chapter 11 EXAM 5 Flashcards
safety requirements are required for:
tube housing xray control panel radiographic exam table source to image distance indicator other devices and accessories
xray tube housing requirements
lead lined metal tube housing to protect patient and tech from leakage, scatter and off focus radiation
leakage radiation
radiation that does not exit form the collimator opening but penetrates the protective tube housing and the sides of the collimator
primary beam (direct radiation, useful beam)
radiation that emerges directly from the xray tube and moves without deflection toward a wall, door, etc
the tube housing is constructed so that leakage radiation measured at a distance of what from the xray source does not exceed what when the tube is at it’s highest mAs/Kv
1m from the xray source does not exceed 1 mGya/hr
the tube housing also does what
it confines high voltage to prevent electric shock
makes cooling of xray tube possible
control panel/console requirements
must indicate the conditions of exposure and indicate when xray tube is energized
usually satisfied with use of kVp and mAs
radiographic exam table requirements
must be strong and support the pt.
thickness must be uniform and radiolucent to absorb only a minimal amount of radiation
what is the xray table typically made of
carbon fiber material
source to image receptor distance indicator requirements
can be a tape measure or a laser
distance and centering indicators
distance and centering indicators must be accurate to within what of the SID
within 2% and 1% of the SID
t or f
under no circumstances should the xray beam exceed the size of the image receptor
true
collimation is accomplished primarily through the use of
a light localized, variable aperture collimator
other types of collimators include
aperture diaphragms
cones and cylinders
all collimation devices reduce what and improve what
scattered radiation and improves image quality
light localizing variable aperture rectangular collimators have 2 sets of adjustable lead shutters they do what
- reduces the amount of off focus radiaiton
2. consists of 2 pairs that can be adjusted independently
light localizing variable aperture rectangular collimators have
a light source and a mirror
skin sparing :
to minimize skin exposure to electrons produced by photon interaction with the collimator, the patient’s skin surface should be at least
15 cm below the collimator
space bars
some collimators contain space bars to prevent collimators from being closer than 15 cm to the pt.
what is luminance
the brightness and quantifies the intensity of a light source
brightness is measured in
nit or candela per square meter
a reading of 15 foot candles corresponds to a collimator light source with a lumincance of about
161 nit or 161 candela per square meter
the xray beam and the light beam must coincide to within what of the SID
within 2% of the SID
Positive Beam Limitation PBL
ensures that the radiographic beam is no larger than the IR