Chapter 10: Quality Control Flashcards
- Also known as “Internal Quality Control or Statistical Process Control”
- is a process to periodically examine a measurement procedure to verify that it is performing according to pre-established specifications.
Quality Control
Clinical Laboratory Test
- To evaluate the pathophysiologic condition of an individual patient
- To assist with diagnosis
- To guide or monitor therapy
- To assess risk for a disease or for the progression of a disease.
Factors that contribute to TOTAL ERROR
- Pre-analytic variability
- Analytic variability
- Interfering substances
sample collection, transportation, processing, and storage
Pre-analytic variability
test performance
Analytic variability
drugs or metabolic components
Interfering substances
- Called “QC sample materials”
- are measured periodically in the same manner as clinical samples, and their results are examined to determine that the measurement procedure meets performance requirements appropriate for patient care.
Surrogate samples
- Dispersion of results
- SD is a measure of expected imprecision in a measurement procedure when it is performing correctly
Imprecision
- is the difference between the observed mean and the expected value for a QC material.
- If the calibration changes for any reason, a systematic bias is introduced
Systematic Bias
a method which eliminates systematic bias (within uncertainty limits)
-Correct calibration
- Closeness of measured value to a standard or known value
Accuracy
- Refers to the closeness of two or more measurements to each other, regardless of whether those measurements are close to the known value
Precision
Primary Purpose of Measuring QC Samples
- statistically evaluate the measurement process to verify that the method continues to perform within the specifications consistent with acceptable systematic bias and imprecision
- identify that a change in performance occurred that needs to be corrected.
The performance of a new method can be assessed for accuracy by _____
Assaying patient specimens or interlaboratory survey materials with known values
- Used for individual results and is a combination of systematic bias and imprecision that occurred for that specific measurement.
Assaying Patient Specimens
used to refer to an average systematic bias that may be present in a given method
Trueness
most often performed by the laboratory using calibrator materials provided by the method or instrument manufacturer
Calibration
require calibration or calibration verification at least every 6 months, or more frequently if recommended by the method manufacturer
Clinical Laboratory Improvement Act (CLIA) regulations, Section 493.1255
- component of a quality management system
- part of the process management component of the quality system that integrates good laboratory practices to ensure correct patient results.
Statistical Quality Control
- required for all aspects of laboratory operation, including statistical quality control
Standard Operating Procedures (SOPs)
If the QC result is within acceptable limits of the known value
the measurement procedure is verified to be stable and results for patient samples can be reported
If a QC result is not within acceptable limits
the measurement procedure is not performing correctly, results for patient samples are not reported, and corrective action is necessary
- also called Shewhart plot (Shewhart, 1931)
- most common presentation for evaluating QC results
Levey-Jennings (Levey, 1950)
±1 SD
68.3% of observations
±2 SD
95.4% of observations
±3 SD
99.7% of observations
QC materials should be selected to
Provide analyte concentrations that monitor the analytic measurement range of the method
Common response of quantitative measurement procedures
Linear Response
In the case of non-linear method response, it may be necessary to
Use additional controls at intermediate concentrations
The QC materials selected must be
Manufactured to provide a stable product that can be used for an extended time period, preferably 1 or more years for stable analytes.
Use of a single lot for an extended period allows
reliable interpretive criteria to be established that will permit efficient identification of an assay problem, avoid false alerts due to poorly defined expected ranges for the QC results and minimize limitations in interpreting values following reagent and calibrator lot changes
Limitations of Quality Control Materials
o Non-commutable
o Deterioration of the analyte during storage
o Multi-constituent control materials caused by solubility considerations or potential interactions between different constituents, particularly at higher concentrations.
Several Parameter for the Frequency to Assay QC Samples
- Analytic stability of the measurement procedure
- Risk of harm to a patient from clinical action being taken before a significant error is detected
- Number of patient results produced in a period of time when an error condition existed but was not yet detected
- Events such as recalibration or maintenance that may alter the current performance condition of the measurement system
- Training and competency of the test operator, particularly for manual or semi-automated methods
- Risk of failure of the measuring device
The more stable the system
The less frequently a statistical QC evaluation needs to be performed