CE Flashcards

1
Q

CE What is it used for?

A

Rx Sample

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

CE What is the impurity?

A

T-3003

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

CE volage

A

30k watts

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

CE lamp calibration is done how often, how is it done?

A

Monthly, .Page 13 Agilent lab advisor connect (right upper side), services and diagnostics, DAD, detector test (holmium, intensity) DO ONE AT A TIME, start run, print results write it up in logbook

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

CE sample diluent ACN prep

A

Molecular sieves, sit for 1-day lasts a month, filter with 0.45 mm into NIR vial, 320g per 4L of acn

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

CE lamp details hwo to turn on

A

Needs 30 min to equilibrate, control panel, launch (online)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

CE SOP IOP number

A

DS-SOP-11311 page 5 vials

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

CE NaOH concentration details

A

Rxn method 0.1N, preconditioning 1N

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

CE 3 methods type

A

Preconditioning, rxn method, shutdown

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

CE it is a __________ system

A

Pressure we need a specific volume, full 600 ml , 300 ml we use the 20-300 ml micropipette

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

CE BE SLN is use to measure the

A

Background

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

CE method is the ____, sequence is the ___

A

How, what

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

CE how to run the preconditioning

A

Run control, sample info, empty vial location

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

CE TIP when you change your method you will know because

A

It will change your run time

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

CE run test injection

A
1- change method 
2- run control
3-sample info 
4- vial 2 
5- sample name: SLN - # 
6-fill comment, run method
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

CE review sequence method buttons

A

-

17
Q

CE sequence up tab

A

Save sequence template as,

Sequence table to show table

18
Q

CE run control, how to run sequence

A

Run sequence make sure everything (method, sequence) is good

19
Q

CE how to lock

A

Up tab user lock session - non privately

20
Q

CE Zach training remember that before applying anything to the chromes you need to select all what was his mistake

A

Don’t hit overlay select just to the left of that

21
Q

CE how to take out the capillary cartridge

A

Tilt to the front until you hear a click take it out.

To put another one don’t click it back leave it tilted. Remember to adjust the right side of the capillary with your pinkie.

22
Q

CE Remember to u—d —ls before turning off

A

Unload vials

23
Q

CE peak tailing should always be

A

.6 if it’s not that it might indicate a problem

24
Q

CE how to put the injections through a report

A
  • you have to highlight every injection before going to the report tab
  • report templates
  • there are two reports
  • print both