Final Final - Rad II Flashcards
Overall image blackness
Density
The visibility of fin anatomical structures is
Detail
Changes in human anatomy that result in decreased tissue density
Destructive pathology
The area where electrons impact the anode
True focal spot
All of the following are synonymous with detail
Resolution, definition, sharpness
NOT contrast
The photon strength in the polyenergetic primary beam is expressed as
KVP
The disparity in xray intensity between cathode and anode ends of the X-ray tube is known as
Anode heel effect
The number of electrons thermionically emitted from the filament is known as
MA
The total number of electrons traveling from cathode to anode is measured as
MAS
To manipulate the density/brightness of a digital image the tech will adjust
Window level
When the patient contributes to the image contrast is known as
Subject contrast
The number of X-rays in the polyenergetic primary beam is expressed as
MAS
The true focal spot projected towards the patient is known as
Effective focal spot
Materials that temporarily decrease the atomic number of an anatomical structure are known as
Negative contrast agents
Intentionally magnifying the area of anatomical interest is known as
Macroradiography
Changes in human anatomy that result in increased tissue density are known as
Additive pathology
An image that is mainly black and white and possesses few grays exhibits
High contrast
The portion of the radiographic image that represents an objects true size is known as
Umbra
To manipulate the contrast of a digital image the tech must adjust the
Window width
The height of the lead strips related to the distance between the lead strips is the formula for calculating
Grid ratio
The minimum xray energy that will adequately penetrate a given body part is known as
Effective KV
Blooming of focal spot
Electrons as they move from cathode to anode
Exposure latitude
Margin for technique error
Size distortion
Too much OID magnified the object
Shape distortion
Penumbra
Quantum mottle
Grainy image as a result of not enough X-rays
Positive contrast agent
Materials that temporarily increase the atomic number of an anatomical structure
Long scale contrast
Many shades of gray
Differential absorption
Differing habitus absorb xray at differing rates
High contrast
Black and white image with few grays
The nine penny test is performed to evaluate
Xray to light field alignment
Focal spot size is allowed to vary no more than
+/-50%
The required frequency for performing protective apparel testing is
Annually
Exposure reproducibility is allowed to vary no more than
+/-5%
The required frequency for performing screen speed uniformity testing is
Annually
KVP accuracy is allowed to vary by no more than
+/-10%
The ESE for general fluoroscopy may not exceed
10R/min
The ESE for interventional fluoroscopy may not exceed
20R/min
Exposure timer accuracy for greater than 10ms exposures is allowed to vary by no more than
+/-5%
The density difference on a darkroom fog test general radiography may not be greater than
.08 OD
The developer temperature recorded during sensitometry is allowed to vary by no more than
+/-2 degrees F
The required frequency for performing sensitometry is
Semiannually
The required frequency for performing film/screen contact testing is
Annual
The required frequency for performing collimator dial accuracy testing is
Semiannual
With sensitometry contrast is allowed to vary by no more than
.15
Xray to light field alignment is allowed to vary by no more than
+/-2%
The standard light bulb for an xray illuminator box is
15 watt daylight fluorescent
SID accuracy is allowed to vary by no more than
+/-2%
The required frequency for performing reproducibility testing is
Annual
The required frequency for performing linearity testing is
Annual
The required frequency for performing HVL testing is
Annual
Focal spot size may be evaluated by
Pin hole camera
Slit camera
Star pattern
The required frequency for performing darkroom fog testing is
Semiannual
Exposure linearity is allowed to vary no more than
+/-10%
The wire mesh test is performed to evaluate
Film screen contact
The required frequency for performing filtration testing is
Annual
The minimum allowable total xray beam filtration is
2.5mm AL/EQ
The required frequency for performing KVP accuracy is
Annual
The density difference on a darkroom fog test for mammography may not be greater than
.05 OD
Exposure timer accuracy for less than 10ms exposures is allowed to vary no more than
+/-20%
The required frequency for performing exposure timer accuracy testing is
Annual
Screen speed uniformity is allowed to vary no more than
+/-10%
Collimator dial accuracy is allowed to vary no more than
+/-2% of SID
The light intensity across the entire surface of an xray view box is allowed to vary by no more than
+/-10%
When using positive beam limitation the light field is allowed to vary from the film size by no more than
+/-2% of the SID
The number of HVLs required to reduce a 78mR beam to less than 10mR is
3 HVLs
Latent image is made manifest in
Developer
Modern film screen imaging amount of film exposure due to visible light is
90-99%
Function of wash section is to
Remove fixer from the emulsion
Emission of visible light from a phosphor crystal is
Luminescence
Sequence of order for film processor is
Developer, fixer, washer, dryer
Brown tint that develops in processed xray film while in storage is
Smog
Color of visible light emitted by a phosphor crystal is
Spectral emission
Radiographic film contrast increases as film latitude
Decreases
Addition of an electron to a positively charged silver ion is
Reduction
Portion of sensitometric curve representing minimum optical density is
Base + fog
Invisible image held in chemical code within the films emulsion layer is
Latent image
Size if films silver halide crystals increases the films screen speed
Increases
Addition of electron to a positively charged silver ion results in
Black metallic silver
The wash section is filled with
Filtered water
The temperature of developer solution is
95 degrees
Stored film is sensitive to (can be fogged by)
Visible light, high temperature, and chemical vapor
Color of visible light film is most sensitive to known as
Spectral sensitivity
Rate at which sensitometry must be performed is
Daily