Screen Film Radiographic Technique Chapter 13 Flashcards
Prime Exposure Factors
kVp
mA
Exposure Time
Source to Image Distance (SID)
kVp
Controls image contrast Primary control of quality Controls energy and beam penetrability Higher quality means higher energy Greater kVp=more Compton=more scatter=reduced contrast. ie....Chest exam, high kVp, low contrast
mA
Controls image density or brightness
controls number of x rays that are produced
if time does not change, changing the mA will change quantity of x rays proportionately. Double exposure=Double Patient Dose
No change in energy, just quantity
x ray systems are often identified by their maximum mA
Exposure Time
Reducing time reduces motion blur
mAs= mA x time (seconds)
Time is decreased, mA must be increased to retain the same OD/brightness
Falling Load Generators
This type of generator only allows the tech to select mAs
not mA and time separate
Highest mA and shortest time is automatically used for each exposure
Distance
effects exposure and intensity
no effect on quality of radiation
Standard SIDs are used to ensure continuity 40, 44, 72 inches
More SID means:
Less magnification and blur, increased spatial resolution
More mAs must be used with a longer SID
Direct Square Law
Allows rad tech to calculate new mAs after a change in SID
New mAs/old mAs= (SID) suared new SID/ (SID) squared old SID
Direct not inverse relationship
Half the distance= 1/4 mAs
Double the distance= 4x’s the mAs
Imaging system characteristics
Focal spot size
Filtration
High voltage generation
Focal spot size
Tubes have Large and Small Focal spots
Large focal spots are used for general radiography, thick dense objects such as abdomen and pelvis.
Small focal spots are used for detail radiography
Large focal spots
Produce more mAs
shorter time can be used
Small focal spots
produce better record detail
High mAs is not needed
Filtration
Inherent
Added
Compensating
Filtration increases beam quality
Filtration works like kVp: both increase beam quality
Result:
increased scatter
Decreased image contrast
Inherent filtration
2.5 mm Al equivalent is required
Metal or glass envelope provides 0.5 mm Al equivalent
Collimator provides normally 1.0 mm Al equivalent
Another 1 mm Al equivalent is added to meet the minimum requirement
Added filtration
Normally not necessary to change filtration of an imaging system, however, some tubes are equipped to do so.
Higher filtration can or would be used for exams with High Subject Contrast such as extremities, joint, chest
Adding filtration decreases patient dose
Compensating filters
Mounted on the collimator for specific exams, slides in or is magnetic
Wedge shaped filters are useful for spine studies, cross table hips, leg length studies
High voltage generation
x ray quantity and quality are greatly influenced by the type of generator
Single phase, three phase, and high frequency
Half and full wave rectified generators (Single phase)
Only emit x rays half the time
Half wave generators produce the SAME quality as full wave
BUT
half wave produce HALF the quantity of x rays.
Full wave
has NO dead time
Three phase and high frequency generators
produce higher quantity and quality
Most units today are high frequency
Technique selection factors
Patient factors
Image quality factors
Exposure technique factors
Patient factors include
habitus
body thickness
body composition
pathology
Sthenic- Average, strong, active
Hyposthenic- thin but healthy
Hypersthenic- overweight, large frame
Asthenic-small, frail
Body thickness
Thicker patients require more x radiation to penetrate them
Thickness should be measured with calipers
Vary kVp or mAs according to patient thickness