A change in the strength of the main magnetic field can be caused by:
a. All are correct.
b. cryogen boiloff.
c. external ferromagnetic materials.
d. shim coil failure.
A
An acquisition with a 200-mm field of view and 256 x and y encodes yields a pixel size of ____ mm2.
0.78
Luminance uniformity of soft-copy displays used in MRI systems should be within ____ of the maximum brightness measured in the center of the screen.
30%
Nonuniformity in the magnetic field can be caused by:
the presence of ferromagnetic material within the bore of the magnet.
The ACR slice position deviation in MR scanners should not exceed ____ mm.
5
The ability of an MR scanner to resolve small objects is known as:
high-contrast resolution.
The ability of the MR system to differentiate between adjacent tissues having minimal differences in signal intensities is known as:
low-contrast resolution
The amount of geometric distortion in the image is the:
spatial linearity.
What is the maximum luminance for soft-copy display devices as suggested by the ACR?
Greater than 90 nit
What is the resonant frequency in a unit with a 1.0-T magnet and a gyromagnetic ratio of 42.57 MHz/T?
42.57 MHz