Equipment Maintenance and Calibration 1% Flashcards

1
Q

Any equipment with a ________you should use all day without recharging you between use

A

rechargeable battery

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Why shouldn’t rechargeable batteries be charged between each use?

A

Because rechargeable batteries have a memory and if you constantly recharge the battery it never will develop much of a memory.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

You should keep a record book of your ________(can be as simple as a spiral notebook)

A

equipment maintenance and calibration checks(should include serial number of the equipment because equipment gets moved from room to room)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

____instruments when not in use.

A

Cover or store(in a container) and turn off

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Why is it important to cover or store instruments?

A

Because many of the instruments have mirrors and they are dust magnets.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Anything that makes contact with a patient should be….

A

cleaned, disinfected, or sterilize depending on how it comes into contact with the pt.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Equipment that is painted metal, plastic, or vinyl should be…

A

dusted with a soft cloth
dry cloth or with spray cleaner(if spray cleaner, spray on cloth not instrument) and use all solutions sparingly

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Crevices can be cleaned with ….

A

canned air or cotton tip applicators

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Why must you not tilt canned air?

A

Because a cold moist liquid will come out instead of air and this will make plastics milky white

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Why is it important to turn equipment off in between use?

A

Because they get hot and they need to cool down

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

When changing bulbs…..

A

Turn off equipment, unplug, and wait for bulb to cool, handle it with a bulb jacket or tissue.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

If bulbs have corrosion….

A

unplug equipment and scrape off with eraser or file, then blow it out.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Why should you not touch the bulbs?

A

because oil from fingers get on it and shorten the life of the bulb and create shadows in the illumination field and it also protects you in case the bulb breaks

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Some bulbs don’t blow they instead _________

A

diminish in illumination

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Some bulbs can be rotated ____degrees to extend their life
(The Goldman perimeter is one that has a bulb that should be rotated periodically)

A

180

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

If you have a shadow or decrease in you illumination field for projector you should…..

A

-check the bulb(may be over heated, may not be rotated properly, may need replaced, contact may be corroded)
-maybe housing mirror is out of alignment
-maybe the lens or optics in it need cleaned

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

The housing mirror in the projector is adjacent to the ____and is easy to knock out of position

A

bulb

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

What is the formula for calibrating a projector?

A

RF distance in ft. to screen(the distance from pt eye to the mirror to the silver screen behind them) x 4.5mm=height of 20/200 letter in mm on reflecting screen

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

Other than formula what can be used to calibrate a projector and how is it done?

A

“a chart bracket”
( a piece of paper that you can get from the equipment manufacturer)
first measure RF in feet, then take the chart bracket and hold on the silver screen, a second person is at the projector moving the projector tube in and out to change the size of the 20/200 letter and to change the focus of it. If RF is 16’ your 20/200 letter should be the size of the preprinted E. If RF is 14’ yours 20/200 should be the size of the bracket just above the number 14

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

Calibrating projector:
20’ chart should be ____from pt(infinity)
10’ chart should be ___from pt

A

20’
10’

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

Calibrating a projector;
If room is not 20’ then you use 2___surface mirrors or a screen

A

front

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

Calibrating a projector:
Measure distance between pt and ______

A

reflecting screen

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

What should you clean projector slide with?

A

soft dry cloth
(do not touch, never use any liquid)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

Projector slides have___glass plates with the symbols between them

A

2

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
If you use liquid on the projector slides it will....
distort and ruin the symbols
26
We use ___surface mirrors to check VA because they reflect 100% of the light.
front
27
Front surface mirrors are called this because the ....
mirror is literally on the front
28
Frequent wiping will____the reflectivity so you only clean them when they are ______
erode visibly dirty
29
What do you clean front surface mirrors with?
canned air, a camel hair brush, or a 1:4 soln of ether dragged across a lint free cloth.
30
How can you tell if a mirror is a front surface mirror?
Pencil tip touches its reflection
31
How can you tell if a mirror is a back surface mirror?
It will show some space between the pencil and its reflection
32
How do you clean back surface mirrors
any household cleaner(just like mirrors at home)
33
Frequent rubbing or chemicals can on trial lenses will...
cause the painted numbers to rub off
34
How should you clean trial lenses?
With wens wipes or lint free cloth or glass cleaner
35
You should keep trial frames covered to keep them__
dust free
36
Be careful not to drop or jar trial frames because they can...
come out of alignment easily.
37
What should you use to clean trial frames?
alcohol or damp cloth
38
Cross cylinders should be cleaned with.....
Damp cloth
39
Vigorous rubbing on the cross cylinder can cause .... Harsh cleaners can...
loosened lenses and rotate lens rub off dots.
40
____are special lenses that have and antireflective coating on them
bloomed lenses
41
Phoropter lenses are ___
bloomed
42
How do you clean bloomed lenses (phoropter lenses?
with 100% ethyl or methyl alcohol on lens wipe (never wipe dry)
43
Glass lenses lenses should be cleaned with
glass cleaner
44
Plastic should be cleaned with
contact lens cleaner
45
Most prism bars or loose prislms are made of ____
acrylic
46
What should you clean prism bar or loose prism with?
soft lint free cloth (keep them in a case)
47
Why should you not use alcohol or other solutions on prism bar or loose prisms?
because they will fog lenses and turn them white
48
Why should you store unused phoropter at zero and open and?
Because it protects the lenses in the body of the unit. Should also cover with dust cover at night.
49
You should ____ never touch the lenses of the phoropter
never
50
What should you clean the face shield of the phoropter with?
alcohol, soap and water, or sterilize in autoclave
51
Phoropters should be professionally cleaned every ____ or you can do it yourself.
1-2 years
52
Be careful not to get any ___ in auto refractor.
liquid
53
If the measuring window of the auto refractor gets dusty it can...
affect the accuracy of findings
54
Auto refractor measuring lens should be cleaned with...
canned air or lens wipe
55
You can clean forehead rest, chin rest, and face shield of auto refractor with...
alcohol
56
Auto refractor is ___voltage and should be treated as other computers, do not ____. Make sure thermal paper is installed ____
disassemble correct side up
57
The retinoscope, transilluminator, and direct ophthalmoscope are all maintained by...
Leave out of charger during daily use to discharge battery Turn off after use Bulbs will blow if transformer is set too high
58
You should keep the ophthalmoscope viewing lens set at ____to keep lenses clean
zero
59
How do you get the bulb in and out of a Copeland?
push/pull lamp, do not twist-will ruin base
60
With Welch Allyn the bulb____ and you remove by ___
slides into head gently flipping with screwdriver
61
How do you clean the lensometer?
with canned air or camel hair brush and eye piece with lens cloth
62
Must ___lensmeter in between use
turn off
63
Marking device can get gunked up with ink, clean with
and alcohol pad
64
____may need reinked or replaced
ink pad
65
How do you check to see if lensmeter is in calibration?
by reading some trial lenses from one of trial sets
66
If lesmeter needs calibrated it should be done by____
returned to factory for service
67
Automatic lensometer is a ____processor
micro
68
If fluid gets into the auto lensometer it will...
damage it
69
___device on auto lensometer may need replaced or reinked
marking device
70
Marking pens on auto lensometer can be cleaned with
alcohol wipe
71
Replace nose piece cover of auto lensmeter if.. (pry loose with a screwdriver and attach new piece)
it is torn or abraded
72
Check calibration of auto lensmeter with ___ and sent to ___if needs recalibrated
trial lenses factory
73
Lens of indirect ophthalmoscope should be cleaned with..
canned air or alcohol
74
Indirect ophthalmoscope has a __surface mirror
front
75
Band on the head of indirect ophthalmoscope can be cleaned with...
soap and water or alcohol
76
Indirect ophthalmoscope should be stored by...
hanging on hanger or in a case
77
Where is the bulb located in the indirect ophthalmoscope
in the bulb connector next to the core connector
78
facts about Bulb of indirect ophthalmoscope....
-located in bulb connector next to cord connector -pull cord connector out and unscrew bulb connector -bulb will fall out -check bulb contacts and clean if necessary
79
slit lamp should be ___when it is not in use
locked and turned off
80
if chin rest of slit lamp is made of plastic it can be cleaned with ___ but if is made of rubber don't use because it will turn it white
alcohol
81
Slit lamp should be cleaned with..
canned air, camel hair brush, or dust with soft lint free cloth
82
Slit lamp mirror is a ___surface mirror. To clean grasp both sides of the mirror and gently____ and clean with a _____
front pull straight out lens wipe
83
The gliding plate of the slit lamp needs its ball bearing tracks cleaned ____
1-2x per year with alcohol
84
Slit lamp: Every __cleanings of he gliding plate you should lubricate it with ___ to condition it and prevent cracking and keep SL moving smoothly.
5 WD-40
85
How do you extend the life of slit lamp bulb?
by using lowest voltage setting and turning off between use
86
The corneal topographer cone is painted with a___ . It absorbs the oils in your fingers and will scratch easily so keep hands off and keep covered with a dust cover.
special white reflected paint
87
____dust corneal topographer cone but never use _____. Clean chin and forehead rest with____
Lightly alcohol or harsh detergents alcohol
88
When should Topography unit be calibrated?
When it is initially set up, when moved, and at least monthly
89
How do you check calibration of topography unit?
with a calibration object mounted and then capture a series of images(unit will advise if calibration needed)
90
Topography unit: verify and recalibrate every __years by ___ or ____ or ____
q 2 years manufacturer or certified measuring lab or according to manufacturer's instructions
91
Biometry instrument calibration should be checked___
daily before using on pts
92
To calibrate biometry instruments...
use a clean test eye, perform all the tests, ensure results are within tolerance range given for test eye, if not take instrument out of service and contact manufacturer's service dept
93
How should front surface mirror that surrounds the bulb of keratometer be cleaned?
with canned air or camel hair brush
94
Sliding pate of Keratometer requires no maintenance because...
it has permanent grease applied at factory
95
What maintenance does keratometer require?
cleaning mirror, changing bulb, and calibration
96
What is used to check calibration of keratometer?
3 steel balls with a known radii
97
Explain how to calibrate keratometer?
Mount test sphere on magnetic holder, focus eye piece, measure them.
98
How often should calibration be checked with keratometer?
every 2 mos
99
If readings aren't accurate when checking calibration of keratometer....
send for recalibration
100
What should pachymeter probe be cleaned with
alcohol between pts or gas sterilize it with ethylene oxide (EO gas), or used chemicals (cold sterilization) but do not autoclave
101
Be careful with probe of pachymeter-do not jar or drop it and store in...
a case
102
It stimulus on automated visual field machine had blurred edges you need to ....
send for servicwe
103
Disk drives of automated visual field machine needs to be cleaned at least___
3 x per year with a drive cleaner moistened with alcohol
104
Tape heads of the automated visual field machine needs to be....
cleaned periodically with tape cartridge that has cleaning pad, cleaning solution and arm for moving cleaning pad across the head. Apply a few drops of cleaning soln to pad and insert into drive, move handle up and down to clean entire head surface, resume normal operation
105
Bowl type perimeters have a special ___ This absorbs oil from fingers and scratches easily.
white reflectance paint
106
How do you clean a bowl type perimeter
with canned air, camel hair brush, or Extran MA O2 Neutral (soln form MERCK)
107
The ____ perimeter has a projector arm with 2 ball joints that need lubricated__x per year
Goldmann perimeter 2
108
When replacing the lamp housing of goldmann perimeter make sure you align the ___
red marks
109
Clean lamp housing of goldmann perimeter with...
canned air or remove and clean with soft cloth
110
Bulb of Goldmann perimeter must be roatated .....
180 degrees q 1 wk or after 20-30 hrs of use.
111
When should you replace the bulb and calibrated the Golmann perimeter?
Replace bulb when neither side achieves 1000 asb and recalibrate machine after replacing bulb
112
A ___is used to check the calibration of a Goldmann tonometer
Calibration bar
113
Goldmann tonometer is calibrated at ____ and should be logged in log book with date, tonometer serial number and whether the tonometer was calibrated
0, 20, 60 have a set routine to check such as every 3 mos
114
How many marks are on the Goldmann tonometer calibration bar and what do each of them represent?
Each has 5 markings which correspond to the 3 pressures at which you check calibration middle mark = 0 marking on either side of"0" =20 Outer most markings =60
115
To calibrate tonometer: If tonometer point up... If tonometer hang down....
-place the longer end of the calibration bar toward the examiner -Place the longer end of the calibration bar toward the pt
116
Explain how to check calibration of Goldmann tonometer
put tonometry probe into position as if going to measure eye pressure, now move the tonometer knob and note the pressure at which the tonometer tip tilts forward. If it tilts when the knob is dialed to 60 the tonometer is correctly calibrated. Repeat for 0 and 20 and move calibration bar accordingly
117
If tonometer is not in calibration at 0, 20, and 60 then....
check and see if tonometer prism properly inserted, see if calibration bare is precisely adjusted, repeat. If repeat check still shows errors-take tonometer out of service and send for recalibration.