RAPHEX V Flashcards
Mammography x-ray tubes typically use a lower tube voltage (kVp) than those used in general radiography in order to _____. A. reduce dose to the patient B. reduce the image exposure time C. utilize characteristic x-rays exclusively D. reduce shadowing artifacts E. improve image contrast
A lower x-ray tube voltage (i.e., kVp) produced an x-ray beam with lower-energy x-rays than a higher tube voltage. The proportion of x-rays interacting in tissue or other media via the photoelectric effect rather than Compton scatter increases with decreasing x-ray energy. The lower tube voltage used in mammography systems thus improves image contrast by reducing the number of Compton-scattered photons contributing to the image. This strategy (i.e., the use of lower tube voltages) is not adaptable to general radiography, however, because lower-energy x-rays, while able to penetrate the breast, would not penetrate the torso or other thicker body parts in sufficient numbers to produce a statistically reliable image at an acceptably low radiation dose to the patien
The inherent filtration in a diagnostic x-ray tube results from _____ and is usually specified in terms of _____. A. metal sheets interposed between the x-ray tube and the patient; mm Cu equivalent B. metal sheets interposed between the x-ray tube and the patient; mm Al equivalent C. metal sheets interposed between the x-ray tube and the patient; mm W equivalent D. components of the x-ray tube itself; mm Cu equivalent E. components of the x-ray tube itself; mm Al equivalent
E
A hypothetical x-ray beam has a uniform energy spectrum, as shown below. Which of the lettered graphs best represents the beam’s energy spectrum after passing through and exiting a uniform slab of polystyrene? Assume photoelectric effect is the dominating absorption mechanism.
Remember uniform spectrum doesn’t mean mono-energetic! - just means same number of photons at all energies
-beam hardening occurs
The image below illustrates a feature now commonly utilized in CT scanning. This feature is known as _____.
ANSWER
A. spectral filtration B. adaptive beam collimation C. cardiac-motion tracking D. iterative reconstruction E. tube-current modulation
e- look up what it looks like
To compensate for different degrees of attenuation in different anatomic regions of the body, and to maintain patient radiation doses as low as practical, the CT scanner’s x-ray tube current varies. It increases in regions of greater attenuation (such as the shoulders) and decreases where attenuation is not as great (as in the lungs), providing dose reduction while maintaining image quality (Z-modulation). It also may decrease in anterior-posterior (AP) projections and increase in lateral projections (angular modulation), as seen by the ripples in the tube current (mA) profile.
In an x-ray tube, the line focus principle is used to make the effective focal spot _____ the actual focal spot, which _____ the power loading of the anode. A. smaller than; increases B. smaller than; decreases C. the same as; maintains D. larger than; increases E. larger than; decrease
A. By angulating the anode with respect to the direction of travel of the incident electrons, the effective focal spot size will be smaller than the actual focal spot size. This improves spatial resolution by reducing penumbra blurring while dispersing the incident electron beam and associated heat deposition over a larger area of the anode
A prostate-specific membrane antigen (PSMA)-binding small molecule labeled with lutetium-177 (physical half-life, Tp = 6.65 days) is eliminated from the body of a prostate cancer patient with a biological half-life, Tb, of 1.20 days. What is the effective half-life, Te, of lutetium-177 in this patient’s body? A. 1.02 days B. 1.20 days C. 6.65 days D. 7.98 days E. 8.85 days
A. A radioactive material administered to a patient is eliminated from a tissue or organ of the
patient by the simultaneous processes of radioactive decay (with physical half-life Tp = 6.65 days in this example) and metabolism, secretion, excretion, etc. (with biological half-life Tb = 1.20 days in this example). The overall elimination of the radioactivity is more rapid than if it were eliminated only by radioactive decay or only by biological processes, and the resulting effective half-life (Te) is given by the following formula
A thyroid cancer patient is imaged after iodine-131 therapy. What is the cause of the star artifact shown below?
A. A metal object was in the field of view. B. The photopeak energy window was set to that for technetium-99m instead of that for 131I.
C. The collimator septa was penetrated by high-energy 131I gamma-rays.
D. There was crystal damage. E. There was a non-functioning photomultiplier tube (PMT)
C.
In addition to the imageable 364-keV gamma-ray emitted by iodine-131 in 81% of its decays, iodine-131 emits a 637-keV gamma-ray in 7.3% of its decays. This high-energy photon easily passes through the collimation designed for the 364-keV gamma-ray and, when scattered, may appear in the photopeak energy window set for the 364-keV photon, producing the observed star artifac
Assuming the same crystal and collimator pair above, if the source is moved such that the distance between the source and the collimator is increased, what will be the effect on the resolution and sensitivity? A. The spatial resolution improves and the sensitivity decreases. B. The spatial resolution worsens and the sensitivity increases. C. The spatial resolution improves and the sensitivity increases. D. The spatial resolution worsens and the sensitivity stays the same. E. There will be no effect at all.
As a source is moved farther away from the collimator, the inverse-square law results in a lower number of photons reaching the detector. However, this is counterbalanced by the increased total aperture area of the collimator, through which photons can pass. Therefore, the sensitivity does not change. The spatial resolution is best with the source in contact with the collimator. As the source-to-detector distance is increased, photons strike the crystal over a larger area, resulting in degradation of spatial resolution.
In a C-arm type of fluoroscope having a source-to-image distance (SID) less than 45 cm (a mini C-arm), the air-kerma rate (AKR) shall be measured at _____. A. the location specified in the manufacturer’s manual B. 30 cm from the focal spot C. 30 cm from the input surface of the fluoroscopic imaging assembly D. minimum SSD E. 1 cm above the input surface of the fluoroscopic imaging assemb
D. Based on Code of Federal Regulations (CFR) 1020.32 (d)(3)(iv), in a C-arm type of fluoroscope having an SID less than 45 cm, the AKR shall be measured at the minimum SSD
For an air-kerma rate (AKR) exceeding 6 mGy/min, the displayed values shall not differ from the actual values by more than _____. A. the tolerance specified by the manufacturer B. 10% C. 2 mGy/min D. 5% for the range from 6 to 25 mGy/min and 10% for an AKR exceeding 25 mGy/min E. 35%
E.
Based on Code of Federal Regulations (CFR) 1020.32 (k)(6), the displayed AKR shall not deviate from the actual values by more than ±35% over the range of 6 mGy/min to the maximum indication of AKR.
In fluoroscopy, “grid controlled” refers to _____. A. a rapid control of the tube output by a grid inside the tube B. elimination of scatter based on the proper selection of the scattering grid C. dynamic variation of the Bucky speed based on the AKR. D. a fixed selection of fluoro protocol parameters based on the procedure and the patient thickness
E. an evaluation of the SNR as a function of patient thickness
A.
The units for dose-area product (DAP) and air-kerma rate (AKR) are _____ and _____, respectively. A. mGy; mGy/sec B. mGy-cm2; mGy/min C. R-cm2; mGy/R/sec D. mGy-cm; mGy-cm/min E. mGy/R-cm2; R/sec
B. mGy-cm2; mGy/min
The benefit of using additional copper filtration in interventional fluoroscopic units is to _____. A. create a more homogenous x-ray field by attenuating slightly more in the center B. reduce overall patient dose by augmenting the spectrum with the copper K_alpha lines C. reduce patient dose by entrance skin sparing D. soften the contrast to enhance low-visibility blood vessels E. enhance the contrast in iodine contrast studies
C.
A copper filter (e.g., 0.11 mm thick) will significantly reduce the number of low-energy photons, which, based on their energy, are mainly absorbed in the patient’s skin. Therefore, this is used to achieve skin sparing.
Which of the following is not a regulatory requirement for a stationary C-arm type of fluoroscope having a source-to-image distance (SID) less than 45 cm (a mini C-arm)? A. It must be labeled “For extremity use only.” B. The SSD must not be less than 10 cm. C. The fluoroscopic irradiation time has to be displayed within 6 seconds of termination of an exposure.
D. The display of the air-kerma rate (AKR) shall be clearly distinguishable from the display of the cumulative air kerma.
E. During cinefluoroscopy, the AKR shall not exceed 44 mGy/min.
E. There is no regulatory limit specified for AKR in cinefluoroscopy
equation for DQE
(SNRout/SNRin)^2
The brightness gain in a typical fluoroscopy system (12-inch diameter input phosphor) is approximately _____. A. B. C. D.
7.2 72 720 7,200 E. 72,000
D.
Brightness gain = electronic gain × minification gain. In a typical system, the electronic gain is about 50 and the minification gain for a 12-inch input phosphor to a 1-inch output phosphor is 144. Therefore, the brightness gain is 7200.
kVp and mAs- how are they adjusted for a large patient?
kVp increase will spare dose to patient but cotnrast will be worse
increasing mAs and keeping kVp low will be good contrast but more dose
The heel effect can be reduced by _____. A. decreasing the anode angle B. increasing the source-to-image distance (SID) C. increasing the tube current (mA) D. increasing the tube voltage (kVp
B. The heel effect describes the decrease in x-ray intensity toward the anode of the x-ray tube due to the increase in self-filtration of the x-ray beam by the anode itself. It can be reduced by increasing the source-to-image distance (SID). The decrease in beam intensity with distance (due to the inverse square law) will minimize the variation in intensity due to heel effect
- I would argie that d is also right? Or I guess ratio of what is attenuated vs not is still same??? - to me less is attenuated due to high kV so less heel effect
The speed of sound in imaging is affected by the tissue property of _____. A. compressibility B. impedance C. transmittance D. attenuation
A. The speed of sound in a particular medium is affected by the compressibility of the medium. As the compressibility increases, the speed of sound decreases. For example, the speed of sound in air, which is highly compressible, is 330 m/s, while the speed in soft tissue, which is far less compressible, is 1540 m/s.
-Impedance affects what gets reflected/transmitted
effective dose to adult from chest CT scan
~ 10 mGy
Flow void is more likely to occur for _____. A. T1-weighted MR images with slow blood flow B. T1-weighted MR images with fast blood flow C. T2-weighted MR images with slow blood flow D. T2-weighted MR images with fast blood flow
D. Due to blood flow, the excited protons in blood in the tissue cross-section corresponding to the MR image may have exited that tissue section by the time of readout. Also, moving protons (as in flowing blood) accumulate additional phase shift compared to stationary protons. For T2-weighted images, the loss in signal is due to the fact that the fast-moving blood will miss the 180o re-phasing pulse, which reduces the readout signal. T1 has a shorter echo time (TE), so there will be less de-phasing.
The American College of Radiology (ACR) reference dose for an adult head CT scan is _____. A. 15 mGy B. 25 mGy C. 35 mGy D. 40 mGy E. 75 mGy
E. The American College of Radiology (ACR) CT reference dose—actually, the volume CT dose index, CTDIvol—for an adult head study is 75 mGy (7.5 rad). Other ACR CT reference doses are, for example, as follows: pediatric abdomen, 15 mGy (1.5 rad); adult abdomen, 25 mGy (2.5 rad); pediatric head, 35 mGy (3.5 rad
The ACR pass/fail criterion for an adult head CT scan is _____. A. 20 mGy B. 30 mGy C. 35 mGy D. 40 mGy E. 80 mGy
E. The American College of Radiology (ACR) CT pass/fail criterion for an adult head study is 80 mGy (8.0 rad). Other ACR CT pass/fail criteria are, for example, as follows: pediatric abdomen, 20 mGy (2.0 rad); adult abdomen, 30 mGy (3.0 rad); pediatric head, 40 mGy (4.0 rad).
Multi-slice CT scanners are more advantageous than single-slice scanners because of _____. A. lower patient effective dose B. improved high-contrast spatial resolution C. decreased heat load on the machine D. increased scan time E. lower capital (i.e., purchase) cost of the scanner
C.
Decreased heat load on the machine is associated with multi-slice CTs compared to single-slice CT scanners. Patient effective dose is not reduced because the same patient volume is irradiated in both multi-slice and single-slice scanners. High-contrast spatial resolution is not improved. The scan time is decreased because data for more than one tissue section is collected per rotation. The purchase price of a multi-slice scanner is higher than that of a single-slice scanner.
For which of the following CT images would one utilize the widest display window? A. head B. chest C. abdomen D. pelvis
B. The wide range of linear attenuation coefficients and, therefore, Hounsfield units (HUs) among the tissues in the chest—lung, soft tissue, and bone—generally requires a wide display window for optimal visualization of all such tissues
How is the noise in a CT image affected when the patient is scanned with twice the slice thickness and one-half of the mA, assuming all other parameters are the same? A. It remains the same. B. It decreases by a factor of 2 times. C. It decreases by a factor of 4 times. D. It decreases by a factor of 8 times. E. It decreases by a factor of 2 times
A. The CT noise decreases by √2 by doubling the slice thickness. The CT noise increases by √2 by using only one-half of the mA. Therefore, since √2 / √2 = 1, there is no change in the CT image noise.
accuracy of a diagnostic test
(TP + TN) / (TP + FP + TN + FN)
In fluoroscopy, the greatest radiation exposure to the fluoroscopist is _____. A. the exit beam from the patient B. scatter from the patient C. leakage radiation from the x-ray tube D. scatter from the image intensifier
B.
At fluoroscopic x-ray energies, there is considerable side (i.e., 90-degree) scatter from the patient toward the fluoroscopist. The exit-beam dose from the patient is smaller due to attenuation by the patient. The leakage radiation is very small due to federal design requirements. Scatter from the image intensifier would also be a very small fraction of the exit beam from the patient.
The patient scatter at the position of the fluoroscopist from a certain fluoroscopic procedure is 5.5 Gy of air kerma. The fluoroscopist is standing 50 cm from the patient, and the area of the fluoroscopy x-ray field entering the patient is 20 25 cm. If the x-ray field is enlarged to 25 30 cm, the scatter reaching the fluoroscopist would be approximately _____. A. B. C. D.
3.2 Gy 5.5 Gy 8.3 Gy 9.6 Gy
E. 11.3 Gy
C. The amount of scatter radiation produced is proportional to the area of the x-ray field entering the patient, so the new amount of scatter will be [(25) (30)] / [(20) (25)] 5.5 Gy = 1.5 5.5 Gy = 8.25 Gy ≈ 8.3 Gy.
Changing the field of view of an older image intensifier tube from 30-cm mode to 15-cm mode will result in a change in spatial resolution from 1.2 line pairs (lp)/mm to _____. A. 0.3 lp/mm B. 0.6 lp/mm C. 1.2 lp/mm D. 2.4 lp/mm E. 4.8 lp/mm
D.
The spatial resolution of the image from the image intensifier will vary approximately as one over the diameter of the image. Therefore, changing the field of view of an image intensifier from 30 to 15 cm changes the resolution from 1.2 lp/mm to (30 cm / 15 cm) 1.2 lp/mm = 2 1.2 lp/mm = 2.4 lp/mm.
A fluoroscopy unit with a flat-panel detector image intensifier (II) has a spatial resolution of 1.0 line pair/mm (lp/mm). Changing the field of view from 25-cm mode to 15-cm mode will result in a spatial resolution of _____. A. 0.5 lp/mm B. 1.0 lp/mm C. 1.5 lp/mm D. 2.0 lp/mm E. 3.0 lp/mm
B.
The spatial resolution of the image for a flat panel detector does not change with field size unless binning is employed, typically for field sizes of 30 cm and larger.
The maximum allowed patient input air kerma rate with high-level control activated is _____. A. 88 mGy/min B. 176 mGy/min C. 350 mGy/min D. 700 mGy/min E. There is no upper limit
B. The maximum allowable air kerma rate (AKR) is 176 mGy/min (or 20 R/min input exposure rate.