Static Magnetic Fields Flashcards
If an object is magnetically saturated, it means that:
The object is wet and should not be used in the magnetic field
An increase in the applied external magnetic field cannot increase the magnetization of the object further
The object is no longer considered magnetic
Condensation has formed on the object while it was in the static magnetic field
An increase in the applied external magnetic field cannot increase the magnetization of the object further
Within the static magnetic field, where is the rotational force/torque exerted on a metallic object the greatest?
Mouth of the bore
Wall of the bore
Entrance to Zone 4
Isocenter
Isocenter
An object deflects 45 degrees during deflection testing. What force(s) has the most effect on the device?
The external magnetic field
Gravity
Scarring within the patient due to healing
Gravity and the external magnetic field are exerting equal force on the device
Gravity and the external magnetic field are exerting equal force on the device
The twisting force on a ferromagnetic object as it attempts to align parallel to the external magnetic field is called _____.
Missile effect
Translation
Torque
Magnitude
Torque
Mechanical forces related to the static magnetic field include translational (missle), rotational (torque), and Lenz effects. True or False?
True
False
True
1T converted to Gauss would be:
10 Gauss
100 Gauss
1,000 Gauss
10,000 Gauss
10,000 Gauss
rimary concerns with static magnetic fields are __________ .
Translation and rotation (torque) forces
Focal heating
Induced voltages
Retention
Translation and rotation (torque) forces
If there is an area that reads 16T/m on a spatial gradient map, that means:
The strength of the magnetic field would change by 16 Tesla if that same localized rate of change in that small area were to occur over 1 meter of distance.
The strength of the magnetic field would change by 16 Tesla if an implant were to be spread out over 1 foot of distance.
The maximum field strength of that area within the magnet is 16T
The magnet will change from 1T to 16T across the length of the bore.
The strength of the magnetic field would change by 16 Tesla if that same localized rate of change in that small area were to occur over 1 meter of distance.
Within the magnetic field, torque/rotational force will be the greatest at __________.
mouth of the bore
edge of gradient coil
isocenter
within transmit coil
isocenter
What type of map is this?
Spatial Gradient Map
Gradient Map
RF Deposition Map
B0 map
B0 map
Shielding of the magnet has what effect on the magnetic field?
Increases the size of the fringe field and decreases the MSG
Increases both the size of the fringe field and the MSG
Decreases both the size of the fringe field and the MSG
Decreases the size of the fringe field and increases the MSG
Decreases the size of the fringe field and increases the MSG
The concentration of the orbiting magnetic particles in space within the magnetic field is termed:
Proton Density
Field Density
Torque Density
Flux Density
Flux Density
If an object is magnetically saturated, increasing the external magnetic field will have no further effect on the object. True or False?
True
False
True
Potential biologic effects of the static magnetic field include all the following except:
Magnetophosphenes
Peripheral nerve stimulation
Magnetohydrodynamic effect
EKG perturbances
Vertigo/dizziness
Peripheral nerve stimulation
In viewing a spatial gradient map, there will be no translational forces exerted on an object when:
Behind the 5 G line
Below the patient table
The spatial gradient is at its maximum
There is no change in the strength of the main static magnetic field- Like at isocenter
There is no change in the strength of the main static magnetic field- Like at isocenter
How does shielding within the magnet housing affect the system spatial gradient?
it doesn’t, a system’s SG is the same shielded or unshielded
it increases the system SG because it decreases the size of the fringe field, forcing lines of flux closer together
it decreases the SG because the fringe field is smaller
the shielding is in the walls, not in the magnet system
it increases the system SG because it decreases the size of the fringe field, forcing lines of flux closer together
What is defined as “A visual representation of the change of the magnetic field strength over distance?”
Spatial gradient map
Gradient map
B0 map
Deposition map
Spatial gradient map
What line is regarded as the point in which non-cleared implants and devices cannot cross safely?
5 mT
500 G
0.05 mT
5 G
5 G
The static magnetic field has no bioeffects. True or False?
True
False
False
What is the purpose of shielding within the magnet housing/bore?
it makes the magnet safer because the fringe field is smaller
it compresses the fringe field, allowing for a shorter distance to the 5 Gauss line
shielding within the magnet housing replaces shielding within the scan room walls
it increases the cost of the system
it compresses the fringe field, allowing for a shorter distance to the 5 Gauss line
What phrase describes second level in regards to static magnetic fields?
No undo physiologic effects are associated with the static magnetic fields of these strengths
A potential for harm to a patient exists
There is definite potential for harm to a patient
It is safe to scan the patient at second level only
There is definite potential for harm to a patient
Based on the MSG map above, what would be the maximum spatial gradient that the patient’s head might be exposed to in this position while being moved into the bore?
5 T/m
3 T/m
15 T/m
7 T/m
0.5 T/m
3 T/m
A 3T system will always have higher magnetic attractive force than a 1T, regardless of magnet format. True or False?
True
False
False
Does gravity or the static magnetic field have a greater effect on an implant that shows a deflection angle of 39 degrees?
Gravity
Magnetic Field
Both Equally
Gravity
Bioeffects of the static magnetic field are temporary and disappear when the patient is removed from the field. True or False?
True
False
True
The maximum Tesla strength that is safe for infants younger than one month per the FDA is:
8T
1T
4T
1.5T
4T
Spatial gradient map values are typically measured in _______ or ________.
mT/Kg or J/cm
W/Kg or J/cm
G/Kg or W/cm
T/m or G/cm
T/m or G/cm
Lenz forces are more likely to be experienced by electrically conductive material oriented _______ the static magnetic field.
through
parallel to
perpendicular to
diagonally through
perpendicular to
During MR safety testing, an implant demonstrates a deflection angle of 45 degrees. Based solely on this information, the implant would be labeled _____.
MR Safe
MRI Compatible
MR Unsafe
MR Conditional
MR Unsafe
Please convert 100 mT to Gauss
1 G
10 G
100 G
1000 G
1000 G
The maximum Tesla strength that is safe for adults, adolescents, and infants older than one month per the FDA is:
3 Tesla
7 Tesla
8 Tesla
10 Tesla
8 Tesla
Which of the following is NOT a possible bioeffect caused by the static magnetic field?
Magnetophosphenes (flashes of light)
Induced voltages within patient tissues
Nystagmus (repetitive, uncontrolled eye movements)
Vertigo (inner ear effects)
Nystagmus (repetitive, uncontrolled eye movements)
Which object would experience the most translational force when exposed to the static magnetic field?
1 kg spherical object
1 kg elongated object oriented parallel to the static magnetic field
1 kg elongated object oriented perpendicular to the static magnetic field
1 kg wire wound into a tight ball
1 kg elongated object oriented parallel to the static magnetic field
Translational (missile) force is affected only by the spatial gradient of the static magnetic field. True or False?
True
False
True
In horizontal bore scanners, where will the translational force exerted on a metallic object the greatest?
Isocenter
Entrance to Zone 4
Mouth of the bore
Wall of the bore
Mouth of the bore
What possible effects can static magnetic fields have on active implants?
Attractive force, focal heating, induced voltages
Attractive force, affect implant functionality, displacement of implant
PNS, rotational force, translational force
False feedback, affect implant functionality, torque
Attractive force, affect implant functionality, displacement of implant
Fringe fields are typically either measured in ___________ or ___________, depending on the part of the world.
Gauss / Joules
MilliTesla / Joules
Gauss / milliTesla
Watts / decibels
Gauss / Tesla
Gauss / milliTesla
Lenz’s forces are exhibited when a large conductive object moves through the field while oriented perpendicular to the main magnetic field. True or False?
True
False
True
What static magnetic field scanning level would result in “No undo physiologic effects” to the patient?
Normal Mode
First Level Mode
Second Level Mode
MR Safe Mode
Normal Mode
Maximum spatial gradient is proportional to static magnetic field strength, in that higher field strength systems will always have higher MSG. True or False?
True
False
False
A patient complains of a pulling sensation in the area of their implant while they are being advanced into the bore. The sensation goes away when the table motion is stopped. The pulling sensation is due to:
Translational forces of the static magnetic field
Lenz forces caused by the implant moving through the lines of flux
The patient’s imagination
It doesn’t matter because the exam should be cancelled
Lenz forces caused by the implant moving through the lines of flux
The size and shape of a metallic object will affect its potential projectile effect within the magnetic field. True or False?
True
False
True
The spatial gradient of the static magnetic field is consistent throughout the entire fringe field. True or False?
True
False
False
Magnetophosphenes are more easily produced in higher static field strength systems. True or False?
True
False
True
All of the following are true for spatial gradient maps EXCEPT:
Measured in T/m
May vary in appearance depending on the manufacturer
Visual representation of the change of the magnetic field strength over distance
Visual way of telling the field strength of a static magnetic field
Visual way of telling the field strength of a static magnetic field
The potential biologic effects of the static magnetic field are thought to be harmful and irreversible. True or False?
True
False
False
The patient may be exposed to magnetic lines of flux upon entering zone 4. True or False?
True
False
True
t is possible for a vertical field open system to have a higher magnetic attractive force than a 3T system. True or False?
True
False
True
Pure aluminum and titanium do not have any interactions with the static magnetic field. True or False?
True
False
False
During MR safety testing, an implant demonstrates a deflection angle of 25 degrees. Based only on this information, the implant would be labeled _______.
MR Safe
MR Conditional
MR Unsafe
MRI Compatible
MR Conditional
Bioeffects of the static magnetic field are permanent. True or False?
True
False
False
What is 5 Gauss converted to milliTesla?
0.5 mT
0.05 mT
50 mT
500 mT
5000 mT
0.5 mT
A way to demonstrate Lenz’s forces is:
Walk toward the bore with a pen in your pocket
Throw your keys toward the magnet from the scan room door
Walk toward the magnet while holding a large aluminum object oriented perpendicular to the main magnetic field
Quench the magnet and see what happens
Walk toward the magnet while holding a large aluminum object oriented perpendicular to the main magnetic field
Implants that have a ________ will try to align ____________ to the direction of the main static magnetic field.
short axis / perpendicular
long axis / perpendicular
long axis / parallel
long axis / parallel
Which deflection angle(s) would constitute an MR Safe implant?
0 degrees
15 degrees
30 degrees
44 degrees
0 degrees
Conductive, nonmagnetic implants/objects will not experience Lenz forces as they move through the static magnetic field. True or False?
True
False
False
For the magnet diagram shown here, approximately how far from isocenter (in the Z-direction) lies the “pacemaker line” where there is the concern that the static magnetic field could potentially interfere with active implanted medical devices?
1 meter
2.5 meters
3.5 meters
4 meters
4 meters
If you are scanning on a 1.5T scanner, this means that a static magnetic field strength of 1.5T is the maximum a patient will encounter within zone 4. True or False?
True
False
False
Force product is the product of:
The strength of the static magnetic field and the spatial gradient of the static field
The RF field and the slope of the time varying gradients
The weight of the magnet and the length of the bore
The length of the body coil and the gradient slew rate
The strength of the static magnetic field and the spatial gradient of the static field
Which factor will NOT affect the strength of Lenz forces on an object?
Speed of an objects entry into the bore
Object’s electrical conductivity
Cross-sectional size of object
Spatial gradient
Deflection angle of object
Spatial gradient
The translational force exerted on an object in the magnetic field will be constant throughout the field. True or False?
True
False
False
A surgical clip with a long axis oriented in the head to foot direction will experience a __________ force and align ___________ to the magnetic lines of flux when placed into a vertical field / open MRI.
rotational (torque) / parallel
translational / parallel
rotational / perpendicular
translational / perpendicular
rotational (torque) / parallel
It is possible for magnetic lines of flux to extend through:
Floors
Walls
Ceilings
Floors and walls
Ceilings and walls
All of the above
Floors
Walls
Ceilings
Floors and walls
Ceilings and walls
All of the above
Does the shape of an implant or device affect the translational force exerted on the implant/device by the static magnetic field?
Yes
No
Yes
For the electromagnetic fields used during MR imaging, which is of concern beyond the volume of the bore?
Static magnetic field
Gradient magnetic field
Radiofrequency magnetic field
Static magnetic field
5 Gauss translates into _______ mT.
5 mT
0.5 mT
0.005 mT
50 mT
0.05 mT
0.5 mT
0.35 mT translates into _______ G.
35 G
3.5 G
0.035 G
350 G
3.5 G
The torque force exerted on an object is a function of:
The magnetic spatial gradient
The strength of the static magnetic field
The gradient slew rate
The length of time is has been exposed to the magnetic field
The strength of the static magnetic field
Which effects/forces would be exerted on a metallic object as it enters the bore of the magnet?
Lenz, Translational, Induced Voltage
Lenz, Rotational, Translational
Translational, Induced Voltage, Rotational
Torque, Heating, Rotational
Lenz, Rotational, Translational
Translational (missile) force is affected by which field?
Strength of the static magnetic field
Radiofrequency field
Spatial gradient of the static magnetic field
Slope of the time varying gradients
Spatial gradient of the static magnetic field
Ferromagnetic materials do not exhibit Lenz effects. True or False ?
True
False
False
Which deflection angles (without sufficient anchoring forces) would constitute an MR Conditional implant?
1-44 degrees
15-45 degrees
30-65 degrees
46-90 degrees
1-44 degrees
During MR safety testing, an implant demonstrates no deflection. Based solely on this information, the implant would be labeled _____.
MR Conditional
MR Safe
MR Unsafe
The implant cannot be labeled because we do not have enough information
The implant cannot be labeled because we do not have enough information
What static magnetic field level mode would a 9T MRI Scanner fall under?
Normal Mode
First Level Mode
Second Level Mode
MR Safe Mode
Second Level Mode
A patient is known to be free of metallic implants/devices. Possible interactions between the patient and the static magnetic field could be:
Claustrophobia
Nothing, The patient is metal-free and will therefore have no interaction with the static magnetic field
Dizziness, nystagmus, altered EKG
Hearing loss
Dizziness, nystagmus, altered EKG
It is impossible for non-magnetic metals to interact with the static magnetic field. True or False?
True
False
False
What factor does NOT directly contribute to Lenz effects?
Conductivity
Cross-sectional area
Speed of motion
Magnetic field strength
Spatial gradient though which the object is moving
Magnetic field strength
Torque force exerted on an object in the magnetic field will be the greatest as the object enters the bore. True or False?
True
False
False
What is referred to as “stray lines of magnetic flux that extend outside the bore of the MRI scanner”?
Magnetophosphenes
Fringe field
Flux density
Shielding
Fringe field
What would be an example of first level in respect to static magnetic field strength?
0.2T
1.0 T
3.0 T
7.0 T
9.0 T
7.0 T
If the patient is feet first, supine, what spatial gradient will the head be exposed to at approximately 1.5 meters from isocenter?
approx 5 T/m
approx 3 T/m
approx 1 T/m
approx 0.7 T/m
approx 0.5 T/m
approx 0.5 T/m
_______________ is when an object becomes airborne when exposed to a static magnetic field.
Missile effect
Translational force
Flux effect
Torque effect
Missile effect
Characteristics of an object that affect its potential projectile effect within a magnetic field include all of these EXCEPT:
Ferromagnetic property/material
Mass
Age
Size/shape
Age
Because the static magnetic field is homogeneous at isocenter, translational (missile) forces there are:
Unable to be measured
At their maximum
Zero
Unimportant
Zero
Within the magnetic field, translational force will be the greatest at ____________ and the least or zero at ___________.
the location of maximum spatial gradient / isocenter
isocenter / end of table
close to the bore walls / edge of body coil
the location of maximum spatial gradient / isocenter
For most full body MR systems, the homogeneity of the static magnetic field around isocenter is an area the size of:
The entire body coil
A tennis ball
Only isocenter
A basketball
A basketball
During MR safety testing, a metallic implant demonstrates no deflection. Based on this information, the implant would be labeled _____.
MR Safe
MR Unsafe
MR Conditional
MR Compatible
MR Conditional
A paramagnetic object or substance will retain magnetism when removed from the external magnetic field. True or False?
True
False
False
The translational force exerted on an object is a function of:
The strength of the static magnetic field
The magnetic spatial gradient
The gradient slew rate
The length of time it has been exposed to the magnetic field
The magnetic spatial gradient
A ferromagnetic object or substance will retain magnetism when removed from the external magnetic field. True or False?
True
False
True
The ______ and ________ of an implant is very important in regards to rotational (torque) force.
Size / Shape
Composition / Position
Shape / Position
Size / Position
Composition / Size
Deflection testing is performed by:
Placing the object inside a patient then exposing it to the magnetic field
Deciding how the object will behave in the magnetic field by studying its material content
Suspending the object from a string and placing it in the magnetic field
We don’t know, it’s a proprietary process
Suspending the object from a string and placing it in the magnetic field
Which of the following is NOT a tool / procedure that the technologist should use to avoid a static magnetic field injury?
Ferrous metal detector
5mm gap between patient and bore wall
Screening questionnaire
Verbal interview
5mm gap between patient and bore wall
What types of materials are most likely to exhibit noticeable Lenz effects?
PVC, Plastics
Aluminum, Brass, Copper
Surgical Steel, Cobalt, Nickel
Gadolinium
Aluminum, Brass, Copper
Lenz’s forces are harmful and there is no way to mitigate them. True or False?
True
False
False