TPS and LINAC QA Flashcards
Important Features of a TPS
- Dose calculation accuracy
- Dose calculation algorithm
- Geometric accuracy
- Representation of the treatment beam
- Characteristics of CT scanner used changes the electron density to CT number graphs
- CT orientation cube
- DVH, scorecard, isodose lines represented correctly
- Connectivity to CT
- Connectivity to R&V system
TPS Commissioning Process
- Network integration
- Data transfer
- System setup and machine-source configuration
- Patient anatomical representation
- External beam commissioning
- Plan evaluation tools
- Plan output and data transfer (report)
- Overall clinical test
List of TPS QA Tests
- Back-up and recovery
- CT data transfer – demographics, patient orientation, CT HU
- CT density and geometry
- Patient anatomy
- External beam revalidation (using previously calculated plans – recalc to check still the same)
- Monitor Units Check (using previously calculated plans) (update upgraded calc time by farming out to many cores and adding up at the end which resulted in rounding errors)
- Plan Transfer
- Patient Specific Quality Assurance (PSQA)
Most Common PSQA equipment
- matrixX
- Delta4
- ArcCheck
- Octavius
- MapCheck
- EPID
- PerFraction
Factors to Consider when choosing a device for PSQA
- Does it record in Absolute or Relative Dose
- Treatment technique (IMRT, VMAT, FFF)
o Can it handle FFF dose rate? - Resolution
o Small fields will require smaller detector spacing - Gamma Pass Criteria
- Cost – Upfront and Ongoing
- Compatibility with existing equipment (TPS, Linacs etc.)
What is EPIQA
- EPIQA is a commercial software that converts a dosimetric image acquired by an EPID into a dose map and compare it with a reference dose distribution from the TPS.
- EPIQA is a pre-treatment verification tool – it is not capable of in vivo dose verification.
- It can identify potential errors in the calculation of the dose delivery process i.e. MLC not performing
- Specifically developed for Varian amorphous silicone a-Si500 EPID
- Software can be used for verification of static and dynamically modulated fields
- You can perform comparison between EPID vs EPID, or TPS vs TPS as well
o Useful for QA of the EPID/EPIQA software
Common Causes for QA Failure
- The Plan
- Detector Limitations
- Treatment Equipment Failure
Common Causes for QA Failure: The Plan
- Low MU efficiency (too many MU)
- High modulation factor (MU/cGy)
o Daily prescribed dose (i.e. 2Gy) should equal 200 MU
o At 100cm SSD, 1Gy = 100MU for 10x10cm field
o Modulation Factor = if 2Gy = 200MU; so if you need 800MU to deliver plan = 800/200 = 4MF (3MF is the standard for most VMAT plans) - Plan complexity (e.g., High complexity fields)
o High MU/Gy (typically a MF of 3 produces a good VMAT plan)
o Narrow MLC aperture
o Higher uncertainty during delivery
o Apertures smaller than minimum calibrated field for EPIQA
Common Causes for QA Failure: Detector Limitations
Example: highly elongated, high modulated spinal fields
- Solution: commission a beam model specifically for spinal fields as a typical/standard beam model is optimised for “square fields” with low to moderate shielding
o Specific beam model introduced a 1cm minimum leaf gap - Inherent build-up of the Varian EPID is 8mm but images are converted to dose at dmax
o (1.5 cm 6 MV, 2.1 cm at 10 MV)
o Algorithm converts dose measured at 0.8cm (Inherent build up of the Varian EPID) to an equivalent radiological depth of dmax of beam
Common Causes for QA Failure: Treatment Equipment Failure
- MLC failure
- Beam symmetry
- Alignment of optical system to radiation isocentre
Plan Complexity Metrics
Small aperture score
Modulation factor
Modulation Complexity Score
Area/perimeter ratio (low = complex)
When is in-vivo dosimetry done?
- Pacemakers/ICD
- Pregnant patients
- Lens dose
- TBI
What detectors are used for in vivo dosimetry?
- Film, TLD, Alanine and OSLD
Daily Linac QA Checks
Xray output constancy
electron output constancy (weekly)
Laser localisation
distance indicator (ODI) at isocentre (SSD)
Field Light
Door interlocks
AV monitors
Beam on indicator
Beam hold and Beam off
Backup MU Counter
SGRT QA
LINAC Daily QA: Xray Output Constancy: Rationale, Action Level, and Potential QA Tools
Rationale:
Ensure dosimetric treatment accuracy
Action Level:
3% variance
Potential QA Tools
* Constancy meter
* Daily QA 3
* BeamChecker
* MPC Phantom
LINAC Daily QA: Xray Output Constancy: Reasons for Failure and Potential Errors
Reasons for Failure:
* Electronic Drift of Detector
* Damage/Change to the MU Chamber (requiring replacement or recalibration)
Potential Errors:
* Incorrect dose delivered to all patients
LINAC Daily QA: Laser Localisation: Rationale, Action Level, and Potential QA Tools
Rationale:
* Treatment set-up accuracy to ensure accurate tumour localisation for optimal outcome
Action Level:
* 2mm variance
Potential Tools:
* Cube Phantom
* Quasar Penta Guide and Tilt Plate