chapter 11 EXAM 5 Flashcards
safety requirements are required for:
tube housing xray control panel radiographic exam table source to image distance indicator other devices and accessories
xray tube housing requirements
lead lined metal tube housing to protect patient and tech from leakage, scatter and off focus radiation
leakage radiation
radiation that does not exit form the collimator opening but penetrates the protective tube housing and the sides of the collimator
primary beam (direct radiation, useful beam)
radiation that emerges directly from the xray tube and moves without deflection toward a wall, door, etc
the tube housing is constructed so that leakage radiation measured at a distance of what from the xray source does not exceed what when the tube is at it’s highest mAs/Kv
1m from the xray source does not exceed 1 mGya/hr
the tube housing also does what
it confines high voltage to prevent electric shock
makes cooling of xray tube possible
control panel/console requirements
must indicate the conditions of exposure and indicate when xray tube is energized
usually satisfied with use of kVp and mAs
radiographic exam table requirements
must be strong and support the pt.
thickness must be uniform and radiolucent to absorb only a minimal amount of radiation
what is the xray table typically made of
carbon fiber material
source to image receptor distance indicator requirements
can be a tape measure or a laser
distance and centering indicators
distance and centering indicators must be accurate to within what of the SID
within 2% and 1% of the SID
t or f
under no circumstances should the xray beam exceed the size of the image receptor
true
collimation is accomplished primarily through the use of
a light localized, variable aperture collimator
other types of collimators include
aperture diaphragms
cones and cylinders
all collimation devices reduce what and improve what
scattered radiation and improves image quality
light localizing variable aperture rectangular collimators have 2 sets of adjustable lead shutters they do what
- reduces the amount of off focus radiaiton
2. consists of 2 pairs that can be adjusted independently
light localizing variable aperture rectangular collimators have
a light source and a mirror
skin sparing :
to minimize skin exposure to electrons produced by photon interaction with the collimator, the patient’s skin surface should be at least
15 cm below the collimator
space bars
some collimators contain space bars to prevent collimators from being closer than 15 cm to the pt.
what is luminance
the brightness and quantifies the intensity of a light source
brightness is measured in
nit or candela per square meter
a reading of 15 foot candles corresponds to a collimator light source with a lumincance of about
161 nit or 161 candela per square meter
the xray beam and the light beam must coincide to within what of the SID
within 2% of the SID
Positive Beam Limitation PBL
ensures that the radiographic beam is no larger than the IR
PBL was required on all xray systems in the US between the years of
1974 and 1994
not required anymore
state regulartoy standards require accuracy of
2% of the SID with PBL
some state may require only 3%
aperture diaphragm
simplest of all beam limiting devices
rectangular shape is most common and can be used in many systems
cones and cylinders
modifications of the aperture diaphragm
restricts the useful beam to the required size
reduce scatter and improve contrast resolution
t or f
sharper size restriction is achieved when the cone or cylinder is longer
true
filtration (also called hardening of the xray beam)
reduces exposure to patient’s skin and superfical tissue by absorbing most of the lower energy photons
increases the quality of the beam
2 types of filtration
- inherent filtration
2. added filtration
examples of inherent filtration
glass envelope
insulating oil
glass window
the inherent filtration material amounts to about what aluminum equivalent
0.5mm aluminum equivalent
the collimator provides an additional inherent filtration of what aluminum equivalent
0.1mm aluminum equivalent
added filtration
sheets of aluminum or the equivalent that are added to the tube
the inherent filtration and added filtration combine to equal
the required amount necessary to filter the useful beam
does the kVp of an xray unit determine the amount of filtration required?
yes
the regulatory standard for total filtration of 2.5 mm Aluminum equivalent is required for fixed xray unit operating above
70 kVp
because each xray tube and collimator has a total inherent filtration of 1.5mm aluminum equivalent, the manufacturer place and additional
1mm Al Eq filter between the tube housing and collimator to meet the minimum regulatory requirement
stationary radiographic equipment operating at 50-70 kVp require a total filtration of
1.5 mm Al Eq