Exam 6 Flashcards
Which of the following is not an x-ray beam limitation device?
- Extension cylinder
- Collimator
- Filter
- Cone
Filter
If the luminance of the collimator light source is adequate, the localizing light beam will satisfactorily outline the margins of the radiographic beam on the area of clinical interest on
- children only.
- teenagers and adults only. - - infants only.
- all patients.
all patients.
When using digital fluoroscopy systems, making use of the last image hold feature can
- lead to a 50% reduction of technical exposure factors.
- be an effective dose reduction technique
- eliminate the need for body or part immobilization.
- eliminate the need to collimate the primary beam.
be an effective dose reduction technique
What is the purpose of radiographic beam filtration?
- To reduce exposure to the patient’s skin and superficial tissues by absorbing most of the lower-energy photons from the heterogeneous beam.
- To reduce exposure to the patient’s skin and superficial tissues by removing all the lower-energy photons from the homogeneous beam.
- To reduce exposure to the patient’s skin and superficial tissues by absorbing all of the higher-energy photons from the homogeneous beam.
- To reduce exposure to the patient’s skin and superficial tissues by absorbing most of the higher-energy photons from the heterogeneous beam.
To reduce exposure to the patient’s skin and superficial tissues by absorbing most of the lower-energy photons from the heterogeneous beam.
Depending on the area of the body being examined with a fluoroscopic image intensification system, a range of ___________ kVp is generally used for adult patients.
- 110 to 140
- 60 to 75
- 30 to 60
- 75 to 110
75 to 110
During a fluoroscopic examination, a resettable cumulative timing device measures the x-ray beam-on time and sounds an audible alarm or in some cases temporarily interrupts the exposure after the fluoroscope has been activated for what length of time?
- 10 Minutes
- 3 Minutes
- 1 minute
- 5 Minutes
5 Minutes
When obtaining a dorsoplantar projection of a foot, which of the following types of filters should be used to provide a greater uniform density of the anatomy?
- Thoraeus filter
- Bilateral wedge filter
- Wedge filter
- Trough filter
Wedge filter
Total filtration of _______________________ for fixed x-ray units operating above 70 kVp is the regulatory standard.
- 1.0-mm aluminum equivalent
- 2.0-mm aluminum equivalent
- 2.5-mm aluminum equivalent
- 1.5-mm aluminum equivalent
2.5-mm aluminum equivalent
In digital radiography the number of different shades of gray that can be stored in memory and displayed on a computer monitor is termed
- contrast matrix.
- grayscale.
- shadows.
- pixels.
grayscale.
The control panel, where technical exposure factors are selected and visually displayed for the equipment operator, must be located
- at the head end of the radiographic examining table.
- at the foot end of the radiographic examining table.
- behind any regular wall that has a regular glass window that permits observation of the patient during any procedure.
- behind a suitable protective barrier that has a radiation-absorbent window that permits observation of the patient during any procedure.
behind a suitable protective barrier that has a radiation-absorbent window that permits observation of the patient during any procedure.
Any image that must be performed more than once because of human or mechanical error during the production of the initial image is considered to be a repeat image. What effect does a repeat image have on the radiation dose received by the patient?
- The patient’s skin and possibly the gonads, if they were in the included imaged area, receive a double radiation dose.
- The patient receives no additional radiation dose.
- The patient’s superficial tissues only receive a radiation dose that is four times as great as that of the original radiation dose.
- The patient’s critical organs receive a radiation dose that is 10 times greater than the initial radiation dose.
The patient’s skin and possibly the gonads, if they were in the included imaged area, receive a double radiation dose.
During a radiographic examination, which of the following combinations of technical exposure factors reduce patient radiation dose?
- Higher kVp, lower mAs
- Lower kVp, lower mAs
- Lower kVp, higher mAs
- Higher kVp, higher mAs
Higher kVp, lower mAs
When a pregnant patient must undergo a radiographic procedure, which of the following practices will minimize radiation exposure?
- Opening the x-ray beam collimator shutters as wide as possible to ensure complete coverage of the image receptor
- Precisely collimate, and use appropriate technical exposure factors for the anatomical part
- Selecting technical exposure factors that are appropriate for the part of the body to be radiographed
- Precisely collimating the radiographic beam to include only the anatomic area of interest and shielding the lower abdomen and pelvis when this area does not need to be included in the area to be irradiated
Precisely collimate, and use appropriate technical exposure factors for the anatomical part
Which of the following are considered to be benefits of an aggressive repeat analysis program?
1. Increased awareness among staff and student radiographers of the need to produce optimal quality recorded images from the start.
2. Radiographers become more careful in producing radiographic images because they are aware that images are being reviewed.
3. When the repeat analysis program identifies problems or concerns, in-service education programs covering these specific topics may be designed for imaging personnel.
- 2 only
- 1 only
- 3 only
- 1, 2, and 3
1, 2, and 3
Which of the following must always be the first step in the protection of the reproductive organs?
- Use of a mobile protective shielding device
- Use of gonadal shielding
- Use of digital imaging equipment
- Adequate and precise collimation of the radiographic beam to include only the anatomy of interest
Adequate and precise collimation of the radiographic beam to include only the anatomy of interest