RAPHEX IX Flashcards
mean free path
1/u
convert Sv to rem
1 Sv = 100 rem
The inherent filtration in a diagnostic x-ray tube results from _____ and is usually specified in terms of _____. A. metal sheets interposed between the x-ray tube and the patient; mm Cu equivalent. B. metal sheets interposed between the x-ray tube and the patient; mm Al equivalent. C. metal sheets interposed between the x-ray tube and the patient; mm W equivalent. D. components of the x-ray tube itself; mm Cu equivalent. E. components of the x-ray tube itself; mm Al equivalent.
e
Typical x-ray transformers for general radiography have a power rating of about _____
100 kW
Which one is CT?
C.
Graph C is a single x-ray energy spectrum with a maximum x-ray energy of 120 keV, which would be produced when the x-ray tube voltage is set to 120 kV. An x-ray tube voltage of 120 kV is typical for CT scanning.
-no ripple
-low energy photons filtered out
which graph is for P-A chest radiograph?
A
The energy spectra shown in Graph A would be for a tube voltage setting of 100 kV and tube current settings of 300 and 600 mA. Graph A is a better choice than Graph D because posterior-anterior (P-A) chest radiographs are typically obtained with a higher tube voltage (kV) and current (mA) to yield a greater x-ray intensity and, consequently, a shorter exposure time. The spectrum in Graph D corresponds to a relatively low tube voltage of only 80 kV
The intensity of an x-ray beam at the patient entrance surface would be increased most by _____. A. increasing tube current from 100 mA to 125 mA B. increasing tube voltage from 70 kVp to 85 kVp C. increasing exposure time from 0.6 second to 1.0 second D. decreasing the distance from the target to the patient surface from 120 cm to 60 cm E. decreasing filtration by 1 HV
D. decreasing the distance from the target to the patient surface from 120 cm to 60 cm
Due to the inverse square law, exposure will increase by a factor of 4 (22) when the target-to-patient distance is reduced by a factor of 2. X-ray intensity and exposure increases linearly with tube current and, therefore, the tube current increase in answer A will only increase patient exposure by a factor of 125 mA / 100 mA = 1.25. X-ray intensity increases as the square of the tube voltage and, therefore, the voltage increase in answer B will increase exposure by a factor of about (85/70)2 = 1.47. Decreasing filtration by one half-value layer (HVL) will increase x-ray intensity and exposure by a factor of 2
The image below illustrates a feature now commonly utilized in CT scanning. ANSWER
This feature is known as _____.
tube-current modulation
To compensate for different degrees of attenuation by different anatomic regions of the body while maintaining patient radiation doses as low as practical, the CT scanner’s x-ray tube current is varied. It increases in regions of greater attenuation (such as the shoulders) and decreases where attenuation is not as great (as in the lungs), providing dose reduction while maintaining image quality (Z-modulation). It also may decrease in anterior-posterior (AP) projections and increase in lateral projections (angular modulation), as seen by the ripples in the tube current (mA) profile.
Mammography units that can perform digital breast tomosynthesis are most likely to have a tube target composed of _____.
tungsten
For tomosynthesis units, _____ is (are) commonly used as the x-ray tube filter. A. silver (Ag) B. aluminum (Al) C. rhodium (Rh) D. All of the above are true. E. None of the above is true.
All of the above are true.
As indicated in the table above, the x-ray units for digital breast tomosynthesis used for DBT (as well as full-field digital mammography) utilize an aluminum (Al), rhodium (Rh), or silver filter.
An image intensifier with a 16-inch-diameter input screen, a 1-inch-diameter output screen, and a flux gain of 60 will have an overall brightness gain of about _____.
15,300
The brightness gain of an image intensifier is the product of the flux gain (60) and the minification gain (162), (16 in / 1 in)2 60 = 256 60 = 15,360 15,300. Note that the minification gain equals the square of the ratio of the linear dimensions of the input and output screens (in this case, 16 inches and 1 inch, respectively).
Compared to that of a digital mammography detector, the linear dimension of a detector element (del) of a flat-panel fluoroscopy detector is _____. A. about half the size B. about 25% smaller C. about the same size D. about 25% larger E. about twice as large
E. about twice as large
The detector element (del) size (linear dimension) of a digital mammography detector is about 0.05 to 0.1 mm or smaller, while that of a flat-panel fluoroscopy detector is about 0.14 to 0.2 mm in size (i.e., about twice as large)
If a CT scan is performed with 64 slices per rotation, a slice width of 0.625 mm, and a table increment of 6 cm per rotation, the pitch is _____.
1.5
For a CT scan, the pitch is the table increment divided by the product of the slice width and the number of slices. In this case, the pitch = 6 cm / (0.625 mm 64) = 60 mm / 40 mm = 1.5
In the American College of Radiology (ACR) CT Accreditation Program, which of the following types of examinations is allowed to have the lowest contrast-to-noise ratio (CNR) and receive a passing score? A. adult head B. adult abdomen C. pediatric head D. pediatric abdomen
pediatric abdomen
Images of the American College of Radiology (ACR) CT accreditation phantom acquired with the pediatric abdomen technique must have a contrast-to-noise ratio (CNR) of at least 0.4. Images of this phantom acquired with the pediatric head technique must have a CNR of at least 0.7, and those acquired with adult head and abdomen settings must have a CNR of at least 1.0. (Reference: American College of Radiology CT Accreditation Program Requirements, American College of Radiology, Reston, VA, 201
The U.S. Food & Drug Administration (FDA) regulatory limit for the entrance air kerma (Kair) rate in fluoroscopy is _____. A. 0.1 R/min B. 0.5 R/min C. D.
1 R/min 5 R/min
E. 10 R/min
10 R/min
The focal-spot size of the x-ray tube in a fluoroscopy system
0.6 mm
Compared to conventional (i.e., non-pulsed) fluoroscopy, pulsed fluoroscopy at a frame rate of 30 frames per second _____ motion blurring and _____ patient radiation dose. A. increases; reduces B. reduces; does not change C. does not change; reduces D. reduces; reduces
reduces; does not change
In pulsed fluoroscopy at 30 frames per second (fps), the x-ray tube current (i.e., mA) is only on for a short time, but it is increased to maintain the same exposure (i.e., mAs) per frame, as in continuous fluoroscopy. This reduces motion blurring (since the x-ray exposure for each frame is of shorter duration) but does not reduce or otherwise change the patient radiation exposure (since mAs per frame is uncha
The U.S. Food & Drug Administration requires that all fluoroscopes sold in the United States after May 2006 have which of the following indicators related to radiation dose from a fluoroscopic procedure? A. effective dose B. peak skin dose C. focal spot-to-skin distance D. total air kerma at the reference point
total air kerma at the reference point
The breast radiation dose in breast tomosynthesis is _____. A. about one-tenth of that from conventional mammography B. about one-third of that from conventional mammography C. about the same as that from conventional mammography D. three times that from conventional mammography E. 10 times that from conventional mammography
about the same as that from conventional mammography
In graph below of multiple x-ray energy spectra, what x-ray tube parameter was changed in transitioning from the top spectrum to the bottom spectrum
A. tube current (mA) B. tube voltage (kV) C. filtration material D. target material E. acquisition time
B. tube voltage (kV)
In fluoroscopy, the size of the field of view (FOV) can be changed to improve spatial resolution. In changing the FOV from the 30-cm mode to the 23-cm magnification mode, the exposure rate will _____. A. decrease by a factor equal to the square of the FOV ratios, (23 cm / 30 cm)2 B. increase by a factor equal to the square of the FOV ratios, (30 cm / 23 cm)2 C. be unchanged D. be reduced by approximately one-half
increase by a factor equal to the square of the FOV ratios, (30 cm / 23 cm)2
A technique that can be used to help limit patient dose in fluoroscopy is _____. A. increasing the frame rate B. decreasing the field of view C. decreasing the frame rate D. opening the collimation E. using cine mode instead of fluoroscopy
C. decreasing the frame rate
Pulsed fluoroscopy may help reduce patient radiation dose by reducing the duration of the “beam-on” time. By lowering the frame rate, the dose to the patient dose is typically decreased.
Pinhole collimators have long been used in nuclear medicine to produce magnified images of small structures, such as the thyroid. A patient is positioned near the pinhole aperture such that the distance from the patient’s neck to the pinhole aperture is less than the collimator cone length, yielding a magnified image. If the patient is moved farther away from the collimator, how do the detection sensitivity and spatial resolution change? A. The sensitivity increases and the resolution decreases. B. The sensitivity decreases the resolution increases. C. The sensitivity decreases and the resolution decreases. D. The sensitivity increases and the resolution increases. E. The sensitivity remains the same and the resolution decreases.
The sensitivity decreases and the resolution decreases.
As the distance of the object from the pinhole aperture is increased, the number of photons passing through the pinhole is reduced, reducing the sensitivity. Resolution decreases as the source-to-collimator distance is increased.
remember clara that patient is the source
M= SID/SOD
As one proceeds from low-energy to medium-energy to high-energy collimation for gamma camera imaging, _____. A. the collimator septa are made longer and thicker B. the collimator apertures are made wider C. the system sensitivity decreases D. the system spatial resolution worsens E. All of the above are tr
E. All of the above are true.
As one images radionuclides emitting x-rays and gamma rays with higher and higher energy, the mean path length through the septa must be increased to reduce the septal penetration (and resulting mispositioning) of the increasingly energetic radiations to an acceptably low level. This is accomplished in practice by making the collimator septa both longer and thicker. To compensate, at least in part, for the reduced number of radiations reaching the crystal, the apertures are widened. Fewer photons nevertheless reach the crystal, resulting in lower system sensitivity. At the same time, a spatial resolution is degraded because of the wider apertures.
The most likely cause of the gamma camera uniformity (or “flood”) image shown here is _____. A. a crack in the crystal B. an uncoupled PMT C. one or more software corrections being outdated, corrupted, or turned off
D. radioactive contamination E. incorrect collimation
C. one or more software corrections being outdated, corrupted, or turned off
A pattern of nonuniformity artifacts over the entire crystal (such as the PMT artifacts in the image shown) is very likely due to one or more software corrections being outdated, corrupted, or turned off.
A new technologist attempts to count a sample containing 200 mCi of carbon-14, a pure beta-particle emitter, in a sodium iodide (NaI(Tl)) scintillation well counter. What will be the result? A. The observed net count rate will be accurate. B. The observed net count rate will overestimate the true count rate. C. The observed net count rate will underestimate the true count rate. D. The observed net count rate will be zero.
The observed net count rate will be zero.
Carbon-14 emits only “soft” (i.e., low-energy) beta-rays that are so non-penetrating as to be completely absorbed in much less than a one-mm thickness of any solid or liquid. Therefore, carbon-14 beta rays would likely be absorbed by any sample being counted, and they would certainly be completely absorbed by the housing of the crystal in a scintillation well counter. For this reason, pure soft beta-ray emitters such as hydrogen-3 and carbon-14 can only be counted by the destructive technique of liquid scintillation counting, in which the radioactive sample is completely solubilized and mixed in solution with an organic scintillator molecule. On the other hand, radionuclides such as phosphorus-32, strontium-89, and yttrium-90—which emit much harder (i.e., higher-energy, >~1 MeV) beta-rays—may be counted in a scintillation well counter by counting the bremsstrahlung (brake x-rays) that these harder beta-rays produce as they pass through matter
In preparation for radioembolic therapy of liver cancer by infusion of yttrium-90 (90Y)-labeled microspheres via the hepatic artery, a pre-therapy imaging study was performed using technetium-99m (99mTc)-macroaggregated albumen (MAA) to estimate lung shunting. The images of and liver and lung region-of-interest (ROI) counts were as follow
liver has 485,912 counts, lung has 5,222
what is lung shunting?
1%
The lung shunt is calculated as the ratio of the 99mTc-MAA counts in the lung region of interest (ROI) divided by the sum of the counts in both the lung and liver ROIs multiplied by 100%. In this case, [5,222 / (5,222 + 485,912)] 100% = 1%.
In lutetium-177 (177Lu) dotatate (LUTATHERATM) treatment of neuroendocrine tumors, the beta-rays from 177Lu in a lesion account for _____ percent of the radiation dose to the lesion. A. 0% B. <10% C. 25–50% D. >75%
D. >75%
Lutetium-177 (177Lu) beta particles, like beta particles and other particulate radiation generally, have a very short range (about 1 mm or less) in tissue. Therefore, they deposit all of their energy and deliver all of their radiation dose within 1 mm or less than their point of emission. Thus, -rays emitted by 177Lu within a neuroendocrine lesion will deposit virtually all of their energy and deliver virtually all of their dose to the lesion itself. The gamma-rays emitted by 177Lu, like other high-energy x-rays and gamma-rays, travel many centimeters in tissue, and they will deposit much of their energy and deliver much of their dose outside the lesion, contributing very little to the lesion dose.
The current regulation for radionuclide therapy patients states that such patients can be released from the hospital when the projected dose to individuals around the patient is less than or equal to _____. A. 100 mrem B. 500 mrem C. D.
1 rem 5 rem
E. 10 rem
B. 500 mrem
Which of the following are theranostic radionuclides used in nuclear medicine? A. fluorine-18 (18F) B. yttrium-90 (90Y) C. iodine-123 (123I) D. lutetium-177 (177Lu) E. None of the above is true.
D. lutetium-177 (177Lu)
As its name implies, a theranostic radionuclide can be used for diagnostic imaging and for internal radiation therapy. Such a radionuclide must, therefore, emit one or more “image-able”
gamma-rays and beta-particles or other particulate radiations in reasonably high abundance. 177Lu satisfy both of these requirements and is used for imaging and therapy. The other isotopes listed are used exclusively for PET (18F), gamma camera imaging (123I), or therapy (90Y)
The gamma camera performance parameter known as multi-peak spatial registration can be measured using the _____ isotope. A. fluorine-18 (18F) B. technetium-99m (99mTc) C. iodine-123 (123I) D. iodine-131 (131I) E. indium-111 (111In)
Multi-peak (or multi-window) spatial registration is a gamma camera performance parameter that applies only to isotopes that emit multiple “imageable” gamma-rays, such as indium-111 (111In), which emits a 173-keV gamma-ray (90% abundance) and 247-keV gamma-ray (94%). All of the other isotopes listed are either a positron emitter used in PET (18F) or emit a single imageable gamma-ray (99mTc, 123I, 131I).