STE #1 Flashcards
Which of the following is a required pre-operational check?
- Physical condition of the instrument.
2 . Calibration due date
And 3. Battery check
Scintillation detector operate in the G-M region of the gas amplification curve? True or False?
False
Zinc-Sulfide detectors are good for detecting what type of radiation?
Alpha
Sodium Iodine scintillation detector is used for detecting gamma radiation. True or False?
True
You are using an RO-2 survey meter to measure the activity on a smear taken from the refueling canal. The closed window reading is 4 mR/hr and the open window reading is 22.4 mR/hr. Given a beta correction factor of 4.0 what is the dose rate in mrad/hr?
73.6
How does N-16 gamma’s affect G-M type detectors?
Can cause them to over respond
You took an air sample and count it for 10 minutes and it had a reading of 22000 counts. The 60 minute background count on the instrument was 1400 counts.
What is sample net counts per minute (round your answer)?
2177
Entering a high gamma dose rate field with an extendable G-M detector on low range scale could result in which of the following…
Instrument could become saturated which would result in a low reading
Ionizing radiation interacts with phosphor crystalline material to produce light is called scintillation. True or False?
True
In a noble gas atmosphere a vented ion chamber could…
Become saturate giving a false reading
When alpha frisking how close should the detector be to the surface of the object
¼ inch
In a G-M detector, the amount of time from the initial pulse until the detector can produce another pulse is called
Dead time
In a G-M detector, resolving time is the time from the initial measured pulse until another pulse can be measured. True or False?
True
In a G-M detector, saturation is when a detector reads high in a high field of radiation. True or False?
False
Neutron detector use some type of moderator to slow down the fast neutrons so they can measured. The moderator most often used is
Sodium