Q)

A

FiltersCM5 Standard leads 1/2/3Unipolar V1-6Augmented unipolar aVR, aVL, aVF

QTcMax 440ms menMax 460ms women >500 risk of TdP <350 is short
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Explain Decontamination

A

Cleaning, disinfection and sterilisation (in order of strength of decontamination and how critical the relevant equipment is - non-critical, semi-critical, critical)

Cleaning: physical removal of organic matter with water and detergent

Sterilisation: steam or chemical (ethylene oxide, glutaraldehyde) Prions (infectious proteins): single use equipment (takes 10 decontamination cycles to reduce to safe levels).

Equipment can be quarantined pending diagnosis.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Pacemakers and ICDs

A
Pacemaker nomenclature (pacing/sensing/response/rate modulation/anti-tachycardia)  Rate modulation - via inbuilt accelerometer
Problems: pt, PPM and surgical factors Diathermy - EMI and local heating Fixed cardiac output/pacing dependent Key Qs: indication, mode, are they pacemaker dependent PPM with magnet - asynchronous pacing (non-sensing) ICD nomenclature: shocked chamber/anti-tachycardia pacing/method of sensing/paced chamber.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

POCT (past Q)

A

TEG (and differences from ROTEM)
Blood glucose
ABG
PEFR
ACT

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

What are some of the important features of Scoring systems

A

Ideal: Scores calculated on the basis of easily / routinely recordable variables
Well calibrated and validated
A high level of discrimination
Applicable to all patient populations
Can be used in different countries, health systems or patient cohorts
The ability to predict mortality, functional status or quality of life after discharge
Also considers co-morbidities, organisational aspects, provides a common language for discussion/to evaluate critical care practice, allows ability to compare groups in clinical trials
Uses, features and applications:
Outcome/risk/mortality/prognosis/LoS prediction for individuals and groups Treatment decisions and treatment location/tertiary referral criteria Stratification of patients for clinical trials
Assess/compare ICU/hospital performance over time or against other units Resource allocation
Description of case mix/workload
Features: Measurement of physiological variables
Some measure interventions
Derived from logistic regression from large demographic data sets
Scoring system types
- Generic (outcome prediction, workload, severity of illness)
- Organ/disease specific Assessment of scoring systems
- Discrimination - how effective it is at discriminating survivors/non-survivors (area under receiver operator curve, 1 being perfect; >0.7 considered acceptable)
- Calibration - correlation between predicted and actual outcomes

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

EEG

A
  • Alpha 8-15Hz (awake with eyes closed) - occipital cortex. May represent hypoxia if generalised and unchanged by stimulation. - Beta 15-30Hz (alert) - fronto-temporal region. Primary frequency in drug-induced coma. - Delta 1-3Hz (non-REM sleep). High voltage delta waves indicate metabolic encephalopathy. - Theta 4-8Hz (REM sleep). Dominant waveform in children; decrease with age.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

How do you challenge a Pacing box?

A
  1. RHYTHM/RATE. Reduce rate to 40 to see whether a meaningful intrinsic rhythm exists.
  2. OUTPUT. Put the rate back up until every beat is paced. Reduce the output gradually to determine the capture threshold. Make a note of it and then double it.
  3. SENSITIVITY. Put rate back down below intrinsic rate. Put sensitivity up to max and gradually reduce until indicator on pacing box recognises every intrinsic beat. Make a note of it and then halve it (i.e. double the sensitivity).
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

Pulmonary artery catheter

A

1-1.5ml air
10cm long
5-8Fr
West zone 3
Pressure transducer on distal lumen
Prox port 30cm from tip - when in position, can transduce this port for CVP Another port 26cm from tip Thermal filament 10cm long Thermistor is 4cm from tip
Est. pressures: RA 5/0, RV 25/5, PA 25/15, PAOP 6-12
PA waveform may have dichrotic notch from PV closure
Pulmonary circulation usually reached 40-55cm from IJ insertion point PAOP correlates with LA pressure and is measured with balloon inflated.
PCWP is closer to capillary pressure (and indicates LVEDP) and is measured with balloon deflated. The terms tend to be used interchangeably although they are not technically identical. Measure at end expiration. In conditions of low compliance (eg ARDS), less PEEP is transmitted and the PAOP is less affected.
Factors increasing PASP: anything that raises PVR, hypoxaemia, chronic lung disease, PE, ARDS, PHTN, sepsis, L to R shunts.
Factors increasing PADP: all the above plus hypervolaemia, L heart dysfunction, tamponade, constrictive pericarditis.

PAC should not be positioned beyond hilum on CXR and should have no loops.

Sats: SVC 70%, RA/RV/PA 75%, wedge 98%

PAOP does not equal LVEDP in MR or PE TR makes PAC and CVP readings unreliable

High PAOP: LVF, mitral/aortic valve disease, HOCM, hypervolaemia, L to R shunt, tamponade, RCM, excessive PEEP/iPEEP
Low PAOP: hypovolaemia, large PE PACMAN trial - PAC no mortality or LoS benefit, and 10% complication rate!

Risks: all CVC risks, knotting/kinking/looping, damage to PA/valves/myocardium, risk of tamponade, pulmonary infarction, arrhythmias, balloon rupture and air embolism, thromboembolism, endocarditis. PA rupture = 30% mortality.

PAC indications
Differentiating cardiogenic and non-cardiogenic pulmonary oedema
Guiding inotropic support in PHTN/HF Haemodynamic assessment when multiple conditions exist e.g. sepsis + renal/cardiac failure CO measurements
Intermittent - cold fluid bolus via proximal port; mean of 3 measurements often used; measure at end-expiration
Semi-continuous - heating coil + thermistor; reading every minute. SvO2 can also be measured semi-continuously from a distal fibreoptic probe (works like jugular venous bulb probe).

Info from a PAC
Measured: CVP, RAP, RVP, PAP, PAO/WP, CO, SvO2, ScvO2, HR, core temperature
Derived: CI, SV, SVI, SVR, SVRI, PVR, PVRI, CaO2, CvO2, DO2, VO2 Inaccuracy: incorrect temp/volume of cold fluid bolus, intra-cardiac shunts, measurements not taken at end-expiration. Uses Stewart-Hamilton equation

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

Nerve stimulator

A

Peripheral nerve stimulator (vs. percutaneous nerve stimulator)
Tetanic: 0.1ms, 50Hz
Post-tetanic count: 50Hz for 5s then 1Hz, 3 secs post.
Count of <5 = deep block; >15 = reversal may be possible.
TOF: 0.1s, 2Hz, x4 DBS: burst of 3 pulses, each 0.2ms, 50Hz, x2 (0.75s gap) Negative electrode (black) is most distal!
T4 lost at 75% occupancy T2 lost at 90% occupancy
Reverse when 3-4 twitches present or T4:T1 >0.7.
Percutaneous nerve stimulators use lower currents, around 1.5-2mA at 1-2Hz. Muscle contraction with 0.3mA or lower means needle is in the nerve itself and must be withdrawn.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

Describethe features of an Intra-aortic balloon pump

A

Indications: cardiac failure, cardiogenic shock, difficult wean from CBP, bridge to transplant, bridge to recovery
Coaxial double lumen catheter - femoral artery. Inner lumen measures art pressure, outer lumen opens into balloon.
Volume 30-50ml depending on pt size.
Should occupy 80-90% aortic diameter. Just distal to origin of L subclavian artery. Cardiac cycle monitored by arterial pressure waveform and ECG. Can be triggered by either (use art pressure in cardiac arrest so will time with CPR). Helium as low density - moves quickly, laminar flow.
Risk of gas embolus slightly higher than air/CO2.
Improved end organ perfusion - esp coronary perfusion.
Reduced afterload - reduced myocardial O2 demand.
Risks: insertion risks, anticoagulation, haemolysis, ischaemia, haemorrhage, infection, balloon rupture.
CI: severe AR, aortic dissection/AAA, severe PVD, coagulopathy, sepsis. Inflation: start of diastole - Peak of T wave (ECG) - Dicrotic notch (arterial waveform)
Deflation: just before start of systole - Peak of R wave (ECG)- Just before upstroke (arterial waveform)
Augmented early DBP is higher Augmented end DBP is lower by about 15-20mmHg
Augmented SBP is lower by about 5mmHg
IABP-SHOCK-II - no mortality diff/no better tissue hypoperfusion (mainly STEMI)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

How would you investigate Haemolysis

A

Confirmation
- Spherocytes
- Reticulocytes (>1.5%)
- Unconjugated bilirubinaemia
- Raised LDH
- Low haptoglobin (binds with high affinity to free hemoglobin)

To work out Aetiology
- Direct antigen (Coombs’) test - positive in autoimmune
- Electrophoresis - Hbopathy
- Coag - DIC
- U+E - HUS
- Plasma free Hb - mechanical destruction

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

Lung USS findings?

A

PTX- loss of lung sliding - loss of comet tails - lung point - loss of seashore sign on M-mode

22
Q

Brain stem death testing

A

Preconditions: irreversible brain pathology of known aetiology, coma (GCS 3), apnoea, absence of mitigating factors/reversible causes of coma (drugs, electrolytes, hypothermia etc). Test at least 6h after loss of the last reflex. Beware red flag conditions (neuromuscular, posterior fossa pathology, prolonged fentanyl). 1. Pupils - direct/consensual, unreactive (2–>3)2. Corneal reflex (5i–>7) 3. Supraorbital pressure (5–>7) (and no limb movement) 4. Examine tympanic mems then caloric testing 50ml ice water over 1m (8–>3/4/6) (normal = eyes deviate away)5. Gag (9, 10)6. Cough (10 –> phrenic/intercostals)7. Apnoea test - pre-O2, continuous O2 supply (either via suction catheter or Waters circuit with 5cmH2O CPAP), starting pH<7.4, starting PaCO2>6, ending PaCO2>6.5. If chronic retainer or has had bicarb, starting PaCO2 must be >6.5. CI to apnoea testing: severe hypoxaemia, high C spine injury, haemodynamic instability PaCO2 should rise by 0.4-0.8kPa/min.

23
Q

Chest drains

A

A system that allows drainage of the pleural space using an airtight system to maintain subatmospheric intrapleural pressure. Acts as a one-way valve. Drains undesirable pleural contents to allow lung re-expansion. 1 bottle system: essentially just bottle 2 of the 3-bottle system, attached to the drain. Con (for liquid drainage) is that as it fills, the pressure required to expel more fluid increases as the seal tube becomes more deeply immersed (starts at 3cm underwater so need pressure >+3cmH2O). 3 bottle system: collection (for fluid), underwater seal (for air) and pressure-regulating bottles; modern drains incorporate all 3 into one unit. Can measure output accurately, maintain constant underwater seal, and apply suction if needed. Safety features: wide tubing with low resistance; tubing and water volume both >50% of vital capacity to prevent indrawing of water/air; drain 45cm below chest to prevent contents refluxing into chest. Must keep drain upright to maintain seal. If suction is turned off then tubing must be unplugged so air can escape into atmosphere (otherwise a tension PTX). ComplicationsKinking, occlusion, retrograde fluid flow, tension PTX if clamping. Heimlich valve: unidirectional flutter valve used to replace underwater seal drains (e.g. when being transported).

24
Q

CI for Oesophageal Doppler? Positioning and principles? What can it measure? Assumptions?

A

CI - oesophageal disease, unstable C spine, thoracic AAA, IABP.
Smaller tubes exist for nasal use in unventilated patients (about same size as Ryle’s).
35-40cm from lips, 5cm more nasally
Positioned posterior and parallel to descending aorta (T5/6).
Age, weight, height entered.
US 4-5 MHz - reflected by RBCs and returned to probe.
The greater the RBC velocity, the greater the Doppler shift.
Graph of blood velocity in cm/s on y axis, time in s on x axis.
Data
- Cardiac index (N 2.5-4L/min/m2)
- SV index (N 35-65ml/beat/m2)
- Stroke distance
- AUC - distance travelled by blood per cardiac cycle
- FTc (N 330-360ms) - time taken for the SV to pass a fixed point in the aorta. Corrected for HR by dividing by the square root of the time taken for one cardiac cycle. Low = vasoconstriction, high = vasodilation.
- Preload/afterload. - Peak velocity (N 50-120cm/s)
- contractility.
- SVRI (N 2000-2400 dynes/s/cm5/m2)

  • Assumes: laminar flow, aortic diameter constant, 70% CO enters descending aorta, pt obeys normogram. Operator dependent.
25
Q

Different Ventilator settings

A

Control, trigger, cycle

VCV
Pros: greatest control over MV, Vt guaranteed regardless of compliance
Cons: barotrauma, deep sedation required to tolerate constant flow, no leak compensation

PCV
Pros: partial compensation for leaks, reduced barotrauma, physiological decelerating flow better tolerated and allows lower mean airway pressure
Cons: Vt unpredictable and reliant on compliance

26
Q

TOE

A

180 degree view of heart. Good for valves and aorta. Can exclude vegetations. Good for CHD, RWMAs/LV function intraop. High oesophageal - great vesselsMid oesophageal - main view; 2/4/5 chamber views (NB: will be opposite way up to TTE), long/short axis, bicaval Transgastric - long/short/deep axis, LV/RV function, aortic valve gradients Indications: endocarditis, intracardiac thrombus, peri cardiac surgery, CHD, masses, proximal aortic dissection CI: oesophageal disease, unstable C spine Risks: sedation/GA risks, dental damage, oesophageal trauma/perf, ETT displacement, microshock

27
Q

Sengstaken-Blakemore tube

A

Minnesota: addition of an oesophageal port CI: recent oesophageal surgery, known stricture Insertion: pt (usually) ventilated, at 45 degrees, lubricant and vasoconstrictor to nose (or can go orally), test balloons first, pass with DL to at least 50cm, confirm gastric balloon position on CXR. Aliquots of 50ml air to gastric balloon up to 250-300ml. Pull up to fundus and note length. Apply traction (500ml fluid). Risks: pain, haemorrhage, oesophageal or gastric rupture, pressure necrosis, airway obstruction. Max 12-24h.

28
Q

What is Pulse pressure waveform analysis? How does PiCCO work? Issues with it? What is LiDCO

A

Analysis of the arterial pressure wave in order to estimate the volume that created that wave.
Need info on aortic compliance, capacitance and vascular resistance - usually estimated from population data.

PiCCO Need proximal arterial cannula - femoral/brachial/axillary
- need reliable dichrotic notch so end of systole can be determined
- Calibrated by transpulmonary thermodilution - cold bolus into CVC + thermistor on the cannula
- Gives CO, SV, SVV, SVR, LV contractility (dP/dT)
- Can calculate VD of the cold bolus in the thorax
- intrathoracic thermal volume
- Other values - global end diastolic volume (indication of preload), intrathoracic blood volume (25% higher than GEDV), EVLW (indication of pul oedema), cardiac function indexCan use in all pts >2kg.
- Need special proximal art line, need CVC, regular recalibration, sensitive to damping, cannot produce continuous info in presence of IABP.

LiDCO - Plus:
Li (more accurate).
Calibrated via 2-4micromol/kg Li centrally or peripherally. Inaccurate if pt on Li/trac/roc. CI in early preg/pt <40kg. - Rapid. Calibrates via population data.

Intraop monitoring only.
Does not require manual calibration.

Both use an algorithm called PulseCO, based on aortic compliance data rather than the arterial waveform, so less susceptible to damping
- only need a peripheral art line.
- Can’t be used in AR or with IABP.
- FloTrac Vigileo Similar to LiDCO Rapid. Also continuously measures ScvO2. Not accurate in arrhythmia/IABP.

29
Q

Non-invasive CO monitoring

A

Bioimpedance - small current passed between electrodes on thorax; impedance varies with intrathoracic blood volume. Pulsatile component has to be isolated. Sensitive to interference from movement + other equipment. Less accurate in unstable pts. Bioreactance - NICOM. Uses phase shift as well as reduction in amplitude - better signal:noise ratio than bioimpedance. Still ?not v accurate in critical illness. Finapres/CNAP - Penaz principle. Less accurate in shock/poor peripheral perfusion.

30
Q

ECMO

A

CESAR - H1N1 Indications Reversible cardiac/respiratory failure failing conventional tx PFR<100/13.3 despite optimal ventilator strategy Resp acidosis pH<7.2Cardiogenic shock despite maximal inotropes and IABP Failure to wean from CBP Bridge to transplant/VADCardiac arrest - E-CPR Components Heater, filters, membrane oxygenator, pumpsCan incorporate haemofiltration RisksBleeding, stroke (VA), haemolysis/thrombocytopenia, exsanguination, distal ischaemia, infection, HIT, air embolism ECCO2R/Novalung - CO2 removal

31
Q

Rapid infusors

A

Level 1 - pressurises to 300mmHg
Belmont - roller pumps
Warming and rapid infusion
One bag going in while next prepared
Need reliable IV access - risk of compartment syndrome if tissues
Rate limited by bore of IV access
Risk of air embolism, fluid overload
Not recommended for plts/cryo - damage/inactivation

32
Q

HME

A

Increases resistance to airflow/WoB, adds dead space, can cause rebreathing, heavy/bulky, loses efficiency over time, can become waterlogged, no good in hypothermia or NIVIn BPF would cause excessive flow through BPF as would be path of least resistance Prevent water and heat loss, bacterial/viral filtration, prevent drowning50-80% efficient at steady state (15-20m) Achieves relative humidity of up to 70%

33
Q

Types of Mechanical circulatory support?

A

-Acute VA ECMO/mini ECMO
-IABP
-Chronic - VADs
-Impella - microaxial rotatory device, propels blood LV–> aorta (a R sided Impella also exists)
-TandemHeart - supports LV, have to have good RV
-Hemopump

34
Q

ECGs in MI

A

Hyperacute T waves
ST elevation
T wave inversion
Q waves

35
Q

Liver tests

A

Conventional SyntheticDynamic - LiMON device - indocyanine green used to demonstrate flow

36
Q

CPEX

A

9 panels VO2 max >15 ideally (effort dependent) AT <14 not good but <11 truly high risk (effort independent)

37
Q

CXR/AXR signs

A
  • Deep sulcus sign: abnormally deep and dark costophrenic angle, indictes loculated PTX - Hampton’s hump/Westermark’s sign: PE - Rigler’s sign/double wall sign - air both inside and outside of intestine = perf - Reverse bat’s wings = PJP - Sail sign/double left heart border - LLL collapse (also paeds sail sign - thymus - normal) - Veil-like opacity: LUL collapse - Single lung transplant - the transplanted lung looks wetter (lower resistance - higher hydrostatic pressure) - SCC = central, adeno = peripheral - Old TB treatments: thoracoplasty, plombage, phrenic nerve crush - Multiple tiny calcifications in lung fields - healed varicella pneumonia, silicosis, coal worker’s pneumoconiosis, miliary TB, lymphangitis carcinomatosa, radiation pneumonitis - Effusion on PA/AP CXR: at least 200ml. If obscures diaphragm: at least 500ml. Lateral CXR: at least 50ml.
38
Q

Coagulation

A

DDAVP (desmopressin) 0.3mcg/kg for trauma pts on anti plateletsHigh PT: vit K def, factor 5 def, liver failure High APTT: low fibrinogen, factor 7 def, liver failure Prolonged bleeding time: DIC, uraemia, antiplatelets, thrombocytopenia, beta lactams, NSAIDs, post CPB

39
Q

MRI

A

T1 - water darkT2 - water light FLAIR - like T2 but CSF dark so acute oedema shows up (fluid attenuated inversion recovery) DWI - detects ischaemia much earlier (diffusion weighted imaging)

40
Q

Bronchoscopy indications

A
  • Diagnostic - BAL- Bx
  • Airway assessment for burns, FB
  • Confirmation of airway device (ETT/trache/DLT/BB)
  • Therapeutic - Removal of sputum/blood/FB
  • Reinflation of collapsed segments
  • Stents
  • Assist intubation/trache
41
Q

How do you assess C-spine X-ray?

A

Lines
Anterior vertebral
Posterior vertebral
Spinolaminar
Posterior spinous

Prevertebral soft tissue < 1/3 VB width above C4<100% VB width below C4

42
Q

Post-pneumonectomy CXR

A

Look for the clips!24h: PPS contains only air. Slightly mediastinal shift towards PPS and slight elevation of hemidiaphragm.PPS fills with serosanguinous fluid at a rate of 2 rib spaces/day. If fluid level drops, suspect BPF. 2/52: 80-90% filled. 4/12: complete obliteration. 6-8/12 maximal shift of mediastinum.

43
Q

Synacthen test

A

Primary/Addison’s - no response as problem is in adrenals Secondary/pituitary failure - low baseline, exaggerated response Tertiary/hypothalamic failure

44
Q

High cuff pressure

A

Tube too small for airway Paw higher than cuff pressure will cause leakCuff leak/faulty device Poorly positioned device

45
Q

What is TEG?

A

Rapid point-of-care tests of the viscoelastic properties of clot formation of whole blood

46
Q

What are is one of the issues of TEG?

A

They can’t detect hypothermia or hypocalcaemia as causes of coagulopathy

47
Q

What is R Time

A
  • Time from when test starts, to when clot starts to form
  • Normal is 5-10mins
  • Determined by clotting factors
  • If prolonged, give FFP
48
Q

What is the K Time

A
  • The time from the beginning of the clot forming, until the width or deviation of the graph is 20mm on the screen
  • Normal is 3-5mins
  • Determined by fibrinogen
  • If prolonged, give cryoprecipitate
49
Q

What is the Alpha Angle

A
  • This tells you the rate at which the clot is forming
  • Normal is around 50-70 degrees
  • Determined again by fibrinogen
  • If lower then give cryoprecipitate
50
Q

What is the Maximum Amplitude?

A
  • A measure of how big the clot gets
  • Normal is 50-70mm
  • Determined largely by platelets
  • Unsurprisingly - Give platelets if low
51
Q

What is CL30 or LY30

A
  • Lysis at 30mins
  • A measure of how rapidly the clot is breaking down
  • Should be less than 8% by 30 minutes after maximum amplitude
  • If more than this, give tranexamic acid