AU680 Flashcards

1
Q

Used to process priority stat samples and the fastest way to process a sample.

A

STAT table

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

the sample probe/liquid level detector dispenses samples and or diluent, and detects liquid level. Sample is aspirated from the tube/cup and dispensed into the cuvette wheel

A

Sample transfer unit

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Two reagents probe/liquid level detectors dispense reagent and or diluent, and detect liquid level. The probe assemblies are designed for up and down collision detection. Reagent is aspirated from R1 and R2 compartment and dispensed into the cuvette wheel.

A

Reagent transfer unit

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Mix bars used to mix the reagent and sample in the cuvette

A

Mixing unit

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Consist of a halogen lamp, lenses, a diffraction grafting, and a photodetector to measure the amount of light transmitted through the reaction solution in the cuvette

A

Photometry unit

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Cleans, rinses, and dries the cuvettes automatically

A

wash nozzle unit

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Define ISE unit

A

Diluted samples is passed through sodium, potassium, and chloride ion selective electrode to determine the concentration by comparing the electrical potential different to the reference electrode.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

The screen to check system messages relating to sample status and analyzer status.

A

Home outline

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Check analyzer status menu: Blue

A

no error

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Check analyzer status menu: yellow or orange

A

non fatal error. the analyzer can be started

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Check analyzer status menu: Red

A

Fatal error. The analyzer can’t be started

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

3 important sample requirements for the instrument to sample correctly.

A
  • should not contain fibrous material or fibrin, except for WB.
  • No air bubbles
  • Dispense sample volume in the quantity required for analysis plus the dead volume
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Rack colors: white

A

used for analyzing routine samples

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Rack colors: Blue

A

used for calculation of reagent blanks for creation of calibration curves

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Rack colors- yellow

A

used for creating of calibration curves

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Rack colors- green

A

used for analysis of QC samples

17
Q

Rack colors- orange

A

used for manual repeat run analysis

18
Q

Rack colors- red

A

used for emergency sample analysis

19
Q

Daily variation chart for QC- green

A

normal data

20
Q

Daily variation chart for QC- yellow

A

data outside the range of the single check level programmed in QC specific

21
Q

Daily variation chart for QC- orange

A

abnormal data (data not included in the QC statistical values)

22
Q

What colors do the operator observe on the analyzer maintenance screen if the analyzer requires maintenance and what are the meanings of the colors?

A

Orange- over due
Yellow- expiring
The colors for daily maintenance appear 3 hours before the maintenance check.

23
Q

List 3 reasons that teh instrument may not be able to calculate a result

A
  • mechanical error has occurred
  • outside fo cuvette walls or the cuvette wheel is wet
  • in the rate rxn, fewer than 3 photometric readings statisfy the assay creteria specified in teh specific test parameters
24
Q

List the first possible causes and corrective actiosn to be taken for the errors

A

sample can be severly lipemic, icteric, hemolytic or can contain excessively large amount of the analyte being tests.
dilute the sampel and run the test again.

25
Q

What phase of an enzymatic reaction has been reached if hte not enough data to determine linearity of reaction error message ahs displayed?

A

Reaction might go into substrate depletion due to a high concnetration or a problem with the condition of the reagent.

26
Q

What causes the reuslt (OD) is lower than the dynamic range error?

A

no concentration could be calculated. the OD fo the sample is lower than teh OD of the lwo limit of the dynamic range.

27
Q

What are three corrective actions for the errors in result (OD) is lower than the dynamic range error?

A

review the resul in the clinical context of the patient and repeat if necessary
check the reagent probes and reagent bottles for proper position
check the reagents for bubbles.

28
Q

what is the corrective action if the error result is higher than the dynamic range occurs?

A

Dilute the sample with the appropriate sample diluent and re-analyze
should be diluted so that they yield a value in the middle of the measuring range

29
Q

List 2 common causes of sample related problems with data

A
  • sample evaportation causes unusally high resutls

- incorrect sample handling