Chapter 10: Quality Control (Handout) Flashcards
a process to periodically examine a measurement procedure to verify that it is performing according to pre-established specifications
Quality Control
Dispersion of results for repeated measurements of aliquots of the same QC material
Imprecision
measure of expected imprecision in a measurement procedure when it is performing correctly
Standard Deviation
Interval of ±1SD
68% of measured values
Interval of ±2SD
95% of measured values
Eliminates systematic bias
Correct Calibration
Difference between the observed mean and the expected value for a QC material
- Brought about by changes in calibration
Systematic Bias
- the combination of systematic bias and imprecision that occurred for that specific measurement
Accuracy
refer to an average systematic bias that may be present in a given method
Trueness
most often performed by the laboratory using calibrator materials provided by the method or instrument manufacturer
Method calibration
methods are calibrated during the manufacturing process, and the laboratory performs a verification of that calibration
(E.g. Point-of-care devices)
frequently specify calibration intervals
In vitro device (IVD) manufacturers
Calibration of routine methods should be traceable to a higher-order reference measurement procedure or international reference material
Calibration Traceability to a Reference System and Commutability Considerations
- evaluates a measurement procedure by periodically assaying a QC sample for which the expected result is known in advance.
- part of the process management component of the quality system that integrates good laboratory practices to ensure correct patient results.
Statistical QC
Most common presentation for evaluating QC results
Levey- Jennings or Shewhart Plot