test 1 electric/hemo hw Flashcards
electronics and interventional
when catheter is too deep seated and receives backward pressure from the myocardium
ventricularization of the coronary artery
20 mmhg cardiac pressure recording too low. where was the transducer zeroed?
10 in above pt mid chest. (+2mmHg for every inch above level)
the usual sweep speed for displaying pressure recordings is
10 sec
another name for string gauge is
wheatstone bridge
adding extra lengths of tubing will cause
drop in pressure
balancing is also known as
zero
AO-RV-PA is this pullback or insertion?
insertion
PA-RV-AO is this pullback or insertion
pullback
high frequency filter can cause damping
false
arterial have triangular which have sine waves
none
time shifting the pcw waveform is useful to ensure accuracy when measuring
mitral valve gradient
a. No overshoot
b. No over damping
c. Good signal to noise ratio
d. Appropriate filtering
these are known as
optimal frequency response
a. Inappropriate frequency response caused by damping coefficient
b. Tangled cables, loose connections
c. Fluid leaks
d. Electromagnetic interference
these are considered
electronic artifacts
Pressure applied to a liquid at any point is transmitted equally in all
directions is known as
pascals law
pascals and mmhg are units of
pressure
the filters used in hemodynamic monitoring is ______ band pass filter
low
which waveform would be the most likely for respiration excursion
PCW
proper filtering gets rid of respiratory variation
false
a _____hz filter gets rid of ac power interference
60
strain gauge converts fluid pressure to an electrical signal using
variable resistors
two pressure channels are required to record
simultaneous pressures
what assures proper baseline and top of scale channels
calibration
changing the ____adjusts band pass filter to reduce noise
HZ
40 mm scale is for
Right heart pressures
wheatstone bridge is located
transducer