FLIR R425 Flashcards
What kind of battery does the FLIR R425 use?
Internal lithium ion- runs up to 12 hrs when fully charged( has chamber for back up batt)
After turning on the unit how long does it take to get to a “ready state”?
< 20 sec, reads 5-15 ųR
How do you set background reading?
Press Left button for approximately 10 sec
How do you get back to the main screen
Push dose rate
How long will the meter sample for after pushing “identify”?
At least 30 sec (can be extended if needed)
Where does the meter detect from?
Front and side of the detector when held flat
The screen displays a confidence number between what range?
0-10
What else will the screen display besides the confidence number?
Isotope and source type.
(NORM, INDUSTRIAL, MEDICAL, or SNM [Special Nuclear Material]).
If a Red light/flash mean?
Gamma
What does Blue light/Flash mean?
Neutron
True/False the FLIR is fully deconable?
True