The Hemodialysis Delivery System Flashcards

1
Q

Conductivity and pH alarm causes?

A

Incorrect dialysate bath hooked up than what is programmed in machine.
Empty jug.
Improperly calibrated phoenix meter will give false results.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Interventions for Conductivity and pH alarm?

A

Change to correct bath on machine, get new jug, recalibrate phoenix meter.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

At what temperature will blood hemolyze?

A

42 degrees celsius

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Temp alarm causes and interventions?

A

Dialysate temp too high or too low.

Check temp. manually to verify accuracy of alarm and adjust it on machine.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Blood leak detector alarm causes and interventions?

A

Break in compartment. Air Bubbles (false alarm)

Verify if its a real alarm. Test dialysate w. strip to detect any blood. If alarm is real clamp all lines and don’t return blood.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

TMP alarm causes and interventions?

A

Change in venous pressure (clotting in drip chamber, needle loose)

Troubleshoot machine, readjust needle etc.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Intervention for power outage?

A

Manually return blood and be vigilant for air bubbles.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Arterial pressure high alarm causes?

A

KINKS, CLOTS, NEEDLE SUCKING ON VESSEL WALL, INFLITRATION

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Arterial pressure high alarm intervention?

A

For clots it depends on severity but either saline flush or maybe stop machine if severely clotted.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Arterial pressure low alarm causes?

A

Needle is out.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

How does patients BP affect the arterial pressure?

A

If patients BP is LOW the AP can INCREASE. Is BP is HIGH the AP can DECREASE.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Venous pressure High alarm cause?

A

Clots.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Air/Foam Detector alarm causes and intervention?

A

Drip chamber under 3/4 full, saline bag is empty, improper priming.

Look at PT first and see if they’re okay. Assess machine, check arm and needle. Fill drip chamber, change saline bag etc.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Heparin pump alarm causes and intervention?

A

Incorrect setup

Correct setup

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Blood Pump alarm causes and interventions?

A

Incorrect setup, rollers not calibrated correctly.

Correct the setup, calibrate rollers, get help if necessary.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly