RAPHEX XIII Flashcards
convert Sv to rem
1 Sv = 100 rem
All of the following statements regarding x-ray production are true except for which statement? A. X-rays are produced by two different mechanisms: bremsstrahlung and characteristic x-ray emission.
B. Characteristic x-rays have discrete energies. C. Bremsstrahlung x-rays have a continuous spectrum of energies. D. The efficiency of x-ray production is about 1% for 100-kVp beams. E. None—all of these statements are true.
None—all of these statements are true
Typical x-ray transformers for general radiography have a power rating of about _____. A. 1 W B. 100 W C. 1 kW D. 100 kW
D. 100 kW
According to the electrical power formula, electrical power P (in watts, W) equals the product of the current I (in amperes, A) and voltage V (in volts, V), P = I × V. For a typical x-ray tube for general radiography, the tube voltage is of the order 100 kV = 100,000 V, and the tube current is about 1,000 mA = 1 A, yielding a power P of about 100,000 V × 1 A = 100,000 W = 100 kW
The SI unit of air kerma is the _____. A. roentgen (R) B. sievert (Sv) C. rad D. curie (Ci) E. gray (Gy)
E. gray (Gy)
The SI unit of air kerma is the gray (Gy) (or joule per kilogram of air, J/kg), the same as the unit of absorbed dose. It is formally defined as the sum of the initial kinetic energies of all the charged particles liberated by uncharged ionizing radiation (i.e., indirectly ionizing radiation, such as photons and neutrons) in a volume of air, divided by the mass of the volume of air. However, kerma can be different from absorbed dose, depending on the radiation energies involved. This is because ionization energy is not taken into account. While kerma approximately equals absorbed dose at low energies, kerma is higher than absorbed dose at higher energies, because some energy escapes from the absorbing volume in the form of bremsstrahlung x-rays or high-energy electrons, and is not counted as absorbed dos
_____ is the predominant interaction of 120-kVp x-rays, commonly used in CT scanners, in soft tissues. A. Rayleigh scatter B. Photoelectric interaction C. Pair-production D. Compton scatter
D. Compton scatter
For an x-ray spectrum typically produced in a CT scanner at 120 kVp, the average beam average energy is high enough for Compton scatter to predominate in soft tissues. At the same time, the average energy is well below the threshold energy, 1.022 MeV, for pair production.
In plain radiography, _____ reduces patient dose. A. using a grid B. filtering the beam C. lowering the tube voltage (i.e., kVp) D. decreasing the source-to-object distance
B. filtering the beam
Filtering removes lower-energy, less-penetrating photons from the x-ray beam. These photons, since they are more likely to be absorbed in the patient, increase patient dose. By preferentially removing these x-rays from the beam, filtering reduces the dose. All other choices listed increase the patient dose.
The radiographic modality with the highest beam quality is _____. A. diagnostic radiology B. computed tomography C. fluoroscopy D. mammography
B. computed tomography
Beam quality, which essentially refers to the average x-ray energy, increases with increasing x-ray tube voltage (i.e., by kVp) and filtration. Since CT scanning is performed with a relatively high tube voltage of 100 kVp or greater and with a highly filtered x-ray beam, CT has the highest beam quality among radiographic modalities.
The automatic exposure control system of a CT scanner determines the tube current for a particular scan based on a planning view (scout) image acquired with the tube stationary under the patient’s bed. If the patient centerline is positioned below scanner isocenter, which of the following will be reduced? A. image noise B. low-contrast visibility C. spatial resolution D. patient dose
A. image noise
The closer the patient is to the x-ray tube, the more magnification will occur, with the patient appearing to be larger (i.e., have a greater body thickness) than he/she actually is. This will result in increasing the exposure (i.e., milliampere-second, mA), which reduces the statistical noise in the resulting images, but increasing the radiation dose to the patient’s spatial resolution is not affected by the exposure (i.e., mAs
dose is higher if patient is near source
relative noise is lower because more mA
The device that produced the image shown here was _____. A. a CT scanner B. an MRI scanner C. a C-arm image-intensifier (II) fluoroscopy unit D. a mobile DR E. a bone densitomet
C. a C-arm image-intensifier (II) fluoroscopy unit
The image is an artifact from stray magnetic fields distorting the image in the image intensifier (II) of a C-arm fluoroscopy unit.
In the American College of Radiology (ACR) CT Accreditation Program, which of the following ranges of CT numbers in Hounsfield Units (HUs) would be the allowed (i.e., passing) range for air insert? A. 970 to 1005 B. 850 to 970 C. –7 to 7 D. –850 to –970 E. –970 to –100
E. –970 to –1005
The test tool shown in the image below is used to evaluate _____.
A. spatial resolution B. focal spot size C. contrast D. modulation transfer function (MTF) E. CT slice thickne
B. focal spot size The tool shown is the star pattern used to measure x-ray tube focal spot size
If the flux gain and overall brightness gain of an image intensifier are 50 and 5000, respectively, the image intensifier’s minification gain is _____. A. B.
100 500
C. 2,500 D. 10,000 E. None of the above is true
A. 100
The flux gain of an image intensifier (II) is the number of light photons emitted at the output phosphor for each photon emitted at the emitted phosphor. The minification gain is the increase in image brightness resulting from the reduction in image size from the input to the output phosphor. The overall brightness gain is the product of the flux gain and the minification gain. The minification gain is, therefore, the overall brightness gain divided by the flux gain, in this case, 5000 divided by 50 or 100.
Fluoroscopy is typically performed with an x-ray tube current of about _____. A. 0.3 mA B. 3 mA C. 30 mA D. 300 mA
3 mA
Compared to continuous fluoroscopy, pulsed fluoroscopy _____. A. uses a higher x-ray tube current B. may reduce the patient dose C. results in sharper images D. All of the above are true.
D. All of the above are true.
In conventional fluoroscopy, the x-ray tube current is very low (about 3 mA) but is on, and x-rays are produced continuously. In pulsed fluoroscopy, the tube is on intermittently (3-msec frames) but at a much higher current (about 30 mA). Pulsed fluoroscopy reduces the patient dose by 30% to 40% when acquiring frames at a rate of fewer than 30 frames/sec. And because the frames are so short (about 3 msec/frame for pulsed fluoroscopy versus 30 msec/frame for conventional fluoroscopy), blurring due to patient motion is reduced and the images are, therefore, sharper
An interventional fluoroscope is set up to perform fluoroscopy at 15 frames/sec, cinefluorography at 10 frames/sec, and digital subtraction angiography (DSA) at 3 frames/sec. The correct order of patient entrance dose rates from lowest to highest entrance dose rate, when tested with a 20-cm-thick water phantom, is _____. A. cinefluorography, DSA, fluoroscopy B. cinefluorography, fluoroscopy, DSA C. DSA, cinefluorography, fluoroscopy D. DSA, fluoroscopy, cinefluorography E. fluoroscopy, cinefluorography, DSA
E. fluoroscopy; cinefluorography; DSA
For consistency of image quality, fluoroscopy systems are generally configured such that the same image receptor dose is achieved for 1 DSA frame, 10 cinefluorography frames, or 100 fluoroscopy frames. Patient entrance dose approximately tracks image receptor dose. For a one-minute (or 60-second) run, therefore, the “relative dose” for each modality is as follows:
Fluoroscopy “relative dose” of 1 per frame × 15 frames/sec × 60 seconds → “relative dose” of 900
Cinefluorography “relative dose” of 10 per frame × 10 frames/sec × 60 seconds → “relative dose” of 6000
DSA “relative dose” of 100 per frame × 3 frames/sec × 60 seconds → “relative dose” of 18,000
What is the relationship between the modulation transfer function (MTF) and the line spread function (LSF) of an imaging system? A. The MTF is the derivative of the LSF. B. The MTF is the log transform of the LSF. C. The MTF is the integral of the LSF. D. The MTF is the Fourier transform of the LSF. E. The MTF and LSF are unrelated.
D. The MTF is the Fourier transform of the LSF.
The standard deviation and percent standard deviation of the number of counts per pixel are _____ and _____, respectively. A. 25 counts; 20% B. 50 counts; 14% C. 100 counts; 10% D. 200 counts; 7
C. 100 counts; 10%
I think they meant to ask for number of counts
For an image intensifier-based fluoroscope, switching from 9-inch to 6-inch magnification mode results in a _____. A. dose decrease by a factor of (9/6)2 B. dose decrease by a factor of (9/6) C. dose increase by a factor of (9/6)2. D. dose increase by a factor of (9/6)
C. dose increase by a factor of (9/6)2
For image intensifier-based fluoroscope, the dose increases by a factor determined by the ratio of magnification squared.
For an image intensifier-based fluoroscope, switching from 9-inch to 6-inch magnification mode results in _____. A. poorer resolution (line pairs/mm) by a factor of (9/6)2 B. poorer resolution (line pairs/mm) by a factor of (9/6) C. improved resolution (line pairs/mm) increase by a factor of (9/6)2 D. improved resolution (line pairs/mm) increase by a factor of (9/6)
D. improved resolution (line pairs/mm) increase by a factor of (9/6)
For an image intensifier-based fluoroscope, the resolution increases by a factor determined by the ratio of magnification.
Pure beta particle-emitting radionuclides are best shielded using _____. A. plastic B. lead C. steel D. titanium E. zinc
A. plastic
Beta particles are considered non-penetrating radiation and are best shielded by materials with a low atomic number (Z) such as plastic, a relatively thin layer of which will completely attenuate any incident beta particles with negligible bremsstrahlung production. Higher-Z materials such as metals would produce more bremsstrahlung. High-energy beta particles, such as those emitted by 90Y, can produce more bremsstrahlung, and the low-Z plastic shield can be supplemented by lead shielding to attenuate any bremsstrahlung produced in the plastic. Typically, the amount of bremsstrahlung thus produced will be very small in any case
The device pictured to the right is a/an _____, and it is used to assess the _____ of a gamma camera. A. ACR phantom; sensitivity B. Mobius phantom; uniformity C. transmission phantom; dead time D. four-quadrant bar phantom; spatial resolution
D. four-quadrant bar phantom; spatial resolution
A technetium-99m (99mTc) transmission image of a four-quadrant bar phantom depicted is widely used to assess gamma camera spatial resolution
In preparation for radioembolic therapy of liver cancer by infusion of yttrium-90 (90Y)-labeled microspheres via the hepatic artery, a pre-therapy imaging study was performed using technetium-99m (99mTc)-macroaggregated albumen (MAA) to estimate lung shunting. The images of and liver and lung region-of-interest (ROI) counts were as follows
lung ROI counts = 4168
liver ROI counts = 485912
Therefore, the lung shunting is _____. A. 0.5% B. 0.9% C. 1.6% D. 2.5%
B. 0.9%
The lung shunt is calculated as the ratio of the 99mTc-MAA counts in the lung region of interest (ROI) counts divided by the sum of the counts in both the lung and liver ROIs multiplied by 100%. In this case, [4168 / (4168 + 485,912)] × 100% = 0.85% ≈ 0.9
In iodine-131 (131I)-iodide treatment of metastatic thyroid cancer, the β-particles from 131I in a lesion account for _____ percent of the radiation dose to the lesion. A. 0 B. <10 C. 25–50 D. >75
D. >75
Iodine-131 (131I) beta particles, like beta particles and other particulate radiation generally, have a very short range (about 1 mm or less) in tissue and deposit all of their energy, and deliver all of their radiation dose, within 1 mm or less of their point of emission. Thus, beta-particles emitted by 131I within a thyroid cancer lesion will deposit virtually all of their energy
in and deliver virtually all of their dose to the lesion itself. The 364-keV gammas emitted by 131I, like other high-energy x-and gamma-rays, travel many centimeters in tissue and will deposit much of their energy and deliver much of their dose outside the lesion, contributing very little to the lesion dose.
What is the intensity half-value thickness for a 6-MHz ultrasound beam, assuming the soft-tissue attenuation coefficient is 0.5 dB/cm/MHz? A. 0.5 cm B. 0.75 cm C. 1 cm D. 1.5 cm E. 2.0 cm
C. 1 cm
The intensity half-value thickness (HVT) is defined as the thickness of material that will reduce the initial ultrasound beam intensity by one-half. Relative intensity equals 10 × log(I2/I1) where I1 and I2 are the intensities of the incident and reflected intensities, respectively, of the beam. Solving this equation for a relative intensity that is decreased by one-half will yield a 3-dB reduction. Thus, a half-value thickness (HVT) will reduce the ultrasound beam intensity by 3 dB. In this case, we have 0.5 dB/cm/MHz at 6 MHz, which means an attenuation coefficient of 3 dB/cm. Thus, 1 cm material will reduce the intensity by 3 dB or 50%
The gamma camera performance test known as multi-peak spatial registration is performed using _____ isotopes. A. fluorine-18 (18F) B. gallium-67 (67Ga) C. zirconium-89 (89Zr) D. technetium-99m (99mTc) E. iodine-123 (123I)
B. gallium-67 (67Ga)
Multi-peak (or multi-window) spatial registration is a gamma camera performance test that should be done with isotopes that emit multiple “image-able” gamma-rays, such as gallium-67 (67Ga), which emits a 93-keV gamma-ray (40% abundance), a 185-keV gamma-ray (24%), and a 300-keV gamma-ray (22%). All of the other isotopes listed are either a positron emitter used in PET (18F, 89Zr)) or emit a single image-able gamma-ray (99mTc, 123I). Gamma cameras essentially acquire a separate image for such gamma-ray and then superimpose (i.e., sum) them to yield the final image. Multi-peak spatial registration expresses how accurately these multiple images are spatially registered. Typical values of this parameter are about 1 mm, meaning that the multiple gamma-ray images are registered with each other to within 1 mm.
A lung cancer patient weighing 74 kg undergoing a fluorine-18 (18F)-fluorodeoxyglucose (FDG) PET scan is injected with 370 MBq of 18F-FDG. What is the standard uptake value (SUV) for the lung lesion (arrow) having an activity concentration of 16 kBq/cm3? A. 1.0 B. 2.4 C. 3.2 D. 4.0 E. 4.8
C. 3.2
PET images are routinely parameterized in terms of the standard uptake value (or SUV), defined as the activity concentration in tissue (kBq/g) divided by the activity (kBq) administered per unit body mass (g). For a lesion with an activity concentration of 16 kBq/cm3 in a patient with a total-body mass of 74 kg or 74,000 g who received an administered activity of 370 MBq or 370,000 kBq, the SUV is 16 kBq/g / (370,000 kBq / 74,000 g) = 16 kBq/g / 5 kBq/g = 3.2, assuming (as usual) a tissue mass density of 1 g/cm3.
For modern bone densitometry units, which of the following statements is not true? A. Such units utilize pencil-beam or fan-beam photons. B. Modern bone densitometry utilizes gadolinium-153 (153Gd) gamma-rays for the photon beam.
C. Such units are capable of body composition measurements. D. The effective dose (ED) to the patient for a whole-body bone densitometry study is considerably lower than that for a chest/abdomen/pelvis CT scan.
B. Modern bone densitometry utilizes gadolinium-153 (153Gd) gamma-rays for the photon beam.
Current-day bone densitometry units are capable of body composition measurements utilizing dual-energy x-rays rather than gadolinium-153 (153Gd) gamma-rays. Radioactive 153Gd sources were used at one time in bone densitometry units but were replaced by x-ray tubes a number of years ago. The x-ray beam geometries are either a scanning pencil-beam or scanning fan-beam. Bone densitometry/body composition analysis using dual-energy x-ray absorptiometry is a dosimetrically favorable procedure, delivering considerably lower radiation doses than most other radiological studies, with effective doses (EDs) of well under 100 mrem or 0.1 rem. In contrast, a typical chestabdomen/pelvis CT delivers an ED of ~1.5 rem to an adult patient.
A blood sample counted for 2 minutes yields 1800 counts. A 2-minute background measurement is made yielding 200 counts. What is the standard deviation σ of the net (i.e., background-subtracted) count rate (in counts per minute, cpm)? A. 100 cpm B. C. D. E.
88 cpm 44 cpm 22 cpm 12 cpm
D. 22 cpm
The standard deviation σ of a total of N counts is . For equal counting intervals, the net counts Nn equals the difference of the gross, or total, counts Ng and the background counts Nbg, that is, Nn = Ng – Nbg. By the formula for propagation of error, the standard deviation of the net (or background-subtracted) counts σN equals
N where σg =
gbg 22
σbg = 200 42 4 14 14
22 ..
1800 = 42.4 and
= 14.1. Therefore, the standard deviation σN of the net count counts equals = = 45 counts, and the standard deviation of the net count rate equals
45 counts / 2 minutes = 22 cpm.
propagation of errors!
The American College of Radiology (ACR) phantom, or “Jaczszak,” pictured here is used to evaluate which performance parameters of SPECT systems? A. count rate, spatial resolution, contrast B. spatial resolution, count rate, sensitivity
C. uniformity, contrast, spatial resolution D. sensitivity, uniformity, center-of-rotation alignment
C. uniformity; contrast; spatial resolution
The Jaczszak phantom is widely used to evaluate the overall system performance of SPECT systems. The phantom includes a plain section for the evaluation of the tomographic system uniformity, a section containing non-radioactive, or “cold,” spheres ranging from 9.5 to 31.8 mm in diameter for evaluation of contrast, and a section containing solid (non-radioactive) rods ranging from 4.8 to 12.7 mm in diameter for evaluation of spatial resolution.
A 10-ml syringe containing 2-mCi of technetium-99m (99mTc) in 1 ml is assayed in a dose calibrator. An additional 1 ml of plain saline is added to the syringe, and the syringe is assayed again. This process is repeated, adding 1-ml aliquots of plain saline to the syringe and re-assaying the syringe after the addition of 1 ml aliquot until the total volume in the syringe is 10 ml. These measurements are used in the evaluation of the _____ dose calibrator test. A. geometry B. linearity C. constancy D. accuracy
A. geometry
Dose calibrators are required to be tested for geometry dependence at installation, after major repair, or after being relocated. This is done by adding successive aliquots of plain saline or water to a 1-ml radioactive solution in a syringe or vial and assaying its activity before and after each addition of plain saline or water. Over a total volume range of 1 to 10 ml, the activity readings should not differ by more than 10%.
A 1-microCi iodine-129 (129I) standard source is measured in a scintillation well counter and yields a gross, or total, count rate of 5000 counts per minute (cpm). A background count rate of 100 cpm is then measured. _____ is the efficiency of the well counter for 129I. A. 20% B. 10% C. 1% D. 0.2%
D. 0.2% The efficiency ε of a counting system is defined as follows:
epsilon=(Cradionuclide-Cbackground)/Aradionuclide * 100%
where Cradionuclide is the gross count rate (in cpm) of the standard source, CBackground is the background count rate (in cpm), and Aradionuclide is the activity of the standard source (in disintegrations per minute, dpm). In this example, Aradionuclide = 1 microCi = 37,000 Bq = 37,000 decays/sec × 60 sec/min = 2,220,000 decays per minute (dpm). Therefore, ε = 0.0022 ×
100% = 0.22%. Importantly, this value of the efficiency applies only to this specific isotope, 129I, counted with the particular counter energy window used for this measurement.