Pipet Calibration Semi Annual Flashcards
What is the pipet volume for the semi-annual calibration?
300UL
This refers to the volume capacity of the pipette being calibrated.
What is the serial number of the pipet?
A68031022
This unique identifier helps track the specific pipette for calibration records.
What is the accuracy of the balance used in the calibration?
+0.001g
This indicates the precision of the balance in measuring weight.
What is the method used for volume testing?
Gravimetric Volume Testing
This method involves weighing a liquid to determine its volume.
What is the target volume for the gravimetric volume testing?
0.300 mL
This is the volume of deionized or distilled water used for testing.
What are the individual weight limits for the pipet calibration?
+/-5%: 0.285 - 0.315 g
This range indicates acceptable weight variations during testing.
Where is the pipet located?
Receiving Bay, Room # 1
This specifies the physical location of the pipette in the lab.
What was the temperature at the time of calibration?
25°C
Temperature can affect liquid density and measurement accuracy.
What is the average weight recorded during the calibration?
0.300 g
This average weight indicates that the pipette dispenses the correct volume.
What is the pass/fail result of the calibration?
Pass: V
This indicates that the pipette calibration met the required standards.
What is the next due date for calibration?
07/02/10
This indicates when the next calibration is scheduled.
Fill in the blank: The balance serial number is _______.
1117132085
This unique identifier is essential for maintaining quality assurance.
What is a Traceable Thermometer checked at annually?
Ice Point
The thermometer is tagged with the date, identification, temperature checked, and correction information.
What is the accuracy of the Digital Thermometer?
+/- 2°F / +/- 1°C
This refers to the precision in temperature measurement.
What is the resolution of the Digital Thermometer?
0.1°F / C
This indicates the smallest change in temperature that can be displayed by the thermometer.
What is the temperature range of the 6” Alcohol Thermometer?
58° to 300°F / -50° to 150°C
This defines the minimum and maximum temperatures that can be measured.
What is the sampling time for the Digital Thermometer?
1.5 seconds
This is the time it takes for the thermometer to provide a reading.
How often should the temperature of the incubator be checked?
Daily
Regular checks ensure that the incubator maintains the proper temperature for optimal conditions.
Where can the temperature be checked on the Charm EZ?
Above the Charm EZ label
This is the designated area for monitoring temperature on the device.
Where can the temperature be checked on an external incubator?
Back of the incubator beneath the hinges
This location is specified for external incubator monitoring.
If the temperature is out of range, whom should you contact?
Charm Technical Support
This ensures that any issues are addressed promptly by qualified personnel.
Fill in the blank: The Digital Thermometer uses a _______ battery.
1.5 volt button
This specifies the type of power source used for the thermometer.
What is the maximum allowable temperature for raw milk TC samples upon receipt?
4.4° C
Samples exceeding this temperature should not be tested.
What temperature must the sample be maintained below?
7° C
It is crucial to ensure the sample temperature remains below this limit.
What should be done if a raw milk TC sample is overfilled?
Do not accept the sample
Overfilled samples can lead to inaccurate testing results.
What is required for proper mixing of the sample?
Sufficient head space
Allowing head space is important to facilitate effective mixing.
What is the first instruction regarding the test strip container?
Always reseal container with this lid immediately after opening
This ensures the integrity of the reagents.
What is the storage requirement for the sealed container?
Refrigerate
Proper refrigeration is necessary to maintain reagent stability.
What should be recorded in the reader and on the daily antibiotic log sheets?
Lot # of the test strips
This tracking is essential for quality control and traceability.
What color should the desiccant be?
Blue
Indicates that the desiccant is functioning properly.
How should the test strips be prepared for use?
Set out number of strips to be used and place them in a dry labeled container at room temperature
This preparation helps to avoid moisture contamination.
What should be done with any leftover test strips at the end of the day?
Discard any leftover
This prevents the use of expired or contaminated strips.
What is the sample preparation time for the Beta-lactam Test?
Milk sample and controls 25 times in 7 seconds
This refers to the rapid preparation required for accurate testing.
What should be done to avoid foam when preparing the sample?
Allow foam/bubbles to settle or tilt sample to avoid foam
Proper handling is crucial for accurate results.
How soon should the sample be used after shaking?
Use within 3 minutes of shaking
Timeliness is important to maintain sample integrity.
What is the time limit for pipetting the sample after placing the first strip in the incubator for SL3?
Pipet sample within 1 minute 15 seconds
This is critical for ensuring accurate test results.
What is the time limit for pipetting the sample after placing the first strip in the incubator for SLBL?
Pipet sample within 2 minutes
Adhering to this timeframe is essential for accuracy.
What should each test strip be labeled with?
Sample identification such as sample #, manifest or route #, and indicate F (Front) or R (Rear) if applicable.