Rite ad 1 Flashcards
a 180 deg pulse contains how much more energy than a 90 deg pulse
4x
an aneurysm cip placed after 2019 at a reputible facility will place what risk to the patient
low
which half wavelength at 7t will place a patient at rish from burns
5-7 cm
according to the IEC 60601-233 doc, it is fine to scan a pregnant woman in the first trimester
false
according to the VA Health directive of 2013 it is recommended that how many techs be working within zone 2
two
A/m defines
magnetic field strength
as we increase body size, SAR will increase by how much
squared
as we increase field strength SAR will increase by how much
squared
BO defines
magnetic flux density
B1+rms is measured in what
W/kg J/kg microtesla
can a patient be scanned at 3t if the conditions state the device can be scanned at 128MHz
Yes
A condition states that an device can be scanned at 64MHZ, this means what
it can be scanned at 1.5TY
Copper has a weakly repulsive force in the presence of a magnetic source, it is considered what?
Diamagnetic
The created chance for induced current in a patient would be found at?
peripheral to isocenter
delivering more RF pulses to a patient will do what
increase heating
FPO:B will do what
it will limit B1+rms and slew rate
gadolinium contrast will to what to T1 and T2 relaxation
shorten it
how tightly packed the lines of magnetic flux are descibed how
magnetic flux density
an implant condition states an implant can be scanned only in normal mode. What does this mean
we are limited to 2W.kg whole body SAR
in a gradiant profile which of the following describes a period of time where the gradients are not changing
duty cycle