A visual inspection of a CT unit should be evaluated at least _____ for ACR accreditation programs.
monthly
For ACR CT accreditation, the Water CT Number and Standard Deviation (Noise) should be evaluated by the CT technologist
daily
For a slice thickness of 5 mm or less, the slice thickness should not vary by more than ____ mm.
0.5
Patient dose values for CT scans should remain within ____ of the manufacturer’s specifications.
±10%
Section thickness should not vary by more than ____ mm for sections greater than 5 mm.
±1.0
The CT number of water should not deviate by more than ____ CT numbers from 0.
3
The CTDIvol pass/fail criterion for an adult abdomen is ____ mGy.
30
The laser light field and radiation field should coincide to within ____ mm.
2.0
The minimum distance between two objects that allows them to be seen as separate and distinct would define what CT imaging parameter?
High-contrast resolution
The uniformity of the CT number of water should not vary by more than ____ HU from the center of the phantom to the periphery.
5