radiation output consistency test Flashcards
1
Q
what is the radiation output consistency test
A
measures the output of the xr tube
2
Q
how frequent is it conducted
A
every month
3
Q
equipment used
A
output meter
4
Q
the setup
A
- ensure tube and table are level
- place output meter on table
- switch on the meter and reset
- centre tube to meter and open collimation to include them
- expose
- record kV, time and dose
- repeat expose and record to compare baseline values
- repeat steps using the set exposure for a chest examination and lumbar spine examination
5
Q
if recorded values exceed +-10%
A
of baseline values, contact RPA/ MPE