Equipment Flashcards
Check valve in pipeline inlet assembly is a floating valve and seats according to pressure. What is the function of this valve?
Prevents oxygen cylinder gas from escaping the anesthesia machine
Power outlet valve for ventilator
DISS fitting
Ball and spring type valve
YOU supply the energy
Hanger yoke valve
Pins : Air - 1,5/ O2 - 2,5/ N2O - 3,5
Free floating valve
Prevents a full cylinder from emptying into an empty cylinder or wall oxygen from entering a cylinder
What is the name of the gauge that shows oxygen tank pressure?
Bourdon
You have a dual hanger yoke with 2 oxygen E cylinders. Both cylinders are turned on; which one will supply the gas machine?
The E cylinder with the highest pressure
Pressure Regulators
Diaphragm valves
ONLY role = REDUCE PRESSURE
First Stage Regulators
High (cylinder) pressure decreased to intermediate pressure (40-50 psig)
*Diaphragm valve
Second Stage Regulators
Intermediate pressure decreased to low pressure (16 psig)
Receives oxygen from first stage regulator OR wall pressure oxygen
*Diaphragm valve
Oxygen Flush Valve
Flow rate: 35-75 L/min
Pressurized: 40-50 psig (intermediate pressure)
Ball and spring valve
YOU supply the energy
What are the 2 negatives of the oxygen flush valve?
- May cause barotrauma
2. Dilutes anesthetic gas
Pressure Sensor Shut-Off Valve
Oxygen Failure Pressure Device
Senses oxygen pressure at intermediate pressure
Shuts off nitrous oxide if oxygen PRESSURE falls (< 25 psi)
Ball and spring valve
Oxygen Flow Meter
Flow Control Valve
Receives low pressure (16 psig)
Thorpe tube is gas specific and tapered
Check Valve Located b/t the Vaporizer Outlet and the Machine CGO
*Ohmeda machine only!
Prevents reversal of flow to vaporizer (back pressure)
Free-floating type valve
What happens if the inspiratory valve sticks open?
Expiratory volume will exhaust through the inspiratory limb
ETCO2 waveform will become elevated
What happens if the expiratory valve sticks open?
Inspired volume will not enter the ETT
It will exhaust through the expiratory limb
Inspiration and Expiration Check Valves
Flutter valves
Unidirectional flow
APL valve
Pop off
Attached to the exhalation check valve
Adjusts the limit of pressure in that patient circuit and rebreathing bag
Waste Gas Scavenging System
Adjusting the needle valve alters the FLOW of waste gases into the vacuum
What will happen when the vaporizer is tipped?
Liquid vapor will get into the vaporizer chamber
Carrier flow will carry MORE agent
Do NOT use, needs to be re-calibrated
High Pressure (cylinder)
Cylinder pressure
Bourdon gauge
Check valves in the hanger yoke
Index pins on the hanger yoke
Intermediate Pressure (40-50 psi)
Flush valve Second stage regulator Oxygen pressure "failsafe" shut-off valve Pressure gauges from pipeline DISS Flow control valve of manifold Flow meter valves Ventilator power inlet
Low Pressure (16 psi)
Vaporizer
CGO
Manifold/Flow meters
Check valves
Free Floating Valve
DISS
Dual hanger yoke systems
Prevent gases from leaking out of the system
Ball and Spring Valve
Oxygen flush valve
Oxygen pressure sensor valve
Diaphragm valve
First and second stage regulators
Reduce pressure
The path…
Oxygen cylinder - open cylinder valve - hanger yoke system - free floating valve - first stage regulator - NOW 4 options
What are the 4 options following the first stage regulator?
- Oxygen pressure sensor valve
- Flush valve
- DISS
- Second stage regulator - flow meters - vaporizer - CGO - soda lime - inspiratory valve - inspiratory limb - ETT
What type of anesthesia machines should be tested with a negative-pressure leak test?
Machines with check valves
Suction bulb is attached to the CGO - squeezed repeatedly until the bulb is fully collapsed - leak-free if the bulb remains collapsed for a least 10 sec
Your next patient has MH. What 3 steps will you take?
- Flush gas machine for 10 min
- Replace the breathing circuit and the CO2 canister
- Remove vaporizers
What is the purpose of the interlock system?
Prevents more than one vaporizer from being turned on at a time
What are your actions when the oxygen low-pressure alarm sounds?
D/c pipeline
Open E cylinder
Use low flow FGF
The vacuum control valve should be adjusted to allow the evacuation of what volume of waste gas per min?
10-15 L/min
What is the purpose of the proportioning system?
Prevent hypoxic mixture
N2O and O2 are mechanically/pneumatically linked
Min O2 concentration at CGO is b/t 23-25%
Max N2O-O2 flow ratio of 3:1
Who controls such processes as the filling and manufacturing of gas cylinders?
The US Department of Transportation
What is the working pressure of the hospital pipeline system?
50 psig
What is the most common contaminant of medical gas lines?
Water
Which 2 gases are in liquid form in pressurized cylinders?
- N2O
2. CO2
When should you change the O2 cylinder?
When < 1000 psi
Half full
The nitrous oxide tank reads 700 psi. What is the significance of this?
N2O pressure below 745 psi indications that the cylinder has NO liquid and is < than 1/4 full (400 L)
Change!
*Does NOT follow Boyle’s Law
Calibration of flowmeters is based upon what physical property of gases: density or viscostiy?
Low flows - viscosity - laminar
High flows - density - turbulent
All compressed gas cylinders are constructed according to…
Interstate Commerce Commission specifications
ICC3S
Must have 7 cylinder markings
Each cylinder must be subject to a test by interior hydrostatic pressure at least once every…
5 years
Explain the safety device on each valve stem on the cylinder.
Under hazardous conditions of excessive heat or fire the cylinder will become exhausted
This is a simple plug of soft metal alloy called Wood’s metal
What does it mean when you crack the cylinder?
Open the cylinder slightly to clear the outlet of possible dust
Open System
No Mask On Face
No dead space
No rebreathing or reservoir bag
No valves
Pediatric induction
Semi-Open
Mask On Face
No rebreathing
Unidirectional valve
Reservoir bag
Room pollution
High FGF
Open drop technique
Semi-Closed
Mask On Face
No room air inspired
CO2 absorber
Unidirectional valves
Reservoir bag
FGF can be < MV
Circle system
Closed
Mask On Face
Rebreathing gases - pop off closed
CO2 absorber
Unidirectional valves
Flow 150-500 mL/min - physiological requirements
Flow 150-250 mL/min - under anesthesia
Specific Semi-Open Systems
Mapleson Systems - 5 Components
No rebreathing Unidirectional valve Reservoir bag Room pollution High FGF
- Face mask
- Spring-loaded pop-off valve
- Reservoir tubing
- Fresh gas inflow tubing
- Reservoir bag
Mapleson A is best for what? Worst for what?
Best for spontaneously breathing patient
Worst for controlled ventilation
What is the most commonly used breathing system today?
Bain Circuit
Modification of the Mapleson D system
Which Mapleson systems are rarely used today?
A, B, C
Which Mapleson systems can you ventilate an apneic patient?
All Mapleson systems
Mapleson Systems
Preventing rebreathing during spontaneous ventilation
A > DFE > CB
Mapleson Systems
Preventing rebreathing during controlled ventilation
DFE > BC > A
Which Mapleson System has the fresh gas inlet and popoff valve at opposite ends?
A
Which Mapleson System has the fresh gas inlet at mask end, next to popoff valve?
B, C
Which Mapleson System has fresh gas inlet at mask end and popoff valve at other end — also called the T-piece group?
D, E, F
American Society for Testing and Materials standards for reservoir bags require that with a bag distended to ___x its normal capacity, the pressure will not exceed ___cmH2O.
4x
50 cmH2O
What are the 5 components of the circle system?
- Gas reservoir bag
- 2 corrugated tubes
- 2 unidirectional valves
- CO2 absorbent
- Overflow valve
What is the most common site for breathing circuit disconnection?
B/t breathing system and ETT
Where does the FGF enter the breathing circuit in the circle system?
Semi-closed system
B/t the absorbers and the inspiratory valve
In a circle system, where is dead space located?
B/t Y piece and the patient
When does a semi-closed, or semi-open, system exist?
When high FGF are used with a circle system
Which breathing circuits, open or closed, have the slowest induction time?
Closed