EXAM #5 Flashcards
[What is fluoroscopy]
_____ time viewing of anatomical structures
_____ picture x-ray, can stop and take static images,
Most of it IS/IS NOT captured
DOES/DOES NOT have the spatial resolution as normal x-ray (static image)
[What is fluoroscopy]
Real time viewing of anatomical structures
Motion picture x-ray, can stop and take static images,
Most of it IS NOT captured
DOES NOT have the spatial resolution as normal x-ray (static image)
3 Components of fluoroscopy
Image intensifier
TV
CRT tube
[Fluoroscopy]
Is flouro dynamic?
[Fluoroscopy]
YES
[Fluoroscopy]
Discovered in _____ by _____
[Fluoroscopy]
Discovered in 1896 by Edison
[Fluoroscopy]
Continuous energizing of x-ray tube is needed
[Fluoroscopy]
Continuous energizing of x-ray tube is needed
[Fluoroscopy]
What is something we can do to reduce heat in fluoroscopy
[Fluoroscopy]
Drop mAs
[Fluoroscopy]
Low mAs in fluoro = _____ images
[Fluoroscopy]
Low mAs in fluoro = noisy images
[Fluoroscopy]
DA stands for…
DSA stands for…
[Fluoroscopy]
Digital Angiography
Digital Subtraction Angiography (subtracts anatomy we don’t want to look at)
[Fluoroscopy]
Where is the tube for fluoroscopy? Except for?
Where should shielding be?
[Fluoroscopy]
Tube is in the table (w/ Bucky), except for remote rooms
Shielding should be between the pt and the table
[Fluoroscopy]
What exams do not get a curtain
[Fluoroscopy]
Anything that is a sterile field does not get the curtain
[Fluoroscopy]
What happens when the Bucky slot cover is broken off?
[Fluoroscopy]
Leakage from tube
[Fluoroscopic Technical Factors ]
Common mAs?
Warning bell purpose?
[Fluoroscopic Technical Factors]
1 - 5 mAs
5 minute warning bell to let practitioner know because most cases should be done in less time than that
[Fluoroscopic Technical Factors]
Patient dose is considerably high/low
_____, is selected for body type being imaged (chest, belly) and what _____ material we’re using
[Fluoroscopic Technical Factors]
Patient dose is considerably high
kVp, is selected for body type being imaged (chest, belly) and what contrast material we’re using
[Fluoroscopic Technical Factors]
If we’re using iodine, and I’m looking at vessels or kidneys we want to be around __kVp
If I’m looking at your GI tract, we have to be above kVp
Has to do with z number and k-edge, the average is about 35 and the k-edge is 33
pick ___ to match the contrast
[Fluoroscopic Technical Factors]
If we’re using iodine, and I’m looking at vessels or kidneys we want to be around 70KeV
If I’m looking at your GI tract, we have to be above 100KeV
Has to do with z number and k-edge, the average is about 35 and the k-edge is 33
pick kVp to match the contrast
[Fluoroscopic Technical Factors]
Image Brightness level is selected, and factors (mAs & kVp) are adjusted automatically to meet the required brightness level know as _____.
The advantage of ABC over AEC?
Disadvantage?
[Fluoroscopic Technical Factors]
Image Brightness level is selected, and factors (mAs & kVp) are adjusted automatically to meet the required brightness level know as ABC.
It’s advantage is that it will adjust the KV and the mA (the quality and quantity) while AEC only adjusts mA
It’s disadvantage is that it’s about a second behind (delayed)
[How do we provide a moving image? ]
Image intensification is necessary to provide _____ images.
Illumination is measured in _____ (L) and _____ (mL).
[How do we provide a moving image?]
Image intensification is necessary to provide quality images.
Illumination is measured in Lambert (L) and mililambers (mL).
[How do we provide a moving image?]
An aperture is used to control the _____ between the image intensifier and the television camera tube.
[How do we provide a moving image?]
An aperture is used to control the light between the image intensifier and the television camera tube.
[How do we provide a moving image?]
Human eyes use _____ and _____ to perceive light levels.
_____ - sensitive to low light 10-6 mL, night or scotopic, not good at seeing detail
_____ sensitive to bright light 10-2 mL, daylight or photopic, better at seeing detail
[How do we provide a moving image?]
Human eyes use rods and cones to perceive light levels.
Rod- sensitive to low light 10-6 mL, night or scotopic
Cones sensitive to bright light 10-2 mL, daylight or photopic
[How do we provide a moving image?]
With out intensification, _____ vision was used, requiring time to adjust to darkness.
Brightness increase allow for sight with _____ vision, faster and better (higher detail with cone vision).
mAs, kVp, and the anatomy control _____
[How do we provide a moving image?]
With out intensification, rod vision was used, requiring time to adjust to darkness.
Brightness increase allow for sight with cone vision, faster and better (higher detail with cone vision).
mAs, kVp, and the anatomy control brightness
[Brightness- why is it so important? ]
With out intensification, rod vision was used, requiring time to adjust to darkness.
Brightness increase allow for sight with cone vision, faster and better (higher detail with cone vision).
mAs, kVp, and the anatomy control brightness
[Brightness- why is it so important?]
With out intensification, rod vision was used, requiring time to adjust to darkness.
Brightness increase allow for sight with cone vision, faster and better (higher detail with cone vision).
mAs, kVp, and the anatomy control brightness
[Image Intensifier ]
Converts image forming x-rays into _____ photons of high _____.
The components to do this are housed in an envelope, creating a _____.
X-rays interact with phosphors (_____) creating light.
SCINTILLATOR the very first layer of this is made of _____ _____
Similar to the process seen with radiographic scintillators, however, CsI phosphors are designed to have little divergence and higher spatial resolution.
[Image Intensifier ]
Converts image forming x-rays into light photons of high intensity.
The components to do this are housed in an envelope, creating a vacuum.
X-rays interact with phosphors (CsI) creating light.
SCINTILLATOR the very first layer of this is made of cesium iodine
Similar to the process seen with radiographic scintillators, however, CsI phosphors are designed to have little divergence and higher spatial resolution.
The photocathode is adjacent to the intensifier and made of metal. The light emitted by the _____ (material) causes electrons to be emitted by the photocathode.
The number of electrons emitted is _____ proportional to the intensity of the light (which is proportional to the incident x-ray); however, many light photons are needed to emit one electron.
Starts out dim and spread out, shrunk down and sped up to be focused at anode (made of _____ _____) to be made brighter at the output phosphor
The photocathode is adjacent to the intensifier and made of metal. The light emitted by the CsI (material) causes electrons to be emitted by the photocathode.
The number of electrons emitted is directly proportional to the intensity of the light (which is proportional to the incident x-ray); however, many light photons are needed to emit one electron.
Starts out dim and spread out, shrunk down and sped up to be focused at anode (made of zinc cadmium sulfate) to be made brighter
A potential of 25 kV is maintained across the vacuum, so the electrons will travel to the opposite end or the anode. This anode has a small hole, where the electrons pass through to the _____ _____.
A potential of 25 kV is maintained across the vacuum, so the electrons will travel to the opposite end or the anode. This anode has a small hole, where the electrons pass through to the output phosphor.
The output phosphor then converts the electrons back to _____ _____.
The output phosphor then converts the electrons back to light photons.
Electrostatic focusing lenses are used to _____ the electron beam while maintaining its _____. The electrons have high velocity and are minified at the _____ _____.
Electrostatic focusing lenses are used to focus the electron beam while maintaining its integrity. The electrons have high velocity and are minified at the output phosphor.
The intensification is realized at this point where every electron creates __-__ times more light than it took to create it.
The intensification is realized at this point where every electron creates 50-75 times more light than it took to create it.
Flux gain is used to describe the ______ _____ at _____ compared to the number of incident x-rays.
Flux gain is used to describe the light photons at output compared to the number of incident x-rays.
Brightness gain is used to describe the intensifiers ability to _____ _____. Usually _____-_____.
Brightness gain is used to describe the intensifiers ability to increase illumination. Usually 5,000-30,000.
Minification gain is used to describe the _____ in _____ of _____ phosphor verses _____ phosphor.
Minification gain = Big dull image and I squished you down into a small intense image
Minification gain is used to describe the difference in size of output phosphor verses input phosphor.
Minification gain = Big dull image and I squished you down into a small intense image
What is convergence factor?
Its usual range?
Convergence factor - ratio of the illumination intensity at output to the radiation incident on input phosphor.
Usually 50-300.
Dualfield input phosphor size __ - __
Trifield input phosphor size __ - __ - __
Dualfield input phosphor size 25-17
Trifield input phosphor size 25-17-12
_____ and _____ = total brightness gain
FLUX and MINIFICATION = total brightness gain
Between _____ and _____ there is a screen (filter) that allows photons go through without coming back
Between ANODE and PHOSPHOR there is a screen (filter) that allows photons go through without coming back
Multi-field Image Intensification is also known as…
MAG MODE
The electrostatic lenses charge is increased when a SMALLER/LARGER input phosphor is used. Resulting in a SMALLER/LARGER electron beam at the output phosphor. The focal point is also moved CLOSER/FARTHER from the output phosphor.
The electrostatic lenses charge is increased when a SMALLER input phosphor is used. Resulting in a SMALLER electron beam at the output phosphor. The focal point is also moved FARTHER from the output phosphor.
If a curved input phosphor wasn’t used, the image would be _____
If a curved input phosphor wasn’t used, the image would be DISTORTED
Electronic collimation = _____ _____, two important things happen
1 - The area irradiated INCREASES/DECREASES, and INCREASES/DECREASES FoV
2 - Focusing lenses need to change their _____ to narrow the beam to not shrink down as much and image appears BIGGER/SMALLER
MORE/LESS technique?
MORE/LESS ESE?
MORE/LESS pt. dose? Why?
Electronic collimation = MAG MODE, two important things happen
1 - The area irradiated DECREASES, and DESCREASES FoV
2 - Focusing lenses need to change their CHARGE to narrow the beam to not shrink down as much and image appears BIGGER
MORE technique
MORE PT ENTRANCE SKIN EXPOSURE
SAME pt. dose because more technique, less FoV
The output phosphor must be connected to a viewing devise. Most commonly a _____ _____ _____. _____.
_____ and same size as output phosphor.
The output phosphor must be connected to a viewing devise. Most commonly a TV CAMERA TUBE. VIDICON.
SENSITIVE and same size as output phosphor.
The TV camera takes the _____ from output phosphor and converts it to an electrical _____ to be reconstructed as a TV image.
The TV camera takes the light from output phosphor and converts it to an electrical signal to be reconstructed as a TV image.
TV monitors can have _____ and _____ adjusters, which is an advantage.
Exams can also be stored _____.
TV monitors can have BRIGHTNESS and CONTRAST adjusters, which is an advantage.
Exams can also be stored electronically.
[Vidicon]
Components are housed in an envelope that creates a _____.
Electrons are released from the electron gun and a beam is constructed by the _____/_____ grids.
[Vidicon]
Components are housed in an envelope that creates a vacuum.
Electrons are released from the electron gun and a beam is constructed by the control/acceleration grids.
[Vidicon]
External electromagnetic coils _____ and _____ the beam.
At the anode the beam interacts with a _____ layered target.
The signal plate conducts electricity to the _____/_____ layer and produces an image by emitting _____.
[Vidicon]
External electromagnetic coils focus and direct the beam.
At the anode the beam interacts with a three layered target.
The signal plate conducts electricity to the target/photoconductive layer and produces an image by emitting light.
[Fiber optics]
Advantage?
Disadvantage?
[Fiber optics]
Advantage small, compact easy to move tower, sturdy
Disadvantage- no additional imaging devices, only spot film can be used.
[Lenses]
If _____ or _____ (reel of film) is used lens coupling is required. Careful handling is required.
Object lens converts output phosphor light into two _____ beams. This beam interrupted by beam splitting _____. A portion goes to the spot film and the other to the Vidicon.
[Lenses ]
If spot or cine (reel of film) is used lens coupling is required. Careful handling is required.
Object lens converts output phosphor light into two parallel beams. This beam interrupted by beam splitting mirror. A portion goes to the spot film and the other to the Vidicon.
(Beam splitting mirrors are movable and in place only when needed.)
[Lenses]
The parallel light beams are then _____ by lenses from the respective devices, camera or spot film. The lens alignment is critical in point for image quality.
[Lenses]
The parallel light beams are then focused by lenses from the respective devices, camera or spot film. The lens alignment is critical in point for image quality.
With photospots, doses are HIGHER/LOWER
LOWER
[Television Monitor]
The signal is _____ and transmitted by cable to the monitor, where a visible image is made.
The TV monitor and camera are the two ends of a _____ circuit system
The TV Monitor is essentially a _____ ray tube. It is very similar to the TV camera tube; however, it is larger and its anode is a _____ screen with a graphite lining.
[Television Monitor]
The signal is amplified and transmitted by cable to the monitor, where a visible image is made.
The TV monitor and camera are the two ends of a closed circuit system
The TV Monitor is essentially a cathode ray tube. It is very similar to the TV camera tube; however, it is larger and its anode is a florescent screen with a graphite lining.
[Television Monitor]
The input signal can be amplified and is proportional to the light intensity received by the TV camera.
The modulation is controlled by the _____ _____.
This _____ and the _____ _____ focus the electron beam onto the output phosphor, to produce a burst of _____.
[Television Monitor]
The input signal can be amplified and is proportional to the light intensity received by the TV camera.
The modulation is controlled by the control grid.
This grid and the external coil focus the electron beam onto the output phosphor, to produce a burst of light.
[The Image]
As the electron beams, TV camera and TV tube, move in synchronous timing to create an image the electromagnetic coil direct its pattern
This pattern is known as a _____ pattern.
This pattern starts at the _____ _____ part of the screen an works in a diagonal fashion down the screen.
[The Image]
As the electron beams, TV camera and TV tube, move in synchronous timing to create an image the electromagnetic coil direct its pattern
This pattern is known as a Raster pattern.
This pattern starts at the upper left part of the screen an works in a diagonal fashion down the screen.
[The Image]
Again, the pattern starts at the _____ of the screen and works its way _____, although , the second pattern is worked in _____ the lines of the first.
This is known as _____, two of these fields create one TV _____.
- The _____ lines are one field, the _____ lines are one field, and they make up together an entire _____
[The Image]
Again, the pattern starts at the top of the screen and works its way down, although , the second pattern is worked in between the lines of the first.
This is known as interlacing, two of these fields create one TV frame.
- The odd lines are one field, the even lines are one field, and they make up together an entire frame
Fields and frames per second for digital progressive and traditional interlaced?
Digital progressive:
fields = 1/30 sec
frames = 1/30 sec
Traditional interlaced:
fields = 1/60 or 2/30 sec
frames = 1/30 sec
As the TV signal is created, it immediately starts to fade, each new TV frame represents __ ms of new information.
___ line systems- ___ lines of active trace per frame ___ are available for hi resolution images
As the TV signal is created, it immediately starts to fade, each new TV frame represents 33 ms of new information.
525 line systems- 525 lines of active trace per frame 1024 are available for hi resolution images
What controls quality of the picture?
1 - vertical resolution (how many _____ in system) more _____ = closer together = BETTER/WORSE resolution
2 - horizontal resolution (_____) = how many times can I change the _____ _____ as I’m crossing the field. The more I can change it, the more specific I can be to represent an object
The ideal monitor has _____ vertical and horizontal resolution.
What controls quality of the picture?
1 - vertical resolution (how many lines in system) more lines = closer together = BETTER resolution
2 - horizontal resolution (bandpass) = how many times can I change the electron gun as I’m crossing the field. The more I can change it, the more specific I can be to represent an object
The ideal monitor has equal vertical and horizontal resolution.
Monitors display __ lp/mm, intensifiers display __ lp/mm. The monitor is the weakest link in the _____ chain.
Images are taken during Fluoroscopy to take advantage of the intensifiers superior _____.
Monitors display 2 lp/mm, intensifiers display 5 lp/mm. The monitor is the weakest link in the resolution chain.
Images are taken during Fluoroscopy to take advantage of the intensifiers superior resolution.
[How do we take images during Fluoroscopy? ]
Spot film cameras- Places receptor between _____ and _____. The receptor is removed from it’s lead housing and exposed using a HIGH/LOW mAs radiographic technique.
These mechanical alterations may take __-__ sec.
[How do we take images during Fluoroscopy?]
Spot film cameras- Places receptor between Patient and intensifier. The receptor is removed from it’s lead housing and exposed using a HIGH mAs radiographic technique.
These mechanical alterations may take 1-2 sec.
[How do we take images during Fluoroscopy?]
Several types of film use patterns are available. 1 on 1, 2,3,4, on one.
- Quality images are produced however, dose is LOWER/HIGHER and time is a factor.
[How do we take images during Fluoroscopy?]
Several types of film use patterns are available. 1 on 1, 2,3,4, on one.
- Quality images are produced however, dose is HIGHER and time is a factor.
Photospot camera - exposes one frame of film per x-ray exposure.
Image is derived form intensifier, using the beam splitting mirror therefor, less dose is needed. (BETTER/WORSE)
_____ time is required to take image.
70 and 105 mm film are used, _____ film generates better images.
Photospot camera - exposes one frame of film per x-ray exposure.
Image is derived form intensifier, using the beam splitting mirror therefor, less dose is needed. (BETTER)
Minimal time is required to take image.
70 and 105 mm film are used, larger film generates better images.
[Digital Fluoroscopy]
- The x-ray tube is _____, so that multiple static images are obtained. 1-__ per second (fields, common 4, 7, 15).
- _____ time is the measure of time it takes the tube to be turned on.
[Digital Fluoroscopy]
- The x-ray tube is pulsed, so that multiple static images are obtained. 1-60 per second (fields, common 4, 7, 15).
- Interrogation time is the measure of time it takes the tube to be turned on.
(we no longer use continuous fluoro, the x-ray tube is going to pulse)
[Digital Fluoroscopy]
_____ mode reading of the camera tube is necessary.
One entire field is transmitted in __ms. The signal to noise ratio is also very _____ (___:1) – providing _____ image quality.
- When do we use it?
[Digital Fluoroscopy]
Progressive mode reading of the camera tube is necessary.
One entire field is transmitted in 33ms. The signal to noise ratio is also very HIGH (1000:1) – providing HIGH image quality.
- We use progressive mode during digital fluoro.
1) The TV camera tube transmits the analog signal to the _____ to _____ Converter, where it is put in digital form. _____ then transfers information to memory for storage.
2) A CCD (charged coupled device) can be connected via _____, acquires the image and sends to the computer as a digital signal.
3) A FPIR (flat panel image receptor) can be used with out an _____.
1) The TV camera tube transmits the analog signal to the Analog to Digital Converter, where it is put in digital form. ADC then transfers information to memory for storage.
2) A CCD can be connected via optics, acquires the image and sends to the computer as a digital signal.
3) A FPIR (flat panel image receptor) can be used with out an II.
Image subtraction accomplished with digital images.
An _____ beam is needed for subtraction imaging
Two types are available:
• _____ subtraction
• _____ subtraction
Image subtraction accomplished with digital images.
An area beam is needed for subtraction imaging
Two types are available:
• Temporal subtraction
• Energy subtraction