According to ACR Mammography Accreditation standards, what is the minimum frequency for evaluating the compression thickness indicator by the technologist?
Monthly
During ACR phantom scoring, which of the following will result in a full point being awarded when viewing the fibers?
≥8 mm long
During ACR phantom scoring, which of the following will result in a full point being awarded when viewing the specks?
4-6 specks
During ACR phantom scoring, which of the following will result in a full point being awarded when viewing the masses?
≥¾ border
Mammographers must perform phantom image evaluation at least
Weekly
The MQSA does not require facilities to have:
ACR accreditation
The average glandular dose delivered during a single craniocaudal view of an FDA-accepted phantom simulating a standard breast shall not exceed ____ mGy per exposure.
3
The mAs recorded during phantom image testing of mammographic equipment should not vary by more than ____ from the initial control value.
±15%
The maximum allowable dose per exposure for mammographic images is ____ rad.
0.3
The recommended maximum tolerance of kVp variance for mammographic x-ray generators is
5%